使用保形超声波贴片进行经颅容积成像

IF 50.5 1区 综合性期刊 Q1 MULTIDISCIPLINARY SCIENCES
Nature Pub Date : 2024-05-22 DOI:10.1038/s41586-024-07381-5
Sai Zhou, Xiaoxiang Gao, Geonho Park, Xinyi Yang, Baiyan Qi, Muyang Lin, Hao Huang, Yizhou Bian, Hongjie Hu, Xiangjun Chen, Ray S. Wu, Boyu Liu, Wentong Yue, Chengchangfeng Lu, Ruotao Wang, Pranavi Bheemreddy, Siyu Qin, Arthur Lam, Keith A. Wear, Michael Andre, Erik B. Kistler, David W. Newell, Sheng Xu
{"title":"使用保形超声波贴片进行经颅容积成像","authors":"Sai Zhou, Xiaoxiang Gao, Geonho Park, Xinyi Yang, Baiyan Qi, Muyang Lin, Hao Huang, Yizhou Bian, Hongjie Hu, Xiangjun Chen, Ray S. Wu, Boyu Liu, Wentong Yue, Chengchangfeng Lu, Ruotao Wang, Pranavi Bheemreddy, Siyu Qin, Arthur Lam, Keith A. Wear, Michael Andre, Erik B. Kistler, David W. Newell, Sheng Xu","doi":"10.1038/s41586-024-07381-5","DOIUrl":null,"url":null,"abstract":"Accurate and continuous monitoring of cerebral blood flow is valuable for clinical neurocritical care and fundamental neurovascular research. Transcranial Doppler (TCD) ultrasonography is a widely used non-invasive method for evaluating cerebral blood flow1, but the conventional rigid design severely limits the measurement accuracy of the complex three-dimensional (3D) vascular networks and the practicality for prolonged recording2. Here we report a conformal ultrasound patch for hands-free volumetric imaging and continuous monitoring of cerebral blood flow. The 2 MHz ultrasound waves reduce the attenuation and phase aberration caused by the skull, and the copper mesh shielding layer provides conformal contact to the skin while improving the signal-to-noise ratio by 5 dB. Ultrafast ultrasound imaging based on diverging waves can accurately render the circle of Willis in 3D and minimize human errors during examinations. Focused ultrasound waves allow the recording of blood flow spectra at selected locations continuously. The high accuracy of the conformal ultrasound patch was confirmed in comparison with a conventional TCD probe on 36 participants, showing a mean difference and standard deviation of difference as −1.51 ± 4.34 cm s−1, −0.84 ± 3.06 cm s−1 and −0.50 ± 2.55 cm s−1 for peak systolic velocity, mean flow velocity, and end diastolic velocity, respectively. The measurement success rate was 70.6%, compared with 75.3% for a conventional TCD probe. Furthermore, we demonstrate continuous blood flow spectra during different interventions and identify cascades of intracranial B waves during drowsiness within 4 h of recording. A conformal ultrasound patch can be used for hands-free volumetric imaging and continuous monitoring of cerebral blood flow","PeriodicalId":18787,"journal":{"name":"Nature","volume":"629 8013","pages":"810-818"},"PeriodicalIF":50.5000,"publicationDate":"2024-05-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Transcranial volumetric imaging using a conformal ultrasound patch\",\"authors\":\"Sai Zhou, Xiaoxiang Gao, Geonho Park, Xinyi Yang, Baiyan Qi, Muyang Lin, Hao Huang, Yizhou Bian, Hongjie Hu, Xiangjun Chen, Ray S. Wu, Boyu Liu, Wentong Yue, Chengchangfeng Lu, Ruotao Wang, Pranavi Bheemreddy, Siyu Qin, Arthur Lam, Keith A. Wear, Michael Andre, Erik B. Kistler, David W. Newell, Sheng Xu\",\"doi\":\"10.1038/s41586-024-07381-5\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Accurate and continuous monitoring of cerebral blood flow is valuable for clinical neurocritical care and fundamental neurovascular research. Transcranial Doppler (TCD) ultrasonography is a widely used non-invasive method for evaluating cerebral blood flow1, but the conventional rigid design severely limits the measurement accuracy of the complex three-dimensional (3D) vascular networks and the practicality for prolonged recording2. Here we report a conformal ultrasound patch for hands-free volumetric imaging and continuous monitoring of cerebral blood flow. The 2 MHz ultrasound waves reduce the attenuation and phase aberration caused by the skull, and the copper mesh shielding layer provides conformal contact to the skin while improving the signal-to-noise ratio by 5 dB. Ultrafast ultrasound imaging based on diverging waves can accurately render the circle of Willis in 3D and minimize human errors during examinations. Focused ultrasound waves allow the recording of blood flow spectra at selected locations continuously. The high accuracy of the conformal ultrasound patch was confirmed in comparison with a conventional TCD probe on 36 participants, showing a mean difference and standard deviation of difference as −1.51 ± 4.34 cm s−1, −0.84 ± 3.06 cm s−1 and −0.50 ± 2.55 cm s−1 for peak systolic velocity, mean flow velocity, and end diastolic velocity, respectively. The measurement success rate was 70.6%, compared with 75.3% for a conventional TCD probe. Furthermore, we demonstrate continuous blood flow spectra during different interventions and identify cascades of intracranial B waves during drowsiness within 4 h of recording. A conformal ultrasound patch can be used for hands-free volumetric imaging and continuous monitoring of cerebral blood flow\",\"PeriodicalId\":18787,\"journal\":{\"name\":\"Nature\",\"volume\":\"629 8013\",\"pages\":\"810-818\"},\"PeriodicalIF\":50.5000,\"publicationDate\":\"2024-05-22\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Nature\",\"FirstCategoryId\":\"103\",\"ListUrlMain\":\"https://www.nature.com/articles/s41586-024-07381-5\",\"RegionNum\":1,\"RegionCategory\":\"综合性期刊\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"MULTIDISCIPLINARY SCIENCES\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nature","FirstCategoryId":"103","ListUrlMain":"https://www.nature.com/articles/s41586-024-07381-5","RegionNum":1,"RegionCategory":"综合性期刊","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MULTIDISCIPLINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

摘要

准确、连续地监测脑血流对于临床神经重症监护和神经血管基础研究非常重要。经颅多普勒(TCD)超声造影是评估脑血流的一种广泛使用的无创方法1,但传统的刚性设计严重限制了复杂三维(3D)血管网络的测量精度和长时间记录的实用性2。在此,我们报告了一种用于免提容积成像和连续监测脑血流的保形超声贴片。2 兆赫的超声波可减少头骨造成的衰减和相位畸变,铜网屏蔽层可与皮肤保形接触,同时将信噪比提高 5 分贝。基于发散波的超快超声成像可准确呈现三维威利斯圈,最大程度地减少检查过程中的人为误差。聚焦超声波可连续记录选定位置的血流频谱。与传统的 TCD 探头相比,保形超声贴片的高准确性在 36 名参与者身上得到了证实,显示收缩期峰值速度、平均流速和舒张末期速度的平均差和标准差分别为 -1.51 ± 4.34 cm s-1、-0.84 ± 3.06 cm s-1 和 -0.50 ± 2.55 cm s-1。测量成功率为 70.6%,而传统 TCD 探头的成功率为 75.3%。此外,我们还展示了不同干预过程中的连续血流频谱,并在记录的 4 小时内确定了嗜睡时颅内 B 波的级联。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Transcranial volumetric imaging using a conformal ultrasound patch

Transcranial volumetric imaging using a conformal ultrasound patch

Transcranial volumetric imaging using a conformal ultrasound patch
Accurate and continuous monitoring of cerebral blood flow is valuable for clinical neurocritical care and fundamental neurovascular research. Transcranial Doppler (TCD) ultrasonography is a widely used non-invasive method for evaluating cerebral blood flow1, but the conventional rigid design severely limits the measurement accuracy of the complex three-dimensional (3D) vascular networks and the practicality for prolonged recording2. Here we report a conformal ultrasound patch for hands-free volumetric imaging and continuous monitoring of cerebral blood flow. The 2 MHz ultrasound waves reduce the attenuation and phase aberration caused by the skull, and the copper mesh shielding layer provides conformal contact to the skin while improving the signal-to-noise ratio by 5 dB. Ultrafast ultrasound imaging based on diverging waves can accurately render the circle of Willis in 3D and minimize human errors during examinations. Focused ultrasound waves allow the recording of blood flow spectra at selected locations continuously. The high accuracy of the conformal ultrasound patch was confirmed in comparison with a conventional TCD probe on 36 participants, showing a mean difference and standard deviation of difference as −1.51 ± 4.34 cm s−1, −0.84 ± 3.06 cm s−1 and −0.50 ± 2.55 cm s−1 for peak systolic velocity, mean flow velocity, and end diastolic velocity, respectively. The measurement success rate was 70.6%, compared with 75.3% for a conventional TCD probe. Furthermore, we demonstrate continuous blood flow spectra during different interventions and identify cascades of intracranial B waves during drowsiness within 4 h of recording. A conformal ultrasound patch can be used for hands-free volumetric imaging and continuous monitoring of cerebral blood flow
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Nature
Nature 综合性期刊-综合性期刊
CiteScore
90.00
自引率
1.20%
发文量
3652
审稿时长
3 months
期刊介绍: Nature is a prestigious international journal that publishes peer-reviewed research in various scientific and technological fields. The selection of articles is based on criteria such as originality, importance, interdisciplinary relevance, timeliness, accessibility, elegance, and surprising conclusions. In addition to showcasing significant scientific advances, Nature delivers rapid, authoritative, insightful news, and interpretation of current and upcoming trends impacting science, scientists, and the broader public. The journal serves a dual purpose: firstly, to promptly share noteworthy scientific advances and foster discussions among scientists, and secondly, to ensure the swift dissemination of scientific results globally, emphasizing their significance for knowledge, culture, and daily life.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信