{"title":"利用负重 CT(WBCT)获取双侧足、踝、膝和髋关节三维成像的方案","authors":"N.A. Segal ,&nbsp;J.A. Lynch","doi":"10.1016/j.ostima.2024.100200","DOIUrl":null,"url":null,"abstract":"<div><h3>INTRODUCTION</h3><p>MRI and CT can provide valuable 3D information on joint structures but are generally acquired in a non-functional position. The Multicenter Osteoarthritis Study (MOST) will address this serious limitation through using low-dose weight-bearing CT (WBCT) to obtain 3D images of the bilateral hips, knees, ankles/feet in a standing position, while also maintaining continuity with previous MOST exams by acquiring both 3D WBCT and knee radiographs using the Synaflexer frame used for all previous MOST knee x-rays. This imaging protocol takes about the same time as PA and lateral knee x-rays, while providing a wealth of additional information that will enable future opportunities to investigate novel questions about OA using 3D imaging of all the joints in the lower limb kinetic chain in a functionally loaded position.</p></div><div><h3>OBJECTIVE</h3><p>To standardize the 3D lower limb joints (hips, knees, feet/ankles) imaging procedures using WBCT and knee radiographs using the Planmed XFI scanner.</p></div><div><h3>METHODS</h3><p>The XFI permits efficient scans (46sec per scan for the hips, knees, and feet/ankles respectively) with excellent image quality (least artifact among WBCT scanners and resolution 150–300µm), with effective algorithms for correction of participant motion and metal artifacts, thereby minimizing the need to repeat scans. The gantry diameter of 85cm permits larger bodies or people with knee flexion contractures to be scanned and has a 43.2 × 43.2cm flat panel detector with pixel size 148µm and source distance of 80cm, providing superior imaging capabilities. Average effective radiation doses are acceptable (20µSv, 26.5µSv, and 629µSv for 3D WBCT of the feet/ankles, knees, and hips and 7µSv total for bilateral knee and hand radiographs), automatic exposure control (AEC) optimizes image quality vs. radiation dose.</p></div><div><h3>RESULTS</h3><p>Participants stand on a platform facing the vertical table, with their feet pressed against the Perspex Synaflexor frame with aluminum beads to position their feet in 5 degrees external rotation (10° between medial sides of feet) and their great toes at the edge of the positioner closest to the table. They stand with equal weight on each foot and with their patellae and front of their thighs pressed against the vertical table, holding handholds and with a Velcro strap around the table and their thighs for safety during the standing scan. The XFI and participant position is adjusted so that: the coronal laser passes through the greater trochanters, posterior to the knees and through the midfoot; the sagittal laser passes in the center between the lower limbs and the axial laser is positioned for each joint of interest. For an average size adult, 3D images of the hip are acquired at 120 kV, 450 mAs, bowtie filter on and using AEC, and knees, feet/ankles at 96 kV, 180 mAs. To scan the hips, the gantry is lowered until the axial laser is at the level 2 inches above the greater trochanters. Scout images are used to confirm inclusion of the hip region of interest, collimate the exposure and set the AEC. Following the hip scan, the gantry is lowered until the axial laser passes through the popliteal fossa crease. Following knee image acquisition, the gantry is lowered until the axial laser passes through the fifth metatarsal and the foot/ankle 3D image acquisition is completed.</p><p>Knee radiographs are acquired using 2D imaging protocols. The gantry is rotated to allow the participant to be positioned against an 80° angled Synaflexor frame placed against the x-ray detector. This frame was modified from the 90° Synaflexor frame to account for the horizontal beam angle of the XFI. For knees that would require a 10° beam angle to optimize imaging of the medial tibial plateau (MTP) on fixed-flexion x-ray, no additional adjustment is necessary. For knees that require a 5° or 15° x-ray beam angles, a 5° wedge is placed under the base of the modified Synaflexor frame to achieve those equivalent relative angles between the beam and the MTP when acquiring the standing bilateral fixed-flexion PA radiograph (Figure). Following this, lateral view radiographs of the left and then the right knee are acquired. The participant's leg of interest is positioned next to and parallel to the detector, with the tip of the great toe in contact a vertical Perspex sheet and the contralateral toes are level with the back of the heel, resulting in a 40-50°flexion angle of the knee being radiographed.</p></div><div><h3>CONCLUSIONS</h3><p>MOST4 will deliver a wealth of 3D weight-bearing imaging data for the bilateral hips, knees, feet, and ankles, using low-dose volumetric cone-beam WBCT technology at an acceptably low radiation dose.</p></div>","PeriodicalId":74378,"journal":{"name":"Osteoarthritis imaging","volume":"4 ","pages":"Article 100200"},"PeriodicalIF":0.0000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S277265412400028X/pdfft?md5=b1164690ab7719afdf46cd5261234fc0&pid=1-s2.0-S277265412400028X-main.pdf","citationCount":"0","resultStr":"{\"title\":\"PROTOCOL FOR ACQUISITION OF 3D IMAGING OF BILATERAL FOOT, ANKLE, KNEE, AND HIP JOINTS WITH WEIGHT-BEARING CT (WBCT)\",\"authors\":\"N.A. Segal ,&nbsp;J.A. Lynch\",\"doi\":\"10.1016/j.ostima.2024.100200\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>INTRODUCTION</h3><p>MRI and CT can provide valuable 3D information on joint structures but are generally acquired in a non-functional position. The Multicenter Osteoarthritis Study (MOST) will address this serious limitation through using low-dose weight-bearing CT (WBCT) to obtain 3D images of the bilateral hips, knees, ankles/feet in a standing position, while also maintaining continuity with previous MOST exams by acquiring both 3D WBCT and knee radiographs using the Synaflexer frame used for all previous MOST knee x-rays. This imaging protocol takes about the same time as PA and lateral knee x-rays, while providing a wealth of additional information that will enable future opportunities to investigate novel questions about OA using 3D imaging of all the joints in the lower limb kinetic chain in a functionally loaded position.</p></div><div><h3>OBJECTIVE</h3><p>To standardize the 3D lower limb joints (hips, knees, feet/ankles) imaging procedures using WBCT and knee radiographs using the Planmed XFI scanner.</p></div><div><h3>METHODS</h3><p>The XFI permits efficient scans (46sec per scan for the hips, knees, and feet/ankles respectively) with excellent image quality (least artifact among WBCT scanners and resolution 150–300µm), with effective algorithms for correction of participant motion and metal artifacts, thereby minimizing the need to repeat scans. The gantry diameter of 85cm permits larger bodies or people with knee flexion contractures to be scanned and has a 43.2 × 43.2cm flat panel detector with pixel size 148µm and source distance of 80cm, providing superior imaging capabilities. Average effective radiation doses are acceptable (20µSv, 26.5µSv, and 629µSv for 3D WBCT of the feet/ankles, knees, and hips and 7µSv total for bilateral knee and hand radiographs), automatic exposure control (AEC) optimizes image quality vs. radiation dose.</p></div><div><h3>RESULTS</h3><p>Participants stand on a platform facing the vertical table, with their feet pressed against the Perspex Synaflexor frame with aluminum beads to position their feet in 5 degrees external rotation (10° between medial sides of feet) and their great toes at the edge of the positioner closest to the table. They stand with equal weight on each foot and with their patellae and front of their thighs pressed against the vertical table, holding handholds and with a Velcro strap around the table and their thighs for safety during the standing scan. The XFI and participant position is adjusted so that: the coronal laser passes through the greater trochanters, posterior to the knees and through the midfoot; the sagittal laser passes in the center between the lower limbs and the axial laser is positioned for each joint of interest. For an average size adult, 3D images of the hip are acquired at 120 kV, 450 mAs, bowtie filter on and using AEC, and knees, feet/ankles at 96 kV, 180 mAs. To scan the hips, the gantry is lowered until the axial laser is at the level 2 inches above the greater trochanters. Scout images are used to confirm inclusion of the hip region of interest, collimate the exposure and set the AEC. Following the hip scan, the gantry is lowered until the axial laser passes through the popliteal fossa crease. Following knee image acquisition, the gantry is lowered until the axial laser passes through the fifth metatarsal and the foot/ankle 3D image acquisition is completed.</p><p>Knee radiographs are acquired using 2D imaging protocols. The gantry is rotated to allow the participant to be positioned against an 80° angled Synaflexor frame placed against the x-ray detector. This frame was modified from the 90° Synaflexor frame to account for the horizontal beam angle of the XFI. For knees that would require a 10° beam angle to optimize imaging of the medial tibial plateau (MTP) on fixed-flexion x-ray, no additional adjustment is necessary. For knees that require a 5° or 15° x-ray beam angles, a 5° wedge is placed under the base of the modified Synaflexor frame to achieve those equivalent relative angles between the beam and the MTP when acquiring the standing bilateral fixed-flexion PA radiograph (Figure). Following this, lateral view radiographs of the left and then the right knee are acquired. The participant's leg of interest is positioned next to and parallel to the detector, with the tip of the great toe in contact a vertical Perspex sheet and the contralateral toes are level with the back of the heel, resulting in a 40-50°flexion angle of the knee being radiographed.</p></div><div><h3>CONCLUSIONS</h3><p>MOST4 will deliver a wealth of 3D weight-bearing imaging data for the bilateral hips, knees, feet, and ankles, using low-dose volumetric cone-beam WBCT technology at an acceptably low radiation dose.</p></div>\",\"PeriodicalId\":74378,\"journal\":{\"name\":\"Osteoarthritis imaging\",\"volume\":\"4 \",\"pages\":\"Article 100200\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S277265412400028X/pdfft?md5=b1164690ab7719afdf46cd5261234fc0&pid=1-s2.0-S277265412400028X-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Osteoarthritis imaging\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S277265412400028X\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Osteoarthritis imaging","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S277265412400028X","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

引言MRI 和 CT 可以提供关节结构的宝贵三维信息,但通常是在非功能性体位下获取的。多中心骨关节炎研究(MOST)将通过使用低剂量负重 CT(WBCT)获取双侧髋关节、膝关节、踝关节/足部的站立位三维图像来解决这一严重的局限性,同时还将通过使用以往所有 MOST 膝关节 X 光片所使用的 Synaflexer 架获取三维 WBCT 和膝关节 X 光片来保持与以往 MOST 检查的连续性。该成像方案所需的时间与 PA 和膝关节侧位 X 光片差不多,但却能提供大量的额外信息,从而为将来利用功能负荷位置下肢运动链中所有关节的三维成像研究有关 OA 的新问题提供机会。方法XFI可进行高效扫描(髋关节、膝关节和足/踝关节的每次扫描时间分别为46秒),图像质量极佳(在WBCT扫描仪中伪影最少,分辨率为150-300µm),并采用有效算法对参与者的运动和金属伪影进行校正,从而最大限度地减少重复扫描的需要。该扫描仪的龙门直径为 85 厘米,可对较大的身体或膝关节屈曲挛缩者进行扫描,其平板探测器尺寸为 43.2 × 43.2 厘米,像素大小为 148 微米,光源距离为 80 厘米,具有卓越的成像能力。平均有效辐射剂量是可接受的(脚/踝、膝和髋关节的 3D WBCT 分别为 20µSv、26.5µSv 和 629µSv,双侧膝关节和手部射线照片的总剂量为 7µSv),自动曝光控制 (AEC) 优化了图像质量与辐射剂量的关系。结果参试者站在一个平台上,面向垂直的工作台,双脚紧贴带有铝珠的 Perspex Synaflexor 框架,双脚外旋 5 度(双脚内侧之间为 10°),大脚趾位于最靠近工作台的定位器边缘。在站立扫描过程中,他们每只脚的重量相等,髌骨和大腿前部紧贴垂直台面,手持支架,台面和大腿上有尼龙搭扣带,以确保安全。调整 XFI 和参与者的位置,使冠状激光穿过大转子、膝盖后方和中足;矢状激光穿过下肢之间的中心;轴向激光定位到每个感兴趣的关节。对于普通体型的成年人,髋关节的三维图像是在 120 kV、450 mAs、弓形滤波器开启并使用 AEC 的情况下采集的,膝关节、脚/踝关节的三维图像是在 96 kV、180 mAs 的情况下采集的。扫描髋部时,将龙门架降低,直到轴向激光位于大转子上方 2 英寸处。扫描图像用于确认是否包含髋部感兴趣区、准直曝光和设置 AEC。完成髋部扫描后,降低龙门架,直到轴向激光穿过腘窝皱襞。膝关节图像采集完成后,降低龙门架,直到轴向激光穿过第五跖骨,脚/踝关节三维图像采集完成。旋转龙门架,使受试者靠在与 X 射线探测器成 80° 角的 Synaflexor 框架上。该框架由 90° Synaflexor 框架改装而成,以考虑到 XFI 的水平光束角度。对于需要 10° 光束角以优化固定屈曲 X 光中胫骨内侧平台 (MTP) 成像的膝关节,无需进行额外调整。对于需要 5° 或 15° X 光束角的膝关节,在获取站立双侧固定屈曲 PA X 光片时,在改良型 Synaflexor 架的底部放置一个 5° 楔形,以实现光束与 MTP 之间的等效相对角度(图)。随后,分别采集左膝和右膝的侧视放射线照片。参试者感兴趣的腿放置在探测器旁边并与之平行,大脚趾尖与垂直的 Perspex 片接触,对侧脚趾与脚跟后部持平,从而使膝关节的屈曲角度达到 40-50 度。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
PROTOCOL FOR ACQUISITION OF 3D IMAGING OF BILATERAL FOOT, ANKLE, KNEE, AND HIP JOINTS WITH WEIGHT-BEARING CT (WBCT)

INTRODUCTION

MRI and CT can provide valuable 3D information on joint structures but are generally acquired in a non-functional position. The Multicenter Osteoarthritis Study (MOST) will address this serious limitation through using low-dose weight-bearing CT (WBCT) to obtain 3D images of the bilateral hips, knees, ankles/feet in a standing position, while also maintaining continuity with previous MOST exams by acquiring both 3D WBCT and knee radiographs using the Synaflexer frame used for all previous MOST knee x-rays. This imaging protocol takes about the same time as PA and lateral knee x-rays, while providing a wealth of additional information that will enable future opportunities to investigate novel questions about OA using 3D imaging of all the joints in the lower limb kinetic chain in a functionally loaded position.

OBJECTIVE

To standardize the 3D lower limb joints (hips, knees, feet/ankles) imaging procedures using WBCT and knee radiographs using the Planmed XFI scanner.

METHODS

The XFI permits efficient scans (46sec per scan for the hips, knees, and feet/ankles respectively) with excellent image quality (least artifact among WBCT scanners and resolution 150–300µm), with effective algorithms for correction of participant motion and metal artifacts, thereby minimizing the need to repeat scans. The gantry diameter of 85cm permits larger bodies or people with knee flexion contractures to be scanned and has a 43.2 × 43.2cm flat panel detector with pixel size 148µm and source distance of 80cm, providing superior imaging capabilities. Average effective radiation doses are acceptable (20µSv, 26.5µSv, and 629µSv for 3D WBCT of the feet/ankles, knees, and hips and 7µSv total for bilateral knee and hand radiographs), automatic exposure control (AEC) optimizes image quality vs. radiation dose.

RESULTS

Participants stand on a platform facing the vertical table, with their feet pressed against the Perspex Synaflexor frame with aluminum beads to position their feet in 5 degrees external rotation (10° between medial sides of feet) and their great toes at the edge of the positioner closest to the table. They stand with equal weight on each foot and with their patellae and front of their thighs pressed against the vertical table, holding handholds and with a Velcro strap around the table and their thighs for safety during the standing scan. The XFI and participant position is adjusted so that: the coronal laser passes through the greater trochanters, posterior to the knees and through the midfoot; the sagittal laser passes in the center between the lower limbs and the axial laser is positioned for each joint of interest. For an average size adult, 3D images of the hip are acquired at 120 kV, 450 mAs, bowtie filter on and using AEC, and knees, feet/ankles at 96 kV, 180 mAs. To scan the hips, the gantry is lowered until the axial laser is at the level 2 inches above the greater trochanters. Scout images are used to confirm inclusion of the hip region of interest, collimate the exposure and set the AEC. Following the hip scan, the gantry is lowered until the axial laser passes through the popliteal fossa crease. Following knee image acquisition, the gantry is lowered until the axial laser passes through the fifth metatarsal and the foot/ankle 3D image acquisition is completed.

Knee radiographs are acquired using 2D imaging protocols. The gantry is rotated to allow the participant to be positioned against an 80° angled Synaflexor frame placed against the x-ray detector. This frame was modified from the 90° Synaflexor frame to account for the horizontal beam angle of the XFI. For knees that would require a 10° beam angle to optimize imaging of the medial tibial plateau (MTP) on fixed-flexion x-ray, no additional adjustment is necessary. For knees that require a 5° or 15° x-ray beam angles, a 5° wedge is placed under the base of the modified Synaflexor frame to achieve those equivalent relative angles between the beam and the MTP when acquiring the standing bilateral fixed-flexion PA radiograph (Figure). Following this, lateral view radiographs of the left and then the right knee are acquired. The participant's leg of interest is positioned next to and parallel to the detector, with the tip of the great toe in contact a vertical Perspex sheet and the contralateral toes are level with the back of the heel, resulting in a 40-50°flexion angle of the knee being radiographed.

CONCLUSIONS

MOST4 will deliver a wealth of 3D weight-bearing imaging data for the bilateral hips, knees, feet, and ankles, using low-dose volumetric cone-beam WBCT technology at an acceptably low radiation dose.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Osteoarthritis imaging
Osteoarthritis imaging Radiology and Imaging
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信