{"title":"与机器人辅助全髋关节置换术相关的不良事件:基于 FDA MAUDE 数据库的分析。","authors":"S Bradley Graefe, Gregory J Kirchner, Natalie K Pahapill, Hannah H Nam, Mark L Dunleavy, Nikkole Haines","doi":"10.1177/11207000241263315","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The Food and Drug Administration (FDA) maintains the Manufacturer and User Facility Device Experience (MAUDE) database for reporting adverse events associated with medical devices, including emerging technologies, such as robotic-assisted total hip arthroplasty (THA). Aim of this study was to evaluate the variation of adverse events associated with robotics in THA.</p><p><strong>Methods: </strong>Medical device reports (MDRs) within the MAUDE database were identified between 2017 and 2021. For MDR identification the product class \"orthopaedic stereotaxic equipment\" and terms associated with THA were used. Individual adverse events were identified and organised by type and consequences, such as patient injury, surgical delay, or conversion to the manual technique.</p><p><strong>Results: </strong>521 MDRs constituting 546 discrete events were found. The most common reported complication was intraoperative hardware failure (304/546, 55.7%), among which the most common failure was a broken impaction handle/platform (110, 20.1%). Inaccurate cup placement was the second most common reported complication (63, 11.5%). Abandoning the robot occurred in 13.0% (71/521) of reports. A surgical delay was noted in 28% (146/521) of reports, with an average delay of 17.9 (range 1-60) minutes.</p><p><strong>Conclusions: </strong>Identifying complications that may occur with robotics in THA is an important first step in preventing adverse events and surgical delays. Database analysis provide an overview of the range of complications.</p>","PeriodicalId":12911,"journal":{"name":"HIP International","volume":null,"pages":null},"PeriodicalIF":1.3000,"publicationDate":"2024-08-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Adverse events associated with robotic-assistance in total hip arthroplasty: an analysis based on the FDA MAUDE database.\",\"authors\":\"S Bradley Graefe, Gregory J Kirchner, Natalie K Pahapill, Hannah H Nam, Mark L Dunleavy, Nikkole Haines\",\"doi\":\"10.1177/11207000241263315\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>The Food and Drug Administration (FDA) maintains the Manufacturer and User Facility Device Experience (MAUDE) database for reporting adverse events associated with medical devices, including emerging technologies, such as robotic-assisted total hip arthroplasty (THA). Aim of this study was to evaluate the variation of adverse events associated with robotics in THA.</p><p><strong>Methods: </strong>Medical device reports (MDRs) within the MAUDE database were identified between 2017 and 2021. For MDR identification the product class \\\"orthopaedic stereotaxic equipment\\\" and terms associated with THA were used. Individual adverse events were identified and organised by type and consequences, such as patient injury, surgical delay, or conversion to the manual technique.</p><p><strong>Results: </strong>521 MDRs constituting 546 discrete events were found. The most common reported complication was intraoperative hardware failure (304/546, 55.7%), among which the most common failure was a broken impaction handle/platform (110, 20.1%). Inaccurate cup placement was the second most common reported complication (63, 11.5%). Abandoning the robot occurred in 13.0% (71/521) of reports. A surgical delay was noted in 28% (146/521) of reports, with an average delay of 17.9 (range 1-60) minutes.</p><p><strong>Conclusions: </strong>Identifying complications that may occur with robotics in THA is an important first step in preventing adverse events and surgical delays. Database analysis provide an overview of the range of complications.</p>\",\"PeriodicalId\":12911,\"journal\":{\"name\":\"HIP International\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-08-04\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"HIP International\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1177/11207000241263315\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"ORTHOPEDICS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"HIP International","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/11207000241263315","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

摘要

背景:美国食品和药物管理局(FDA)设有制造商和用户设施设备经验(MAUDE)数据库,用于报告与医疗设备相关的不良事件,包括机器人辅助全髋关节置换术(THA)等新兴技术。本研究旨在评估机器人辅助全髋关节置换术相关不良事件的变化情况:在 MAUDE 数据库中识别了 2017 年至 2021 年期间的医疗器械报告 (MDR)。在识别MDR时,使用了 "骨科立体定向设备 "产品类别以及与THA相关的术语。对单个不良事件进行了识别,并按类型和后果(如患者受伤、手术延迟或改用手动技术)进行了分类:结果:共发现521个MDR,构成546个离散事件。最常见的并发症是术中硬件故障(304/546,55.7%),其中最常见的故障是撞击手柄/平台断裂(110,20.1%)。髋臼杯放置不准确是第二常见的并发症(63 例,11.5%)。13.0%(71/521)的报告中出现了放弃机器人的情况。28%的报告(146/521)指出手术延迟,平均延迟时间为17.9分钟(1-60分钟不等):识别THA机器人手术可能出现的并发症是预防不良事件和手术延迟的重要第一步。数据库分析提供了一系列并发症的概况。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Adverse events associated with robotic-assistance in total hip arthroplasty: an analysis based on the FDA MAUDE database.

Background: The Food and Drug Administration (FDA) maintains the Manufacturer and User Facility Device Experience (MAUDE) database for reporting adverse events associated with medical devices, including emerging technologies, such as robotic-assisted total hip arthroplasty (THA). Aim of this study was to evaluate the variation of adverse events associated with robotics in THA.

Methods: Medical device reports (MDRs) within the MAUDE database were identified between 2017 and 2021. For MDR identification the product class "orthopaedic stereotaxic equipment" and terms associated with THA were used. Individual adverse events were identified and organised by type and consequences, such as patient injury, surgical delay, or conversion to the manual technique.

Results: 521 MDRs constituting 546 discrete events were found. The most common reported complication was intraoperative hardware failure (304/546, 55.7%), among which the most common failure was a broken impaction handle/platform (110, 20.1%). Inaccurate cup placement was the second most common reported complication (63, 11.5%). Abandoning the robot occurred in 13.0% (71/521) of reports. A surgical delay was noted in 28% (146/521) of reports, with an average delay of 17.9 (range 1-60) minutes.

Conclusions: Identifying complications that may occur with robotics in THA is an important first step in preventing adverse events and surgical delays. Database analysis provide an overview of the range of complications.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
HIP International
HIP International 医学-整形外科
CiteScore
4.20
自引率
0.00%
发文量
70
审稿时长
2 months
期刊介绍: HIP International is the official journal of the European Hip Society. It is the only international, peer-reviewed, bi-monthly journal dedicated to diseases of the hip. HIP International considers contributions relating to hip surgery, traumatology of the hip, prosthetic surgery, biomechanics, and basic sciences relating to the hip. HIP International invites reviews from leading specialists with the aim of informing its readers of current evidence-based best practice. The journal also publishes supplements containing proceedings of symposia, special meetings or articles of special educational merit. HIP International is divided into six independent sections led by editors of the highest scientific merit. These sections are: • Biomaterials • Biomechanics • Conservative Hip Surgery • Paediatrics • Primary and Revision Hip Arthroplasty • Traumatology
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信