J Rueckel, Y Ozpeynirci, C Trumm, C Brem, M Pflaeging, T D Fischer, T Liebig
{"title":"使用衍生2heal栓塞装置治疗颅内动脉瘤的初步结果。","authors":"J Rueckel, Y Ozpeynirci, C Trumm, C Brem, M Pflaeging, T D Fischer, T Liebig","doi":"10.1007/s00234-024-03387-y","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>The Derivo 2 Heal Embolization Device (D2HED) is a novel flow diverter (FD) providing a fibrin-/heparin-based surface coating aiming at lower thrombogenicity. We evaluate periprocedural aspects and preliminary aneurysm occlusion efficacy for intracranial aneurysm treatment.</p><p><strong>Methods: </strong>Thirty-four D2HEDs deployments (34 aneurysms, 32 patients) between 04/2021 and 10/2023 were analyzed. All patients were under dual antiplatelet therapy (dAPT). Periprocedural details, adverse events, and follow-up (FU) imaging were reviewed by consultant-level neuroradiologists. Complication rates and aneurysm occlusion efficacy are compared with performance data of other FDs based on literature research.</p><p><strong>Results: </strong>Each intervention succeeded in the deployment of one D2HED. Significant and/or increased intraaneurysmal contrast stagnation immediately after D2HED deployment was seen in 73.5% of cases according to O'Kelly-Marotta (OKM) grading scale. Clinically relevant early adverse events occurred in three patients: Among them two cases with fusiform aneurysms in the posterior circulation (ischemic events, early in-stent-thrombosis) and one patient (ischemic event) out of the majority of 31 treated internal carotid artery aneurysms (3,2%). Regarding mid-term FU (> 165 days), one aneurysm did not show progressive occlusion presumably caused by a prominent A1 segment arising from the terminal ICA aneurysm itself. Apart from that, mid-term complete / partial occlusion rates of 80% / 20% could be demonstrated.</p><p><strong>Conclusion: </strong>Our case series - although suffering from restricted sample size - suggests a potential effectiveness of D2HED in managing intracranial aneurysms. Further studies with larger samples are warranted to quantify long-term occlusion efficacy and the impact of antithrombogenic surface coating on the necessary (d)APT.</p>","PeriodicalId":2,"journal":{"name":"ACS Applied Bio Materials","volume":null,"pages":null},"PeriodicalIF":4.6000,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11424707/pdf/","citationCount":"0","resultStr":"{\"title\":\"Preliminary results of intracranial aneurysm treatment with derivo2heal embolization device.\",\"authors\":\"J Rueckel, Y Ozpeynirci, C Trumm, C Brem, M Pflaeging, T D Fischer, T Liebig\",\"doi\":\"10.1007/s00234-024-03387-y\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>The Derivo 2 Heal Embolization Device (D2HED) is a novel flow diverter (FD) providing a fibrin-/heparin-based surface coating aiming at lower thrombogenicity. We evaluate periprocedural aspects and preliminary aneurysm occlusion efficacy for intracranial aneurysm treatment.</p><p><strong>Methods: </strong>Thirty-four D2HEDs deployments (34 aneurysms, 32 patients) between 04/2021 and 10/2023 were analyzed. All patients were under dual antiplatelet therapy (dAPT). Periprocedural details, adverse events, and follow-up (FU) imaging were reviewed by consultant-level neuroradiologists. Complication rates and aneurysm occlusion efficacy are compared with performance data of other FDs based on literature research.</p><p><strong>Results: </strong>Each intervention succeeded in the deployment of one D2HED. Significant and/or increased intraaneurysmal contrast stagnation immediately after D2HED deployment was seen in 73.5% of cases according to O'Kelly-Marotta (OKM) grading scale. Clinically relevant early adverse events occurred in three patients: Among them two cases with fusiform aneurysms in the posterior circulation (ischemic events, early in-stent-thrombosis) and one patient (ischemic event) out of the majority of 31 treated internal carotid artery aneurysms (3,2%). Regarding mid-term FU (> 165 days), one aneurysm did not show progressive occlusion presumably caused by a prominent A1 segment arising from the terminal ICA aneurysm itself. Apart from that, mid-term complete / partial occlusion rates of 80% / 20% could be demonstrated.</p><p><strong>Conclusion: </strong>Our case series - although suffering from restricted sample size - suggests a potential effectiveness of D2HED in managing intracranial aneurysms. Further studies with larger samples are warranted to quantify long-term occlusion efficacy and the impact of antithrombogenic surface coating on the necessary (d)APT.</p>\",\"PeriodicalId\":2,\"journal\":{\"name\":\"ACS Applied Bio Materials\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":4.6000,\"publicationDate\":\"2024-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11424707/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"ACS Applied Bio Materials\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1007/s00234-024-03387-y\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/7/1 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"MATERIALS SCIENCE, BIOMATERIALS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Bio Materials","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00234-024-03387-y","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/7/1 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"MATERIALS SCIENCE, BIOMATERIALS","Score":null,"Total":0}
Preliminary results of intracranial aneurysm treatment with derivo2heal embolization device.
Introduction: The Derivo 2 Heal Embolization Device (D2HED) is a novel flow diverter (FD) providing a fibrin-/heparin-based surface coating aiming at lower thrombogenicity. We evaluate periprocedural aspects and preliminary aneurysm occlusion efficacy for intracranial aneurysm treatment.
Methods: Thirty-four D2HEDs deployments (34 aneurysms, 32 patients) between 04/2021 and 10/2023 were analyzed. All patients were under dual antiplatelet therapy (dAPT). Periprocedural details, adverse events, and follow-up (FU) imaging were reviewed by consultant-level neuroradiologists. Complication rates and aneurysm occlusion efficacy are compared with performance data of other FDs based on literature research.
Results: Each intervention succeeded in the deployment of one D2HED. Significant and/or increased intraaneurysmal contrast stagnation immediately after D2HED deployment was seen in 73.5% of cases according to O'Kelly-Marotta (OKM) grading scale. Clinically relevant early adverse events occurred in three patients: Among them two cases with fusiform aneurysms in the posterior circulation (ischemic events, early in-stent-thrombosis) and one patient (ischemic event) out of the majority of 31 treated internal carotid artery aneurysms (3,2%). Regarding mid-term FU (> 165 days), one aneurysm did not show progressive occlusion presumably caused by a prominent A1 segment arising from the terminal ICA aneurysm itself. Apart from that, mid-term complete / partial occlusion rates of 80% / 20% could be demonstrated.
Conclusion: Our case series - although suffering from restricted sample size - suggests a potential effectiveness of D2HED in managing intracranial aneurysms. Further studies with larger samples are warranted to quantify long-term occlusion efficacy and the impact of antithrombogenic surface coating on the necessary (d)APT.