Joachim W Fluhr, Agnès Voisard, Dessyslava G Nikolaeva, Leonie Herzog, Georg Wiora, Jeremias Gayer, Nicolas J Lehmann, Razvigor Darlenski
{"title":"使用蓝牙无线探头测量角质层水合作用:与实验室条件下的测量相比,在家中进行的实际生活研究。","authors":"Joachim W Fluhr, Agnès Voisard, Dessyslava G Nikolaeva, Leonie Herzog, Georg Wiora, Jeremias Gayer, Nicolas J Lehmann, Razvigor Darlenski","doi":"10.1159/000539411","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Non-invasive measurement of the stratum corneum hydration (SCH) with capacitance-based instrumentation is established in dermatological and cosmetic studies. We wanted to test the reliability of non-invasive self-measurements for SCH performed under real-life conditions by volunteers with a Bluetooth-based (wireless) probe Corneometer® (CM 825i) transmitting the data via a smartphone application to a central server. Probes and smartphones communicated using Bluetooth Low Energy. Data from the smartphone were securely transferred to a remote server in a different country with TLS encryption using HTTPS protocols. CM 825i values were correlated with the established CM 825 under laboratory conditions. The primary endpoint was the correlation of the two probes. Secondary endpoints were the coefficient of variation (CV) and delta values (before and after treatment).</p><p><strong>Methods: </strong>Eighteen healthy volunteers (f: 8; m: 10) participated in the prospective observational study. The real-world home use of the wireless CM 825i was performed before and after treatments with base cream DAC for 7 days.</p><p><strong>Results: </strong>Both instruments showed a significant and relevant correlation (p < 0.0001; Spearman coefficient of r = 0.8647). CM 825i and CM 825 differentiate significantly between normal and high SCH. Both devices showed comparable robustness in repeated measurements with a CV between 5.6% and 9.2%.</p><p><strong>Conclusion: </strong>We could show a significant correlation between both devices and a comparable differentiation between low and high SCH and comparable CVs. The real-life use demonstrated adequate acquiring and transmitting of in vivo data to a smartphone and subsequently transmitting to the secure server with low numbers of missed transmissions (<0.2%) and missed measurements (<5%).</p>","PeriodicalId":21748,"journal":{"name":"Skin Pharmacology and Physiology","volume":" ","pages":"40-48"},"PeriodicalIF":2.8000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11488832/pdf/","citationCount":"0","resultStr":"{\"title\":\"Stratum Corneum Hydration Measurements with a Bluetooth Wireless Probe: A Real-Life Study at Home Compared to Measurements under Laboratory Conditions.\",\"authors\":\"Joachim W Fluhr, Agnès Voisard, Dessyslava G Nikolaeva, Leonie Herzog, Georg Wiora, Jeremias Gayer, Nicolas J Lehmann, Razvigor Darlenski\",\"doi\":\"10.1159/000539411\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>Non-invasive measurement of the stratum corneum hydration (SCH) with capacitance-based instrumentation is established in dermatological and cosmetic studies. We wanted to test the reliability of non-invasive self-measurements for SCH performed under real-life conditions by volunteers with a Bluetooth-based (wireless) probe Corneometer® (CM 825i) transmitting the data via a smartphone application to a central server. Probes and smartphones communicated using Bluetooth Low Energy. Data from the smartphone were securely transferred to a remote server in a different country with TLS encryption using HTTPS protocols. CM 825i values were correlated with the established CM 825 under laboratory conditions. The primary endpoint was the correlation of the two probes. Secondary endpoints were the coefficient of variation (CV) and delta values (before and after treatment).</p><p><strong>Methods: </strong>Eighteen healthy volunteers (f: 8; m: 10) participated in the prospective observational study. The real-world home use of the wireless CM 825i was performed before and after treatments with base cream DAC for 7 days.</p><p><strong>Results: </strong>Both instruments showed a significant and relevant correlation (p < 0.0001; Spearman coefficient of r = 0.8647). CM 825i and CM 825 differentiate significantly between normal and high SCH. Both devices showed comparable robustness in repeated measurements with a CV between 5.6% and 9.2%.</p><p><strong>Conclusion: </strong>We could show a significant correlation between both devices and a comparable differentiation between low and high SCH and comparable CVs. The real-life use demonstrated adequate acquiring and transmitting of in vivo data to a smartphone and subsequently transmitting to the secure server with low numbers of missed transmissions (<0.2%) and missed measurements (<5%).</p>\",\"PeriodicalId\":21748,\"journal\":{\"name\":\"Skin Pharmacology and Physiology\",\"volume\":\" \",\"pages\":\"40-48\"},\"PeriodicalIF\":2.8000,\"publicationDate\":\"2024-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11488832/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Skin Pharmacology and Physiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1159/000539411\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/5/22 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"DERMATOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Skin Pharmacology and Physiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1159/000539411","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/5/22 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"DERMATOLOGY","Score":null,"Total":0}
Stratum Corneum Hydration Measurements with a Bluetooth Wireless Probe: A Real-Life Study at Home Compared to Measurements under Laboratory Conditions.
Introduction: Non-invasive measurement of the stratum corneum hydration (SCH) with capacitance-based instrumentation is established in dermatological and cosmetic studies. We wanted to test the reliability of non-invasive self-measurements for SCH performed under real-life conditions by volunteers with a Bluetooth-based (wireless) probe Corneometer® (CM 825i) transmitting the data via a smartphone application to a central server. Probes and smartphones communicated using Bluetooth Low Energy. Data from the smartphone were securely transferred to a remote server in a different country with TLS encryption using HTTPS protocols. CM 825i values were correlated with the established CM 825 under laboratory conditions. The primary endpoint was the correlation of the two probes. Secondary endpoints were the coefficient of variation (CV) and delta values (before and after treatment).
Methods: Eighteen healthy volunteers (f: 8; m: 10) participated in the prospective observational study. The real-world home use of the wireless CM 825i was performed before and after treatments with base cream DAC for 7 days.
Results: Both instruments showed a significant and relevant correlation (p < 0.0001; Spearman coefficient of r = 0.8647). CM 825i and CM 825 differentiate significantly between normal and high SCH. Both devices showed comparable robustness in repeated measurements with a CV between 5.6% and 9.2%.
Conclusion: We could show a significant correlation between both devices and a comparable differentiation between low and high SCH and comparable CVs. The real-life use demonstrated adequate acquiring and transmitting of in vivo data to a smartphone and subsequently transmitting to the secure server with low numbers of missed transmissions (<0.2%) and missed measurements (<5%).
期刊介绍:
In the past decade research into skin pharmacology has rapidly developed with new and promising drugs and therapeutic concepts being introduced regularly. Recently, the use of nanoparticles for drug delivery in dermatology and cosmetology has become a topic of intensive research, yielding remarkable and in part surprising results. Another topic of current research is the use of tissue tolerable plasma in wound treatment. Stimulating not only wound healing processes but also the penetration of topically applied substances into the skin, this novel technique is expected to deliver very interesting results.