{"title":"自动微流控白细胞介素-6 免疫分析仪的分析和临床验证","authors":"Jianbo Yang, Mustafa A Barbhuiya, Christopher Hamilton, Kimberly Robyak, Yusheng Zhu","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>We validated an automated microfluidic interleukin-6 (IL-6) immunoassay on the Ella platform for clinical use.</p><p><strong>Methods: </strong>The assay was validated for precision, lower limit of quantification, analytical measurement range, accuracy, specificity, interference of biotin, tocilizumab, GP130, IL-6R<i>α</i>, hemolysis, icterus, and lipemia, and establishing the reference value. The clinical performance was evaluated in 96 COVID-19 patients.</p><p><strong>Results: </strong>The within-run and between-run coefficient of variations (CV) ranged 1.8%-4.8%. This assay has an analytical measurement range (AMR) 0.7-2652 pg/ml. There was a moderate correlation between Ella IL-6 assay (y) and a Luminex Quantitative Multiplex Bead Assay in a reference lab (x): y=8.561x-475.38, R=0.5047, SEE=1592.8 pg/mL, N=48). Measurement of IL-6 in plasma samples from 45 healthy adults showed the upper limit of reference of 5.0 pg/mL. 95.83% (95% CI: 89.67%-98.85%) of COVID-19 patients had IL-6 >5.0 pg/mL. This assay is resistant to the interference from hemoglobin up to 1,144 mg/dL, triglyceride 1,699 mg/dL, bilirubin 35 mg/dL, biotin 1000 ng/mL, tocilizumab 240 μg/mL, IL-11 50,000 pg/mL, GP130 50,000 pg/mL, and IL-6R<i>α</i> 1,000 pg/mL.</p><p><strong>Conclusions: </strong>The IL-6 assay on the Ella platform is robust with minimum manual operation. The analytical and clinical performance characteristics meet the clinical needs.</p>","PeriodicalId":8228,"journal":{"name":"Annals of clinical and laboratory science","volume":"54 3","pages":"388-393"},"PeriodicalIF":1.1000,"publicationDate":"2024-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Analytical and Clinical Validation of an Automated Microfluidic Interleukin-6 Immunoassay.\",\"authors\":\"Jianbo Yang, Mustafa A Barbhuiya, Christopher Hamilton, Kimberly Robyak, Yusheng Zhu\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>We validated an automated microfluidic interleukin-6 (IL-6) immunoassay on the Ella platform for clinical use.</p><p><strong>Methods: </strong>The assay was validated for precision, lower limit of quantification, analytical measurement range, accuracy, specificity, interference of biotin, tocilizumab, GP130, IL-6R<i>α</i>, hemolysis, icterus, and lipemia, and establishing the reference value. The clinical performance was evaluated in 96 COVID-19 patients.</p><p><strong>Results: </strong>The within-run and between-run coefficient of variations (CV) ranged 1.8%-4.8%. This assay has an analytical measurement range (AMR) 0.7-2652 pg/ml. There was a moderate correlation between Ella IL-6 assay (y) and a Luminex Quantitative Multiplex Bead Assay in a reference lab (x): y=8.561x-475.38, R=0.5047, SEE=1592.8 pg/mL, N=48). Measurement of IL-6 in plasma samples from 45 healthy adults showed the upper limit of reference of 5.0 pg/mL. 95.83% (95% CI: 89.67%-98.85%) of COVID-19 patients had IL-6 >5.0 pg/mL. This assay is resistant to the interference from hemoglobin up to 1,144 mg/dL, triglyceride 1,699 mg/dL, bilirubin 35 mg/dL, biotin 1000 ng/mL, tocilizumab 240 μg/mL, IL-11 50,000 pg/mL, GP130 50,000 pg/mL, and IL-6R<i>α</i> 1,000 pg/mL.</p><p><strong>Conclusions: </strong>The IL-6 assay on the Ella platform is robust with minimum manual operation. The analytical and clinical performance characteristics meet the clinical needs.</p>\",\"PeriodicalId\":8228,\"journal\":{\"name\":\"Annals of clinical and laboratory science\",\"volume\":\"54 3\",\"pages\":\"388-393\"},\"PeriodicalIF\":1.1000,\"publicationDate\":\"2024-05-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Annals of clinical and laboratory science\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of clinical and laboratory science","FirstCategoryId":"3","ListUrlMain":"","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
Analytical and Clinical Validation of an Automated Microfluidic Interleukin-6 Immunoassay.
Objective: We validated an automated microfluidic interleukin-6 (IL-6) immunoassay on the Ella platform for clinical use.
Methods: The assay was validated for precision, lower limit of quantification, analytical measurement range, accuracy, specificity, interference of biotin, tocilizumab, GP130, IL-6Rα, hemolysis, icterus, and lipemia, and establishing the reference value. The clinical performance was evaluated in 96 COVID-19 patients.
Results: The within-run and between-run coefficient of variations (CV) ranged 1.8%-4.8%. This assay has an analytical measurement range (AMR) 0.7-2652 pg/ml. There was a moderate correlation between Ella IL-6 assay (y) and a Luminex Quantitative Multiplex Bead Assay in a reference lab (x): y=8.561x-475.38, R=0.5047, SEE=1592.8 pg/mL, N=48). Measurement of IL-6 in plasma samples from 45 healthy adults showed the upper limit of reference of 5.0 pg/mL. 95.83% (95% CI: 89.67%-98.85%) of COVID-19 patients had IL-6 >5.0 pg/mL. This assay is resistant to the interference from hemoglobin up to 1,144 mg/dL, triglyceride 1,699 mg/dL, bilirubin 35 mg/dL, biotin 1000 ng/mL, tocilizumab 240 μg/mL, IL-11 50,000 pg/mL, GP130 50,000 pg/mL, and IL-6Rα 1,000 pg/mL.
Conclusions: The IL-6 assay on the Ella platform is robust with minimum manual operation. The analytical and clinical performance characteristics meet the clinical needs.
期刊介绍:
The Annals of Clinical & Laboratory Science
welcomes manuscripts that report research in clinical
science, including pathology, clinical chemistry,
biotechnology, molecular biology, cytogenetics,
microbiology, immunology, hematology, transfusion
medicine, organ and tissue transplantation, therapeutics, toxicology, and clinical informatics.