{"title":"通过 Von Clauss 方法的价值转移改进凝血酶原时间衍生纤维蛋白原测定法的应用。","authors":"Yan Zhang, Xiaoyong Duan, Xiao Li, Yunfeng Ling, Fan Wu, Sheng Wang, Xingtai Liu, Ying Chen, Hongbin Tong, Dongmei Li","doi":"10.7754/Clin.Lab.2024.231026","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>This study aimed to improve the accuracy of the fibrinogen (Fib) prothrombin time-derived (PT-der) method. To achieve this, a value transfer method was introduced for calibration, and its effectiveness was assessed.</p><p><strong>Methods: </strong>The PT-der Fib assay was calibrated by pooled samples (assigned by the von Clauss method) in three different ways: 1) multipoint calibration using an automatic dilution system, 2) multipoint calibration using a manual dilution method, and 3) manual calibration with multiple concentrations. Three calibration equations (1, 2, and 3) were obtained and an optimal equation was selected by comparing the detection results of the von Clauss method with the PT-der method. Subsequently, the optimal equation was assessed for an accuracy limit, and linear analysis and reference interval verification were performed following the guidelines (EP15-A and EP6-A) issued by the CLSI.</p><p><strong>Results: </strong>Compared with the other two equations (equation 1 and 2), equation 3, available from manual calibration with multiple concentrations, showed a better performance for the PT-der determination in a primary cohort (n = 208), and a good agreement (99% of the results between 1.52 and 6.30 g/L were interchangeable) was validated (n = 3226). The reference interval was also verified in almost all healthy individuals (39/40). However, the discrep-ancy between the two methods was observed in several specific conditions, such as hyperfibrinolysis.</p><p><strong>Conclusions: </strong>Manual calibration with multiple concentrations is better for the Fib PT-der method assay. As a rapid, accurate, and economical test, the performance of the Fib PT-der method has been verified and may be more applicable than before.</p>","PeriodicalId":10384,"journal":{"name":"Clinical laboratory","volume":null,"pages":null},"PeriodicalIF":0.7000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Improving the Application of Prothrombin Time-Derived Fibrinogen Assay Through Value Transfer from the Von Clauss Method.\",\"authors\":\"Yan Zhang, Xiaoyong Duan, Xiao Li, Yunfeng Ling, Fan Wu, Sheng Wang, Xingtai Liu, Ying Chen, Hongbin Tong, Dongmei Li\",\"doi\":\"10.7754/Clin.Lab.2024.231026\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>This study aimed to improve the accuracy of the fibrinogen (Fib) prothrombin time-derived (PT-der) method. To achieve this, a value transfer method was introduced for calibration, and its effectiveness was assessed.</p><p><strong>Methods: </strong>The PT-der Fib assay was calibrated by pooled samples (assigned by the von Clauss method) in three different ways: 1) multipoint calibration using an automatic dilution system, 2) multipoint calibration using a manual dilution method, and 3) manual calibration with multiple concentrations. Three calibration equations (1, 2, and 3) were obtained and an optimal equation was selected by comparing the detection results of the von Clauss method with the PT-der method. Subsequently, the optimal equation was assessed for an accuracy limit, and linear analysis and reference interval verification were performed following the guidelines (EP15-A and EP6-A) issued by the CLSI.</p><p><strong>Results: </strong>Compared with the other two equations (equation 1 and 2), equation 3, available from manual calibration with multiple concentrations, showed a better performance for the PT-der determination in a primary cohort (n = 208), and a good agreement (99% of the results between 1.52 and 6.30 g/L were interchangeable) was validated (n = 3226). The reference interval was also verified in almost all healthy individuals (39/40). However, the discrep-ancy between the two methods was observed in several specific conditions, such as hyperfibrinolysis.</p><p><strong>Conclusions: </strong>Manual calibration with multiple concentrations is better for the Fib PT-der method assay. As a rapid, accurate, and economical test, the performance of the Fib PT-der method has been verified and may be more applicable than before.</p>\",\"PeriodicalId\":10384,\"journal\":{\"name\":\"Clinical laboratory\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.7000,\"publicationDate\":\"2024-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical laboratory\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.7754/Clin.Lab.2024.231026\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical laboratory","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.7754/Clin.Lab.2024.231026","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
Improving the Application of Prothrombin Time-Derived Fibrinogen Assay Through Value Transfer from the Von Clauss Method.
Background: This study aimed to improve the accuracy of the fibrinogen (Fib) prothrombin time-derived (PT-der) method. To achieve this, a value transfer method was introduced for calibration, and its effectiveness was assessed.
Methods: The PT-der Fib assay was calibrated by pooled samples (assigned by the von Clauss method) in three different ways: 1) multipoint calibration using an automatic dilution system, 2) multipoint calibration using a manual dilution method, and 3) manual calibration with multiple concentrations. Three calibration equations (1, 2, and 3) were obtained and an optimal equation was selected by comparing the detection results of the von Clauss method with the PT-der method. Subsequently, the optimal equation was assessed for an accuracy limit, and linear analysis and reference interval verification were performed following the guidelines (EP15-A and EP6-A) issued by the CLSI.
Results: Compared with the other two equations (equation 1 and 2), equation 3, available from manual calibration with multiple concentrations, showed a better performance for the PT-der determination in a primary cohort (n = 208), and a good agreement (99% of the results between 1.52 and 6.30 g/L were interchangeable) was validated (n = 3226). The reference interval was also verified in almost all healthy individuals (39/40). However, the discrep-ancy between the two methods was observed in several specific conditions, such as hyperfibrinolysis.
Conclusions: Manual calibration with multiple concentrations is better for the Fib PT-der method assay. As a rapid, accurate, and economical test, the performance of the Fib PT-der method has been verified and may be more applicable than before.
期刊介绍:
Clinical Laboratory is an international fully peer-reviewed journal covering all aspects of laboratory medicine and transfusion medicine. In addition to transfusion medicine topics Clinical Laboratory represents submissions concerning tissue transplantation and hematopoietic, cellular and gene therapies. The journal publishes original articles, review articles, posters, short reports, case studies and letters to the editor dealing with 1) the scientific background, implementation and diagnostic significance of laboratory methods employed in hospitals, blood banks and physicians'' offices and with 2) scientific, administrative and clinical aspects of transfusion medicine and 3) in addition to transfusion medicine topics Clinical Laboratory represents submissions concerning tissue transplantation and hematopoietic, cellular and gene therapies.