{"title":"前置法康眼内透镜对扫源光学相干断层成像生物测量的影响和校正。","authors":"Bruno Barbosa Ribeiro, Sílvia Monteiro, João Heitor Marques, Ana Carolina Abreu, Maria do Céu Pinto","doi":"10.3928/1081597X-20240716-01","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the impact of anterior chamber phakic intraocular lens (pIOL) on swept-source optical coherence tomography (SS-OCT) biometric measurements and IOL power calculation.</p><p><strong>Methods: </strong>This retrospective analysis of 67 eyes of 49 patients with previous anterior chamber pIOL implantation analyzed the accuracy of automatic segmentation of the anterior surface of the crystalline lens and its impact on anterior chamber depth (ACD, measured from epithelium to lens), lens thickness measurements, and IOL power calculation. The sample was divided into two groups: correct detection of the anterior surface of the crystalline lens and inaccurate detection. Segmentation of eyes from the inaccurate detection group was manually corrected and ACD and lens thickness were calculated using ImageJ software. IOL power was calculated using 7 formulas for both measurements.</p><p><strong>Results: </strong>The anterior surface of the crystalline lens was mis-identified in 13 (19.4%) eyes. ACD was underestimated (Δ -0.85 ± 0.33 mm, <i>P</i> < .001) and lens thickness was overestimated (Δ +0.81 ± 0.25 mm, <i>P</i> < .001). Manual correction changed the target spherical equivalent only in the Haigis formula (<i>P</i> = .009). After correction for segmentation bias, the Pearl DGS, Cooke K6, and EVO 2.0 formulas showed the lowest prediction error, with the Pearl DGS showing greatest accuracy within ±1.00 diopters of prediction error range (81.0%).</p><p><strong>Conclusions: </strong>SS-OCT biometry misidentifies the anterior surface of the crystalline lens in a significant proportion, resulting in significant IOL power calculation error in the Haigis formula. Manual proofing of segmentation is mandatory in every patient with anterior chamber pIOL implantation. After correct segmentation, the Pearl DGS, Cooke K6, and EVO seem to be the best formulas. <b>[<i>J Refract Surg</i>. 2024;40(8):e562-e568.]</b>.</p>","PeriodicalId":16951,"journal":{"name":"Journal of refractive surgery","volume":"40 8","pages":"e562-e568"},"PeriodicalIF":2.9000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Impact and Correction of an Anterior Phakic Intraocular Lens on Swept-Source Optical Coherence Tomography Biometry.\",\"authors\":\"Bruno Barbosa Ribeiro, Sílvia Monteiro, João Heitor Marques, Ana Carolina Abreu, Maria do Céu Pinto\",\"doi\":\"10.3928/1081597X-20240716-01\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To evaluate the impact of anterior chamber phakic intraocular lens (pIOL) on swept-source optical coherence tomography (SS-OCT) biometric measurements and IOL power calculation.</p><p><strong>Methods: </strong>This retrospective analysis of 67 eyes of 49 patients with previous anterior chamber pIOL implantation analyzed the accuracy of automatic segmentation of the anterior surface of the crystalline lens and its impact on anterior chamber depth (ACD, measured from epithelium to lens), lens thickness measurements, and IOL power calculation. The sample was divided into two groups: correct detection of the anterior surface of the crystalline lens and inaccurate detection. Segmentation of eyes from the inaccurate detection group was manually corrected and ACD and lens thickness were calculated using ImageJ software. IOL power was calculated using 7 formulas for both measurements.</p><p><strong>Results: </strong>The anterior surface of the crystalline lens was mis-identified in 13 (19.4%) eyes. ACD was underestimated (Δ -0.85 ± 0.33 mm, <i>P</i> < .001) and lens thickness was overestimated (Δ +0.81 ± 0.25 mm, <i>P</i> < .001). Manual correction changed the target spherical equivalent only in the Haigis formula (<i>P</i> = .009). After correction for segmentation bias, the Pearl DGS, Cooke K6, and EVO 2.0 formulas showed the lowest prediction error, with the Pearl DGS showing greatest accuracy within ±1.00 diopters of prediction error range (81.0%).</p><p><strong>Conclusions: </strong>SS-OCT biometry misidentifies the anterior surface of the crystalline lens in a significant proportion, resulting in significant IOL power calculation error in the Haigis formula. Manual proofing of segmentation is mandatory in every patient with anterior chamber pIOL implantation. After correct segmentation, the Pearl DGS, Cooke K6, and EVO seem to be the best formulas. <b>[<i>J Refract Surg</i>. 2024;40(8):e562-e568.]</b>.</p>\",\"PeriodicalId\":16951,\"journal\":{\"name\":\"Journal of refractive surgery\",\"volume\":\"40 8\",\"pages\":\"e562-e568\"},\"PeriodicalIF\":2.9000,\"publicationDate\":\"2024-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of refractive surgery\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3928/1081597X-20240716-01\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of refractive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3928/1081597X-20240716-01","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:评估前房人工晶体(pIOL)对扫源光学相干断层扫描(SS-OCT)生物测量和人工晶体功率计算的影响:该回顾性分析对 49 名曾植入前房人工晶体的患者的 67 只眼睛进行了分析,分析了晶状体前表面自动分割的准确性及其对前房深度(ACD,从上皮到晶状体的测量值)、晶状体厚度测量和人工晶体功率计算的影响。样本分为两组:晶状体前表面检测正确组和检测不准确组。对检测不准确组的眼睛进行人工校正分割,并使用 ImageJ 软件计算 ACD 和晶状体厚度。两种测量结果均使用 7 种公式计算人工晶体功率:结果:13 只眼睛(19.4%)的晶状体前表面识别错误。ACD被低估(Δ -0.85 ± 0.33 mm,P < .001),晶状体厚度被高估(Δ +0.81 ± 0.25 mm,P < .001)。手动校正只改变了海吉斯公式中的目标球面等值(P = .009)。修正分割偏差后,Pearl DGS、Cooke K6 和 EVO 2.0 公式的预测误差最小,其中 Pearl DGS 在 ±1.00 斜度的预测误差范围内(81.0%)的准确度最高:SS-OCT 生物测量法错误识别晶状体前表面的比例很高,导致海吉斯公式中 IOL 功率计算出现重大误差。对每一位接受前房人工晶体植入术的患者,都必须进行人工晶体分割校对。在正确分割后,Pearl DGS、Cooke K6 和 EVO 似乎是最好的公式。[J Refract Surg. 2024;40(8):e562-e568]。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Impact and Correction of an Anterior Phakic Intraocular Lens on Swept-Source Optical Coherence Tomography Biometry.

Purpose: To evaluate the impact of anterior chamber phakic intraocular lens (pIOL) on swept-source optical coherence tomography (SS-OCT) biometric measurements and IOL power calculation.

Methods: This retrospective analysis of 67 eyes of 49 patients with previous anterior chamber pIOL implantation analyzed the accuracy of automatic segmentation of the anterior surface of the crystalline lens and its impact on anterior chamber depth (ACD, measured from epithelium to lens), lens thickness measurements, and IOL power calculation. The sample was divided into two groups: correct detection of the anterior surface of the crystalline lens and inaccurate detection. Segmentation of eyes from the inaccurate detection group was manually corrected and ACD and lens thickness were calculated using ImageJ software. IOL power was calculated using 7 formulas for both measurements.

Results: The anterior surface of the crystalline lens was mis-identified in 13 (19.4%) eyes. ACD was underestimated (Δ -0.85 ± 0.33 mm, P < .001) and lens thickness was overestimated (Δ +0.81 ± 0.25 mm, P < .001). Manual correction changed the target spherical equivalent only in the Haigis formula (P = .009). After correction for segmentation bias, the Pearl DGS, Cooke K6, and EVO 2.0 formulas showed the lowest prediction error, with the Pearl DGS showing greatest accuracy within ±1.00 diopters of prediction error range (81.0%).

Conclusions: SS-OCT biometry misidentifies the anterior surface of the crystalline lens in a significant proportion, resulting in significant IOL power calculation error in the Haigis formula. Manual proofing of segmentation is mandatory in every patient with anterior chamber pIOL implantation. After correct segmentation, the Pearl DGS, Cooke K6, and EVO seem to be the best formulas. [J Refract Surg. 2024;40(8):e562-e568.].

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
5.10
自引率
12.50%
发文量
160
审稿时长
4-8 weeks
期刊介绍: The Journal of Refractive Surgery, the official journal of the International Society of Refractive Surgery, a partner of the American Academy of Ophthalmology, has been a monthly peer-reviewed forum for original research, review, and evaluation of refractive and lens-based surgical procedures for more than 30 years. Practical, clinically valuable articles provide readers with the most up-to-date information regarding advances in the field of refractive surgery. Begin to explore the Journal and all of its great benefits such as: • Columns including “Translational Science,” “Surgical Techniques,” and “Biomechanics” • Supplemental videos and materials available for many articles • Access to current articles, as well as several years of archived content • Articles posted online just 2 months after acceptance.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信