{"title":"模拟器验证对幼儿佩戴软带骨导助听器有潜在益处。","authors":"Le Sun, Lu Ping, Xinmiao Fan, Jian Wang, Xiaowei Chen","doi":"10.1097/MAO.0000000000004245","DOIUrl":null,"url":null,"abstract":"<p><strong>Hypothesis: </strong>The current study employed a skull-simulator verification method to assess whether the output of softband bone conduction hearing devices (BCHDs) at the manufacturer's default settings deviated widely from the target determined by the fitting formula.</p><p><strong>Background: </strong>Real ear analysis is utilized for the verification of the fitting of air conduction hearing devices (ACHDs) in a variety of institutions. This procedure, however, has not been used in the fitting of BCHDs, largely due to the difficulty of testing the output of these devices to temporal bones. Despite the availability of skull simulators, they have not been utilized clinically to measure BCHD output.</p><p><strong>Materials and methods: </strong>This prospective, single-center study enrolled 42 subjects, aged 3 months to 10 years, with microtia-atresia-associated mild-to-severe bilateral conductive hearing loss. Hearing sensitivity was evaluated behaviorally by pure tone audiometry (PTA) in 22 subjects 4 years or older (the PTA group), and by auditory brainstem response (ABR) in 20 subjects younger than 4 years (the ABR group). Following 6 months of subjects wearing the prescribed softband BCHDs, their dial level (DL) thresholds were reassessed while using their own BCHDs, configured with zero gain across all frequencies, functioning solely as a bone vibrator. These DL thresholds were inputted into the fitting formula, desired sensation level-bone conduction devices (DSL-BCD) for children, to obtain the target values of BCHD output. The simulator output of the BCHD programmed at the manufacturer's default setting was measured in response to speech presented at 55, 65, and 80 dB SPL, followed by gain adjustment based on the differences between the simulator output and the target. Aided speech intelligibility index (SII) was measured before and after the gain adjustment.</p><p><strong>Results: </strong>The softband BCHDs at the manufacturer's settings generally had lower output than the prescribed target values. This difference was larger at low frequencies and low levels. Across the 12 points tested (four frequencies from 500 to 4000 Hz multiplied by three levels), 22 (52.3%) and 42 (100%) BCHDs had deviations of +7 and +5 dB, respectively, at one point or more. The gain adjustments reduced the deviation and improved the SII values at the two lower levels of speech presented.</p><p><strong>Conclusion: </strong>The simulator output of softband bone conduction hearing devices (BCHDs) with the manufacturer's settings may exhibit significant deviations from the formula. Objective output verification should be considered a beneficial step in BCHD fitting and is recommended when applicable.</p>","PeriodicalId":1,"journal":{"name":"Accounts of Chemical Research","volume":null,"pages":null},"PeriodicalIF":16.4000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Simulator Verification Is Potentially Beneficial for the Fitting of Softband Bone Conduction Hearing Devices in Young Children.\",\"authors\":\"Le Sun, Lu Ping, Xinmiao Fan, Jian Wang, Xiaowei Chen\",\"doi\":\"10.1097/MAO.0000000000004245\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Hypothesis: </strong>The current study employed a skull-simulator verification method to assess whether the output of softband bone conduction hearing devices (BCHDs) at the manufacturer's default settings deviated widely from the target determined by the fitting formula.</p><p><strong>Background: </strong>Real ear analysis is utilized for the verification of the fitting of air conduction hearing devices (ACHDs) in a variety of institutions. This procedure, however, has not been used in the fitting of BCHDs, largely due to the difficulty of testing the output of these devices to temporal bones. Despite the availability of skull simulators, they have not been utilized clinically to measure BCHD output.</p><p><strong>Materials and methods: </strong>This prospective, single-center study enrolled 42 subjects, aged 3 months to 10 years, with microtia-atresia-associated mild-to-severe bilateral conductive hearing loss. Hearing sensitivity was evaluated behaviorally by pure tone audiometry (PTA) in 22 subjects 4 years or older (the PTA group), and by auditory brainstem response (ABR) in 20 subjects younger than 4 years (the ABR group). Following 6 months of subjects wearing the prescribed softband BCHDs, their dial level (DL) thresholds were reassessed while using their own BCHDs, configured with zero gain across all frequencies, functioning solely as a bone vibrator. These DL thresholds were inputted into the fitting formula, desired sensation level-bone conduction devices (DSL-BCD) for children, to obtain the target values of BCHD output. The simulator output of the BCHD programmed at the manufacturer's default setting was measured in response to speech presented at 55, 65, and 80 dB SPL, followed by gain adjustment based on the differences between the simulator output and the target. Aided speech intelligibility index (SII) was measured before and after the gain adjustment.</p><p><strong>Results: </strong>The softband BCHDs at the manufacturer's settings generally had lower output than the prescribed target values. This difference was larger at low frequencies and low levels. Across the 12 points tested (four frequencies from 500 to 4000 Hz multiplied by three levels), 22 (52.3%) and 42 (100%) BCHDs had deviations of +7 and +5 dB, respectively, at one point or more. The gain adjustments reduced the deviation and improved the SII values at the two lower levels of speech presented.</p><p><strong>Conclusion: </strong>The simulator output of softband bone conduction hearing devices (BCHDs) with the manufacturer's settings may exhibit significant deviations from the formula. Objective output verification should be considered a beneficial step in BCHD fitting and is recommended when applicable.</p>\",\"PeriodicalId\":1,\"journal\":{\"name\":\"Accounts of Chemical Research\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":16.4000,\"publicationDate\":\"2024-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Accounts of Chemical Research\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MAO.0000000000004245\",\"RegionNum\":1,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/6/26 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q1\",\"JCRName\":\"CHEMISTRY, MULTIDISCIPLINARY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Accounts of Chemical Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MAO.0000000000004245","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/6/26 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
Simulator Verification Is Potentially Beneficial for the Fitting of Softband Bone Conduction Hearing Devices in Young Children.
Hypothesis: The current study employed a skull-simulator verification method to assess whether the output of softband bone conduction hearing devices (BCHDs) at the manufacturer's default settings deviated widely from the target determined by the fitting formula.
Background: Real ear analysis is utilized for the verification of the fitting of air conduction hearing devices (ACHDs) in a variety of institutions. This procedure, however, has not been used in the fitting of BCHDs, largely due to the difficulty of testing the output of these devices to temporal bones. Despite the availability of skull simulators, they have not been utilized clinically to measure BCHD output.
Materials and methods: This prospective, single-center study enrolled 42 subjects, aged 3 months to 10 years, with microtia-atresia-associated mild-to-severe bilateral conductive hearing loss. Hearing sensitivity was evaluated behaviorally by pure tone audiometry (PTA) in 22 subjects 4 years or older (the PTA group), and by auditory brainstem response (ABR) in 20 subjects younger than 4 years (the ABR group). Following 6 months of subjects wearing the prescribed softband BCHDs, their dial level (DL) thresholds were reassessed while using their own BCHDs, configured with zero gain across all frequencies, functioning solely as a bone vibrator. These DL thresholds were inputted into the fitting formula, desired sensation level-bone conduction devices (DSL-BCD) for children, to obtain the target values of BCHD output. The simulator output of the BCHD programmed at the manufacturer's default setting was measured in response to speech presented at 55, 65, and 80 dB SPL, followed by gain adjustment based on the differences between the simulator output and the target. Aided speech intelligibility index (SII) was measured before and after the gain adjustment.
Results: The softband BCHDs at the manufacturer's settings generally had lower output than the prescribed target values. This difference was larger at low frequencies and low levels. Across the 12 points tested (four frequencies from 500 to 4000 Hz multiplied by three levels), 22 (52.3%) and 42 (100%) BCHDs had deviations of +7 and +5 dB, respectively, at one point or more. The gain adjustments reduced the deviation and improved the SII values at the two lower levels of speech presented.
Conclusion: The simulator output of softband bone conduction hearing devices (BCHDs) with the manufacturer's settings may exhibit significant deviations from the formula. Objective output verification should be considered a beneficial step in BCHD fitting and is recommended when applicable.
期刊介绍:
Accounts of Chemical Research presents short, concise and critical articles offering easy-to-read overviews of basic research and applications in all areas of chemistry and biochemistry. These short reviews focus on research from the author’s own laboratory and are designed to teach the reader about a research project. In addition, Accounts of Chemical Research publishes commentaries that give an informed opinion on a current research problem. Special Issues online are devoted to a single topic of unusual activity and significance.
Accounts of Chemical Research replaces the traditional article abstract with an article "Conspectus." These entries synopsize the research affording the reader a closer look at the content and significance of an article. Through this provision of a more detailed description of the article contents, the Conspectus enhances the article's discoverability by search engines and the exposure for the research.