{"title":"用于股骨远端骨折修复的计算优化远皮质锁定钢板与传统植入物的生物力学测试。","authors":"Pawel Brzozowski ,&nbsp;Clayton Inculet ,&nbsp;Emil H. Schemitsch ,&nbsp;Radovan Zdero","doi":"10.1016/j.clinbiomech.2024.106296","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>This study experimentally validated a computationally optimized screw number and screw distribution far cortical locking distal femur fracture plate and compared the results to traditional implants.</p></div><div><h3>Methods</h3><p>24 artificial femurs were osteotomized with a 10 mm fracture gap 60 mm proximal to the intercondylar notch. Three fixation constructs were used. (i) Standard locking plates secured with three far cortical locking screws inserted according to a previously optimized distribution in the femur shaft (<em>n</em> = 8). (ii) Standard locking plates secured with four standard locking screws inserted in alternating plate holes in the femur shaft (n = 8). (iii) Retrograde intramedullary nail secured proximally with one anterior-posterior screw and distally with two oblique screws (n = 8). Axial hip forces (700 and 2800 N) were applied while measuring axial interfragmentary motion, shear interfragmentary motion, and overall stiffness.</p></div><div><h3>Findings</h3><p>Experimental far cortical locking plate results compared well to published computational findings. Far cortical locking femurs contained the highest axial motion within the potential ideal range of 0.2–1 mm and a sheer-to-axial motion ratio &lt; 1.6 at toe-touch weight-bearing (700 N). At full weight-bearing (2800 N), Standard locking-plated femurs had the only axial motion within 0.2–1 mm but had an excess shear-to-axial motion ratio. Nail-implanted femurs underperformed at both forces.</p></div><div><h3>Interpretation</h3><p>For toe-touch weight-bearing, the far cortical locking construct provided optimal biomechanics to allow moderate motion, which has been suggested to encourage early callus formation. Conversely, at full weight-bearing, the standard locking construct offered the biomechanical advantage on fracture motion.</p></div>","PeriodicalId":50992,"journal":{"name":"Clinical Biomechanics","volume":null,"pages":null},"PeriodicalIF":1.4000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Biomechanical testing of a computationally optimized far cortical locking plate versus traditional implants for distal femur fracture repair\",\"authors\":\"Pawel Brzozowski ,&nbsp;Clayton Inculet ,&nbsp;Emil H. Schemitsch ,&nbsp;Radovan Zdero\",\"doi\":\"10.1016/j.clinbiomech.2024.106296\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><p>This study experimentally validated a computationally optimized screw number and screw distribution far cortical locking distal femur fracture plate and compared the results to traditional implants.</p></div><div><h3>Methods</h3><p>24 artificial femurs were osteotomized with a 10 mm fracture gap 60 mm proximal to the intercondylar notch. Three fixation constructs were used. (i) Standard locking plates secured with three far cortical locking screws inserted according to a previously optimized distribution in the femur shaft (<em>n</em> = 8). (ii) Standard locking plates secured with four standard locking screws inserted in alternating plate holes in the femur shaft (n = 8). (iii) Retrograde intramedullary nail secured proximally with one anterior-posterior screw and distally with two oblique screws (n = 8). Axial hip forces (700 and 2800 N) were applied while measuring axial interfragmentary motion, shear interfragmentary motion, and overall stiffness.</p></div><div><h3>Findings</h3><p>Experimental far cortical locking plate results compared well to published computational findings. Far cortical locking femurs contained the highest axial motion within the potential ideal range of 0.2–1 mm and a sheer-to-axial motion ratio &lt; 1.6 at toe-touch weight-bearing (700 N). At full weight-bearing (2800 N), Standard locking-plated femurs had the only axial motion within 0.2–1 mm but had an excess shear-to-axial motion ratio. Nail-implanted femurs underperformed at both forces.</p></div><div><h3>Interpretation</h3><p>For toe-touch weight-bearing, the far cortical locking construct provided optimal biomechanics to allow moderate motion, which has been suggested to encourage early callus formation. Conversely, at full weight-bearing, the standard locking construct offered the biomechanical advantage on fracture motion.</p></div>\",\"PeriodicalId\":50992,\"journal\":{\"name\":\"Clinical Biomechanics\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.4000,\"publicationDate\":\"2024-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical Biomechanics\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0268003324001281\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical Biomechanics","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0268003324001281","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

摘要

背景:本研究通过实验验证了经过计算优化的螺钉数量和螺钉分布远皮质锁定股骨远端骨折钢板,并将结果与传统植入物进行了比较。使用了三种固定结构。(i) 根据先前优化的股骨轴分布,用三根远端皮质锁定螺钉固定标准锁定钢板(n = 8)。(ii) 在股骨轴上交替插入四颗标准锁定螺钉,固定标准锁定钢板(n = 8)。(iii) 逆行髓内钉,近端用一颗前后螺钉固定,远端用两颗斜螺钉固定(n = 8)。在测量轴向节间运动、剪切节间运动和整体刚度的同时,施加轴向髋关节力(700 和 2800 N):实验结果:远皮质锁定钢板的实验结果与已发表的计算结果比较吻合。在 0.2-1 毫米的潜在理想范围内,远端皮质锁定股骨的轴向运动最大,剪切运动与轴向运动的比率也最大:在脚趾接触负重的情况下,远皮质锁定结构提供了最佳的生物力学,允许适度的运动,这被认为能促进早期胼胝的形成。相反,在完全负重时,标准锁定结构在骨折运动方面具有生物力学优势。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Biomechanical testing of a computationally optimized far cortical locking plate versus traditional implants for distal femur fracture repair

Background

This study experimentally validated a computationally optimized screw number and screw distribution far cortical locking distal femur fracture plate and compared the results to traditional implants.

Methods

24 artificial femurs were osteotomized with a 10 mm fracture gap 60 mm proximal to the intercondylar notch. Three fixation constructs were used. (i) Standard locking plates secured with three far cortical locking screws inserted according to a previously optimized distribution in the femur shaft (n = 8). (ii) Standard locking plates secured with four standard locking screws inserted in alternating plate holes in the femur shaft (n = 8). (iii) Retrograde intramedullary nail secured proximally with one anterior-posterior screw and distally with two oblique screws (n = 8). Axial hip forces (700 and 2800 N) were applied while measuring axial interfragmentary motion, shear interfragmentary motion, and overall stiffness.

Findings

Experimental far cortical locking plate results compared well to published computational findings. Far cortical locking femurs contained the highest axial motion within the potential ideal range of 0.2–1 mm and a sheer-to-axial motion ratio < 1.6 at toe-touch weight-bearing (700 N). At full weight-bearing (2800 N), Standard locking-plated femurs had the only axial motion within 0.2–1 mm but had an excess shear-to-axial motion ratio. Nail-implanted femurs underperformed at both forces.

Interpretation

For toe-touch weight-bearing, the far cortical locking construct provided optimal biomechanics to allow moderate motion, which has been suggested to encourage early callus formation. Conversely, at full weight-bearing, the standard locking construct offered the biomechanical advantage on fracture motion.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Clinical Biomechanics
Clinical Biomechanics 医学-工程:生物医学
CiteScore
3.30
自引率
5.60%
发文量
189
审稿时长
12.3 weeks
期刊介绍: Clinical Biomechanics is an international multidisciplinary journal of biomechanics with a focus on medical and clinical applications of new knowledge in the field. The science of biomechanics helps explain the causes of cell, tissue, organ and body system disorders, and supports clinicians in the diagnosis, prognosis and evaluation of treatment methods and technologies. Clinical Biomechanics aims to strengthen the links between laboratory and clinic by publishing cutting-edge biomechanics research which helps to explain the causes of injury and disease, and which provides evidence contributing to improved clinical management. A rigorous peer review system is employed and every attempt is made to process and publish top-quality papers promptly. Clinical Biomechanics explores all facets of body system, organ, tissue and cell biomechanics, with an emphasis on medical and clinical applications of the basic science aspects. The role of basic science is therefore recognized in a medical or clinical context. The readership of the journal closely reflects its multi-disciplinary contents, being a balance of scientists, engineers and clinicians. The contents are in the form of research papers, brief reports, review papers and correspondence, whilst special interest issues and supplements are published from time to time. Disciplines covered include biomechanics and mechanobiology at all scales, bioengineering and use of tissue engineering and biomaterials for clinical applications, biophysics, as well as biomechanical aspects of medical robotics, ergonomics, physical and occupational therapeutics and rehabilitation.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信