评估 ARK Diagnostics 免疫测定法对尿液中甲苯噻嗪的定性检测。

IF 2.3 3区 医学 Q3 CHEMISTRY, ANALYTICAL
Patrick B Kyle, Christopher J Mattiello, Albert Hua, Joanne M Toohey, Warren R Korn, Alana Saldana-Reed, Douglas F Stickle
{"title":"评估 ARK Diagnostics 免疫测定法对尿液中甲苯噻嗪的定性检测。","authors":"Patrick B Kyle, Christopher J Mattiello, Albert Hua, Joanne M Toohey, Warren R Korn, Alana Saldana-Reed, Douglas F Stickle","doi":"10.1093/jat/bkae043","DOIUrl":null,"url":null,"abstract":"<p><p>Xylazine exposure is common in some US cities, but a commercial assay for routine laboratory testing for xylazine is not currently available. We evaluated a pre-release version of the ARK Diagnostics immunoassay for qualitative detection of xylazine/4-hydroxyxylazine in urine. Studies were conducted using either the semi-quantitative assay application (A. Roche Cobas 503 analyzer) or the qualitative assay application (B. Beckman Coulter AU480 analyzer). Study specimens consisted of deidentified patient urine samples submitted for routine drugs-of-abuse testing. Measurements of xylazine (X) were performed by LC-MS-MS to obtain X-NEGATIVE (X <10 ng/mL) and X-POSITIVE (X ≥10 ng/mL). The semi-quantitative ARK assay was calibrated with a 10 ng/mL cutoff for ARK-POSITVE. For (A): among 74 X-POSITIVE samples, there was 1 ARK-NEGATIVE result (false-negative rate = 1.4%); among 78 X-NEGATIVE samples by LC-MS-MS, there were 0% ARK-POSITIVE results (false-positive rate = 0%). For (B), among 74 X-POSITIVE samples, there were 0 ARK-NEGATIVE results (false-negative rate = 0%); among 78 X-NEGATIVE samples there was 1 ARK-POSITIVE sample (false-positive rate = 1.3%). Common sources of interferences were investigated without evidence of interference. The ARK xylazine/4-OH-xylazine immunoassay was found to be suitable for routine use in screening patient urine samples for presence of xylazine >10 ng/mL.</p>","PeriodicalId":14905,"journal":{"name":"Journal of analytical toxicology","volume":" ","pages":"468-471"},"PeriodicalIF":2.3000,"publicationDate":"2024-07-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluation of the ARK Diagnostics immunoassay for qualitative detection of xylazine in urine.\",\"authors\":\"Patrick B Kyle, Christopher J Mattiello, Albert Hua, Joanne M Toohey, Warren R Korn, Alana Saldana-Reed, Douglas F Stickle\",\"doi\":\"10.1093/jat/bkae043\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Xylazine exposure is common in some US cities, but a commercial assay for routine laboratory testing for xylazine is not currently available. We evaluated a pre-release version of the ARK Diagnostics immunoassay for qualitative detection of xylazine/4-hydroxyxylazine in urine. Studies were conducted using either the semi-quantitative assay application (A. Roche Cobas 503 analyzer) or the qualitative assay application (B. Beckman Coulter AU480 analyzer). Study specimens consisted of deidentified patient urine samples submitted for routine drugs-of-abuse testing. Measurements of xylazine (X) were performed by LC-MS-MS to obtain X-NEGATIVE (X <10 ng/mL) and X-POSITIVE (X ≥10 ng/mL). The semi-quantitative ARK assay was calibrated with a 10 ng/mL cutoff for ARK-POSITVE. For (A): among 74 X-POSITIVE samples, there was 1 ARK-NEGATIVE result (false-negative rate = 1.4%); among 78 X-NEGATIVE samples by LC-MS-MS, there were 0% ARK-POSITIVE results (false-positive rate = 0%). For (B), among 74 X-POSITIVE samples, there were 0 ARK-NEGATIVE results (false-negative rate = 0%); among 78 X-NEGATIVE samples there was 1 ARK-POSITIVE sample (false-positive rate = 1.3%). Common sources of interferences were investigated without evidence of interference. The ARK xylazine/4-OH-xylazine immunoassay was found to be suitable for routine use in screening patient urine samples for presence of xylazine >10 ng/mL.</p>\",\"PeriodicalId\":14905,\"journal\":{\"name\":\"Journal of analytical toxicology\",\"volume\":\" \",\"pages\":\"468-471\"},\"PeriodicalIF\":2.3000,\"publicationDate\":\"2024-07-13\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of analytical toxicology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1093/jat/bkae043\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"CHEMISTRY, ANALYTICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of analytical toxicology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/jat/bkae043","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

摘要

在美国的一些城市,接触异丙嗪很常见,但目前还没有用于异丙嗪常规实验室检测的商业测定法。我们评估了 ARK 诊断公司用于定性检测尿液中甲胺嗪/4-羟基甲胺嗪的免疫测定的预发布版本。研究使用了半定量检测应用(A. 罗氏 Cobas 503 分析仪)或定性检测应用(B. 贝克曼库尔特 AU480 分析仪)。研究标本由提交进行常规药物滥用检测的身份已被确认的患者尿样组成。通过 LC-MS/MS 对羟嗪(X)进行测定,以获得 X-阴性(X10 纳克/毫升,X10 毫克/毫升,X10 毫克/毫升,X10 毫克/毫升,X10 毫克/毫升,X10 毫克/毫升)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Evaluation of the ARK Diagnostics immunoassay for qualitative detection of xylazine in urine.

Xylazine exposure is common in some US cities, but a commercial assay for routine laboratory testing for xylazine is not currently available. We evaluated a pre-release version of the ARK Diagnostics immunoassay for qualitative detection of xylazine/4-hydroxyxylazine in urine. Studies were conducted using either the semi-quantitative assay application (A. Roche Cobas 503 analyzer) or the qualitative assay application (B. Beckman Coulter AU480 analyzer). Study specimens consisted of deidentified patient urine samples submitted for routine drugs-of-abuse testing. Measurements of xylazine (X) were performed by LC-MS-MS to obtain X-NEGATIVE (X <10 ng/mL) and X-POSITIVE (X ≥10 ng/mL). The semi-quantitative ARK assay was calibrated with a 10 ng/mL cutoff for ARK-POSITVE. For (A): among 74 X-POSITIVE samples, there was 1 ARK-NEGATIVE result (false-negative rate = 1.4%); among 78 X-NEGATIVE samples by LC-MS-MS, there were 0% ARK-POSITIVE results (false-positive rate = 0%). For (B), among 74 X-POSITIVE samples, there were 0 ARK-NEGATIVE results (false-negative rate = 0%); among 78 X-NEGATIVE samples there was 1 ARK-POSITIVE sample (false-positive rate = 1.3%). Common sources of interferences were investigated without evidence of interference. The ARK xylazine/4-OH-xylazine immunoassay was found to be suitable for routine use in screening patient urine samples for presence of xylazine >10 ng/mL.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
5.10
自引率
20.00%
发文量
92
审稿时长
6-12 weeks
期刊介绍: The Journal of Analytical Toxicology (JAT) is an international toxicology journal devoted to the timely dissemination of scientific communications concerning potentially toxic substances and drug identification, isolation, and quantitation. Since its inception in 1977, the Journal of Analytical Toxicology has striven to present state-of-the-art techniques used in toxicology labs. The peer-review process provided by the distinguished members of the Editorial Advisory Board ensures the high-quality and integrity of articles published in the Journal of Analytical Toxicology. Timely presentation of the latest toxicology developments is ensured through Technical Notes, Case Reports, and Letters to the Editor.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信