{"title":"不同柱塞尺寸和柱塞位置对摘除软性隐形眼镜和硬性透气性隐形眼镜的生物力学评估。","authors":"Min-Chien Hsiao, Yu-Chun Yen, Chun-Hsiang Wang, Yen-Nien Chen, Shun-Ping Wang, Kuo-Chih Su","doi":"10.3233/THC-231983","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>To avoid risks of mucosal infection from contact lenses removal, a contact lens plunger is often used.</p><p><strong>Objective: </strong>Given various types of contact lens plungers available on the market, no study has yet been done on mechanical effects of the contact lens plunger on contact lens removal. Here, this study used finite element analysis to investigate the effects of plunger size and plunger position on the removal of soft and rigid gas permeable (RGP) contact lenses.</p><p><strong>Methods: </strong>First, we established finite element analysis models for the plunger, contact lens, cornea, and aqueous humor. The plunger is made of mostly silicone rubber, and the contact lenses are mainly made of soft and hard material. The part of the plunger used for removal was located either at the central or the edged position, with pulling 1 mm distance. The main parameters observation indicators of in this study were the reaction force at the fixed end of the cornea, aqueous humor, the von Mises stress of the plunger, the contact lenses, and the cornea.</p><p><strong>Results: </strong>Results of this study showed that when a plunger of a larger diameter was used, the reaction force of the plunger was also larger, especially when applied to RGP lenses, which required a slightly larger force (∼0.27 N). Also, when removing a RGP lens from the edge, there was a greater stress at the edge of the contact lens (2.5799 MPa), and this caused a higher stress on both the cornea (0.0165 MPa) and the aqueous humor (0.00114 MPa).</p><p><strong>Conclusions: </strong>When using a plunger with a larger diameter to remove a RGP lens, although a larger force required, the relatively larger contact area likely reduced the stress on the cornea and aqueous humor, thereby reducing the risk of eye injury. In addition, when removing a RGP lens, the results of this study recommended it to be removed from the plunger edge, as that facilitated the removal of contact lens.</p>","PeriodicalId":48978,"journal":{"name":"Technology and Health Care","volume":null,"pages":null},"PeriodicalIF":1.4000,"publicationDate":"2024-07-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Biomechanical evaluation of different plunger size and plunger position on removing soft contact lenses and rigid gas permeable contact lenses.\",\"authors\":\"Min-Chien Hsiao, Yu-Chun Yen, Chun-Hsiang Wang, Yen-Nien Chen, Shun-Ping Wang, Kuo-Chih Su\",\"doi\":\"10.3233/THC-231983\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>To avoid risks of mucosal infection from contact lenses removal, a contact lens plunger is often used.</p><p><strong>Objective: </strong>Given various types of contact lens plungers available on the market, no study has yet been done on mechanical effects of the contact lens plunger on contact lens removal. Here, this study used finite element analysis to investigate the effects of plunger size and plunger position on the removal of soft and rigid gas permeable (RGP) contact lenses.</p><p><strong>Methods: </strong>First, we established finite element analysis models for the plunger, contact lens, cornea, and aqueous humor. The plunger is made of mostly silicone rubber, and the contact lenses are mainly made of soft and hard material. The part of the plunger used for removal was located either at the central or the edged position, with pulling 1 mm distance. The main parameters observation indicators of in this study were the reaction force at the fixed end of the cornea, aqueous humor, the von Mises stress of the plunger, the contact lenses, and the cornea.</p><p><strong>Results: </strong>Results of this study showed that when a plunger of a larger diameter was used, the reaction force of the plunger was also larger, especially when applied to RGP lenses, which required a slightly larger force (∼0.27 N). Also, when removing a RGP lens from the edge, there was a greater stress at the edge of the contact lens (2.5799 MPa), and this caused a higher stress on both the cornea (0.0165 MPa) and the aqueous humor (0.00114 MPa).</p><p><strong>Conclusions: </strong>When using a plunger with a larger diameter to remove a RGP lens, although a larger force required, the relatively larger contact area likely reduced the stress on the cornea and aqueous humor, thereby reducing the risk of eye injury. In addition, when removing a RGP lens, the results of this study recommended it to be removed from the plunger edge, as that facilitated the removal of contact lens.</p>\",\"PeriodicalId\":48978,\"journal\":{\"name\":\"Technology and Health Care\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.4000,\"publicationDate\":\"2024-07-11\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Technology and Health Care\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.3233/THC-231983\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Technology and Health Care","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.3233/THC-231983","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

摘要

背景:为了避免摘除隐形眼镜时粘膜感染的风险,通常使用隐形眼镜活塞:为了避免摘除隐形眼镜时粘膜感染的风险,通常会使用隐形眼镜柱塞:鉴于市场上有各种类型的隐形眼镜柱塞,目前还没有关于隐形眼镜柱塞对摘除隐形眼镜的机械影响的研究。在此,本研究采用有限元分析法研究了柱塞尺寸和柱塞位置对摘除软性和硬性透气(RGP)隐形眼镜的影响:首先,我们建立了柱塞、隐形眼镜、角膜和房水的有限元分析模型。柱塞主要由硅橡胶制成,隐形眼镜主要由软质和硬质材料制成。用于摘除的柱塞部分位于中央或边缘位置,拉动距离为 1 毫米。本研究的主要参数观察指标为角膜固定端的反作用力、房水、柱塞、隐形眼镜和角膜的冯-米塞斯应力:研究结果表明,当使用直径较大的柱塞时,柱塞的反作用力也较大,尤其是在使用 RGP 镜片时,所需的力稍大(∼0.27 N)。此外,从边缘移除 RGP 镜片时,隐形眼镜边缘的应力更大(2.5799 兆帕),这导致角膜(0.0165 兆帕)和房水(0.00114 兆帕)的应力增大:结论:使用直径较大的柱塞摘除 RGP 镜片时,虽然需要较大的力量,但相对较大的接触面积可能会减少角膜和房水所受的压力,从而降低眼睛受伤的风险。此外,在摘除 RGP 镜片时,本研究结果建议从柱塞边缘摘除,因为这有利于摘除隐形眼镜。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Biomechanical evaluation of different plunger size and plunger position on removing soft contact lenses and rigid gas permeable contact lenses.

Background: To avoid risks of mucosal infection from contact lenses removal, a contact lens plunger is often used.

Objective: Given various types of contact lens plungers available on the market, no study has yet been done on mechanical effects of the contact lens plunger on contact lens removal. Here, this study used finite element analysis to investigate the effects of plunger size and plunger position on the removal of soft and rigid gas permeable (RGP) contact lenses.

Methods: First, we established finite element analysis models for the plunger, contact lens, cornea, and aqueous humor. The plunger is made of mostly silicone rubber, and the contact lenses are mainly made of soft and hard material. The part of the plunger used for removal was located either at the central or the edged position, with pulling 1 mm distance. The main parameters observation indicators of in this study were the reaction force at the fixed end of the cornea, aqueous humor, the von Mises stress of the plunger, the contact lenses, and the cornea.

Results: Results of this study showed that when a plunger of a larger diameter was used, the reaction force of the plunger was also larger, especially when applied to RGP lenses, which required a slightly larger force (∼0.27 N). Also, when removing a RGP lens from the edge, there was a greater stress at the edge of the contact lens (2.5799 MPa), and this caused a higher stress on both the cornea (0.0165 MPa) and the aqueous humor (0.00114 MPa).

Conclusions: When using a plunger with a larger diameter to remove a RGP lens, although a larger force required, the relatively larger contact area likely reduced the stress on the cornea and aqueous humor, thereby reducing the risk of eye injury. In addition, when removing a RGP lens, the results of this study recommended it to be removed from the plunger edge, as that facilitated the removal of contact lens.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Technology and Health Care
Technology and Health Care HEALTH CARE SCIENCES & SERVICES-ENGINEERING, BIOMEDICAL
CiteScore
2.10
自引率
6.20%
发文量
282
审稿时长
>12 weeks
期刊介绍: Technology and Health Care is intended to serve as a forum for the presentation of original articles and technical notes, observing rigorous scientific standards. Furthermore, upon invitation, reviews, tutorials, discussion papers and minisymposia are featured. The main focus of THC is related to the overlapping areas of engineering and medicine. The following types of contributions are considered: 1.Original articles: New concepts, procedures and devices associated with the use of technology in medical research and clinical practice are presented to a readership with a widespread background in engineering and/or medicine. In particular, the clinical benefit deriving from the application of engineering methods and devices in clinical medicine should be demonstrated. Typically, full length original contributions have a length of 4000 words, thereby taking duly into account figures and tables. 2.Technical Notes and Short Communications: Technical Notes relate to novel technical developments with relevance for clinical medicine. In Short Communications, clinical applications are shortly described. 3.Both Technical Notes and Short Communications typically have a length of 1500 words. Reviews and Tutorials (upon invitation only): Tutorial and educational articles for persons with a primarily medical background on principles of engineering with particular significance for biomedical applications and vice versa are presented. The Editorial Board is responsible for the selection of topics. 4.Minisymposia (upon invitation only): Under the leadership of a Special Editor, controversial or important issues relating to health care are highlighted and discussed by various authors. 5.Letters to the Editors: Discussions or short statements (not indexed).
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信