新设计的椎间假体植入体(乌尔骨)的有限元分析

Sadeer Anwer Mustafa, Ahmed Z. M. Shammari, Wmieth A. J. Augla, M. N. Mohammed
{"title":"新设计的椎间假体植入体(乌尔骨)的有限元分析","authors":"Sadeer Anwer Mustafa, Ahmed Z. M. Shammari, Wmieth A. J. Augla, M. N. Mohammed","doi":"10.22153/kej.2024.01.001","DOIUrl":null,"url":null,"abstract":"   \nImplantation via an intercalary prosthetic method is a reliable solution for reconstructing a long defect in bones that has been damaged by severe disease or accidents. In the current study, using data from computed tomography (CT) scans, a novel costumed intercalary prosthesis design has been created. To achieve this purpose, the CT scan data of a patient in (DICOM) file format was converted into computer-aided design models and saved in stereolithography (STL) format using the 3D slicer (5.0.3) software. The STL files were loaded into Meshmixer software to design the models of the intercalary prosthesis. Finite element analysis (FEA) was applied to validate the strength of the prosthesis with impact, tensile testing, and torsional testing. According to the results of the impact tests, the highest recorded deformation was 8.8485e-002 m in the area where the implant body interfaces with the bone intramedullary canal, due to the high-stress (Von Mises Stress) value of 3.78 E9 PA. In the torsional loaded the highest deformation recorded was 1.3871e-008 m when the Von Mises stress reached 49006 PA,  and with the application of the tensile test the largest deformation measured as 1.0458e-006 m at maximum Von Mises stress seen was recorded as 6.6012e+006 PA which caused that. A solid rod of Ti6Al4V alloy was selected. Finally, the analysis proved that the implant had enhanced mechanical properties. Based on the findings, it can be inferred that the prosthesis was successfully implanted, and a satisfactory result was obtained by using this design method.","PeriodicalId":7637,"journal":{"name":"Al-Khwarizmi Engineering Journal","volume":"143 5","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Finite Element Analysis of New Designed Intercalary Prosthesis Implant (Ulna Bone)\",\"authors\":\"Sadeer Anwer Mustafa, Ahmed Z. M. Shammari, Wmieth A. J. Augla, M. N. Mohammed\",\"doi\":\"10.22153/kej.2024.01.001\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"   \\nImplantation via an intercalary prosthetic method is a reliable solution for reconstructing a long defect in bones that has been damaged by severe disease or accidents. In the current study, using data from computed tomography (CT) scans, a novel costumed intercalary prosthesis design has been created. To achieve this purpose, the CT scan data of a patient in (DICOM) file format was converted into computer-aided design models and saved in stereolithography (STL) format using the 3D slicer (5.0.3) software. The STL files were loaded into Meshmixer software to design the models of the intercalary prosthesis. Finite element analysis (FEA) was applied to validate the strength of the prosthesis with impact, tensile testing, and torsional testing. According to the results of the impact tests, the highest recorded deformation was 8.8485e-002 m in the area where the implant body interfaces with the bone intramedullary canal, due to the high-stress (Von Mises Stress) value of 3.78 E9 PA. In the torsional loaded the highest deformation recorded was 1.3871e-008 m when the Von Mises stress reached 49006 PA,  and with the application of the tensile test the largest deformation measured as 1.0458e-006 m at maximum Von Mises stress seen was recorded as 6.6012e+006 PA which caused that. A solid rod of Ti6Al4V alloy was selected. Finally, the analysis proved that the implant had enhanced mechanical properties. Based on the findings, it can be inferred that the prosthesis was successfully implanted, and a satisfactory result was obtained by using this design method.\",\"PeriodicalId\":7637,\"journal\":{\"name\":\"Al-Khwarizmi Engineering Journal\",\"volume\":\"143 5\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Al-Khwarizmi Engineering Journal\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.22153/kej.2024.01.001\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Al-Khwarizmi Engineering Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22153/kej.2024.01.001","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

通过椎间假体植入的方法是重建因严重疾病或事故而受损的长骨缺损的可靠解决方案。在当前的研究中,利用计算机断层扫描(CT)数据,设计出了一种新颖的仿真椎间假体。为此,使用 3D slicer (5.0.3) 软件将 DICOM 文件格式的患者 CT 扫描数据转换为计算机辅助设计模型,并保存为立体光刻 (STL) 格式。STL 文件被载入 Meshmixer 软件,用于设计椎间假体模型。应用有限元分析(FEA)通过冲击、拉伸测试和扭转测试来验证假体的强度。根据冲击试验的结果,由于高应力(Von Mises Stress)值为 3.78 E9 PA,在假体本体与骨髓内管交接处记录到的最大变形为 8.8485e-002 m。在扭转加载中,当 Von Mises 应力达到 49006 PA 时,记录到的最大变形量为 1.3871e-008 m,而在拉伸试验中,当 Von Mises 应力达到最大值 6.6012e+006 PA 时,测量到的最大变形量为 1.0458e-006 m。因此,我们选择了 Ti6Al4V 合金实心棒。最后,分析表明该植入物具有更强的机械性能。根据分析结果,可以推断该假体已成功植入,并通过这种设计方法获得了令人满意的结果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Finite Element Analysis of New Designed Intercalary Prosthesis Implant (Ulna Bone)
    Implantation via an intercalary prosthetic method is a reliable solution for reconstructing a long defect in bones that has been damaged by severe disease or accidents. In the current study, using data from computed tomography (CT) scans, a novel costumed intercalary prosthesis design has been created. To achieve this purpose, the CT scan data of a patient in (DICOM) file format was converted into computer-aided design models and saved in stereolithography (STL) format using the 3D slicer (5.0.3) software. The STL files were loaded into Meshmixer software to design the models of the intercalary prosthesis. Finite element analysis (FEA) was applied to validate the strength of the prosthesis with impact, tensile testing, and torsional testing. According to the results of the impact tests, the highest recorded deformation was 8.8485e-002 m in the area where the implant body interfaces with the bone intramedullary canal, due to the high-stress (Von Mises Stress) value of 3.78 E9 PA. In the torsional loaded the highest deformation recorded was 1.3871e-008 m when the Von Mises stress reached 49006 PA,  and with the application of the tensile test the largest deformation measured as 1.0458e-006 m at maximum Von Mises stress seen was recorded as 6.6012e+006 PA which caused that. A solid rod of Ti6Al4V alloy was selected. Finally, the analysis proved that the implant had enhanced mechanical properties. Based on the findings, it can be inferred that the prosthesis was successfully implanted, and a satisfactory result was obtained by using this design method.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信