{"title":"不同血样和抗凝剂类型对 B 型钠尿肽定量检测的影响","authors":"Liangping Yin, Qian Wang, Zhikui Jiang, Xiufang Meng, Yuying Wang","doi":"10.7754/Clin.Lab.2024.240102","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>BNP is a sensitive and widely used biomarker for an early diagnosis of heart failure. Currently, most commercial BNP detection products use EDTA plasma samples. The aim of this study was to evaluate the clinical performance of the BNP test by using whole blood samples compared to plasma samples, and to evaluate the effect of the anticoagulant type on the BNP test result.</p><p><strong>Methods: </strong>In total, 106 patients with different BNP levels from the Dahua Hospital volunteered for this study. Clinically homogenous samples, including EDTA anticoagulant plasma, EDTA whole blood, and heparin anticoagulant plasma, were collected and analyzed by using i-Reader S automatic immuno-analyzer and its supporting reagent kits. Pearson's correlation and weighted least squares linear regression analysis, Bland-Altman plotting, and Kappa test were used for statistical analysis.</p><p><strong>Results: </strong>Correlation analysis showed that BNP concentrations, measured from EDTA anticoagulated plasma samples, had a good linear regression relationship with BNP from whole blood samples, with a slope of 0.9477, r = 0.9978, p < 0.05. A similar correlation was observed between EDTA anticoagulated plasma samples and heparin anticoagulant plasma, with a slope of 0.8413, r = 0.9793, p < 0.05. The BNP concentration measured from the heparin plasma samples were lower than of the EDTA plasma samples. Bland-Altman analysis for assessing BNP concentration agreement showed there was no outlier ratio between EDTA whole blood and EDTA plasma within the range of the detection system, as well as no outlier between EDTA anticoagulated and heparin anticoagulant plasma. Kappa coefficient of BNP concentration between homologous EDTA anticoagulated and heparin anticoagulant plasma was 0.8553 (p < 0.001), and for EDTA anticoagulated plasma and homologous whole blood it was 0.8941 (p < 0.001).</p><p><strong>Conclusions: </strong>The diagnostic performance of EDTA anticoagulated whole blood samples did not differ significantly from EDTA anticoagulated plasma samples for the BNP test. This study showed no big significant difference between EDTA anticoagulated and heparin anticoagulated plasma measurements within 2 hours. The type of anticoagulant should be carefully chosen when performing the BNP test if BNP samples were in vitro for a long time.</p>","PeriodicalId":10384,"journal":{"name":"Clinical laboratory","volume":null,"pages":null},"PeriodicalIF":0.7000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"The Effects of Different Blood Sample and Anticoagulant Types on Quantitative Tests of B-Type Natriuretic Peptide.\",\"authors\":\"Liangping Yin, Qian Wang, Zhikui Jiang, Xiufang Meng, Yuying Wang\",\"doi\":\"10.7754/Clin.Lab.2024.240102\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>BNP is a sensitive and widely used biomarker for an early diagnosis of heart failure. Currently, most commercial BNP detection products use EDTA plasma samples. The aim of this study was to evaluate the clinical performance of the BNP test by using whole blood samples compared to plasma samples, and to evaluate the effect of the anticoagulant type on the BNP test result.</p><p><strong>Methods: </strong>In total, 106 patients with different BNP levels from the Dahua Hospital volunteered for this study. Clinically homogenous samples, including EDTA anticoagulant plasma, EDTA whole blood, and heparin anticoagulant plasma, were collected and analyzed by using i-Reader S automatic immuno-analyzer and its supporting reagent kits. Pearson's correlation and weighted least squares linear regression analysis, Bland-Altman plotting, and Kappa test were used for statistical analysis.</p><p><strong>Results: </strong>Correlation analysis showed that BNP concentrations, measured from EDTA anticoagulated plasma samples, had a good linear regression relationship with BNP from whole blood samples, with a slope of 0.9477, r = 0.9978, p < 0.05. A similar correlation was observed between EDTA anticoagulated plasma samples and heparin anticoagulant plasma, with a slope of 0.8413, r = 0.9793, p < 0.05. The BNP concentration measured from the heparin plasma samples were lower than of the EDTA plasma samples. Bland-Altman analysis for assessing BNP concentration agreement showed there was no outlier ratio between EDTA whole blood and EDTA plasma within the range of the detection system, as well as no outlier between EDTA anticoagulated and heparin anticoagulant plasma. Kappa coefficient of BNP concentration between homologous EDTA anticoagulated and heparin anticoagulant plasma was 0.8553 (p < 0.001), and for EDTA anticoagulated plasma and homologous whole blood it was 0.8941 (p < 0.001).</p><p><strong>Conclusions: </strong>The diagnostic performance of EDTA anticoagulated whole blood samples did not differ significantly from EDTA anticoagulated plasma samples for the BNP test. This study showed no big significant difference between EDTA anticoagulated and heparin anticoagulated plasma measurements within 2 hours. The type of anticoagulant should be carefully chosen when performing the BNP test if BNP samples were in vitro for a long time.</p>\",\"PeriodicalId\":10384,\"journal\":{\"name\":\"Clinical laboratory\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.7000,\"publicationDate\":\"2024-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical laboratory\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.7754/Clin.Lab.2024.240102\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical laboratory","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.7754/Clin.Lab.2024.240102","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
The Effects of Different Blood Sample and Anticoagulant Types on Quantitative Tests of B-Type Natriuretic Peptide.
Background: BNP is a sensitive and widely used biomarker for an early diagnosis of heart failure. Currently, most commercial BNP detection products use EDTA plasma samples. The aim of this study was to evaluate the clinical performance of the BNP test by using whole blood samples compared to plasma samples, and to evaluate the effect of the anticoagulant type on the BNP test result.
Methods: In total, 106 patients with different BNP levels from the Dahua Hospital volunteered for this study. Clinically homogenous samples, including EDTA anticoagulant plasma, EDTA whole blood, and heparin anticoagulant plasma, were collected and analyzed by using i-Reader S automatic immuno-analyzer and its supporting reagent kits. Pearson's correlation and weighted least squares linear regression analysis, Bland-Altman plotting, and Kappa test were used for statistical analysis.
Results: Correlation analysis showed that BNP concentrations, measured from EDTA anticoagulated plasma samples, had a good linear regression relationship with BNP from whole blood samples, with a slope of 0.9477, r = 0.9978, p < 0.05. A similar correlation was observed between EDTA anticoagulated plasma samples and heparin anticoagulant plasma, with a slope of 0.8413, r = 0.9793, p < 0.05. The BNP concentration measured from the heparin plasma samples were lower than of the EDTA plasma samples. Bland-Altman analysis for assessing BNP concentration agreement showed there was no outlier ratio between EDTA whole blood and EDTA plasma within the range of the detection system, as well as no outlier between EDTA anticoagulated and heparin anticoagulant plasma. Kappa coefficient of BNP concentration between homologous EDTA anticoagulated and heparin anticoagulant plasma was 0.8553 (p < 0.001), and for EDTA anticoagulated plasma and homologous whole blood it was 0.8941 (p < 0.001).
Conclusions: The diagnostic performance of EDTA anticoagulated whole blood samples did not differ significantly from EDTA anticoagulated plasma samples for the BNP test. This study showed no big significant difference between EDTA anticoagulated and heparin anticoagulated plasma measurements within 2 hours. The type of anticoagulant should be carefully chosen when performing the BNP test if BNP samples were in vitro for a long time.
期刊介绍:
Clinical Laboratory is an international fully peer-reviewed journal covering all aspects of laboratory medicine and transfusion medicine. In addition to transfusion medicine topics Clinical Laboratory represents submissions concerning tissue transplantation and hematopoietic, cellular and gene therapies. The journal publishes original articles, review articles, posters, short reports, case studies and letters to the editor dealing with 1) the scientific background, implementation and diagnostic significance of laboratory methods employed in hospitals, blood banks and physicians'' offices and with 2) scientific, administrative and clinical aspects of transfusion medicine and 3) in addition to transfusion medicine topics Clinical Laboratory represents submissions concerning tissue transplantation and hematopoietic, cellular and gene therapies.