{"title":"在严重心力衰竭和左上腔静脉持续存在的患者中植入 CRT-D 时,在左心室导联中保留 PTCA 导丝并随后应用心外膜电极:病例报告。","authors":"Pei-Pei Hou, Yu-Hong Liu, Hai-Bo Qu, Jin Meng, Qiang Li, Zhi-Lin Miao","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>One patient with severe heart failure (LV 92 mm, EF 28%) was treated by cardiac resynchronization therapy (CRT).</p><p><strong>Method: </strong>During the operation, it was found that double superior vena cava coexisted, and selective coronary venography cannot clearly show every branch. It was difficult to push ventriculus sinister electrode to sideward vein, so the electrode was released to far end of frontal septal branch along great cardiac vein.</p><p><strong>Result: </strong>However, because of insufficient braced force of ventriculus sinister electrode, 0.014 PTCA guide wire was detained in the electrode. 2 years later, two spots of PTCA guide wire retained in ventriculus sinister electrode broke in atrium dextrum, so the implantation of epicardial electrode was conducted.</p><p><strong>Conclusion: </strong>After the operation, heart failure was relieved. After 43 months, the battery of pacemaker depleted, so the pacemaker was changed. The effect since follow-up visit was good, LV decreased to 86 mm, EF increased to 32%, and SPWMD time limit shortened from 147 ms to 45 ms. The therapeutic experience of this patient indicated that the effect of detaining PTCA guide wire to enhance braced force in implantation of ventriculus sinister is unreliable and inappropriate to be advocated.</p>","PeriodicalId":13892,"journal":{"name":"International journal of clinical and experimental medicine","volume":"8 8","pages":"14472-8"},"PeriodicalIF":0.2000,"publicationDate":"2015-08-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4613122/pdf/","citationCount":"0","resultStr":"{\"title\":\"Retaining of PTCA guide wire in the left ventricular lead and subsequent application of epicardial electrode when CRT-D implantation in a patient with severe heart failure and persistent left superior vena cava: a case report.\",\"authors\":\"Pei-Pei Hou, Yu-Hong Liu, Hai-Bo Qu, Jin Meng, Qiang Li, Zhi-Lin Miao\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>One patient with severe heart failure (LV 92 mm, EF 28%) was treated by cardiac resynchronization therapy (CRT).</p><p><strong>Method: </strong>During the operation, it was found that double superior vena cava coexisted, and selective coronary venography cannot clearly show every branch. It was difficult to push ventriculus sinister electrode to sideward vein, so the electrode was released to far end of frontal septal branch along great cardiac vein.</p><p><strong>Result: </strong>However, because of insufficient braced force of ventriculus sinister electrode, 0.014 PTCA guide wire was detained in the electrode. 2 years later, two spots of PTCA guide wire retained in ventriculus sinister electrode broke in atrium dextrum, so the implantation of epicardial electrode was conducted.</p><p><strong>Conclusion: </strong>After the operation, heart failure was relieved. After 43 months, the battery of pacemaker depleted, so the pacemaker was changed. The effect since follow-up visit was good, LV decreased to 86 mm, EF increased to 32%, and SPWMD time limit shortened from 147 ms to 45 ms. The therapeutic experience of this patient indicated that the effect of detaining PTCA guide wire to enhance braced force in implantation of ventriculus sinister is unreliable and inappropriate to be advocated.</p>\",\"PeriodicalId\":13892,\"journal\":{\"name\":\"International journal of clinical and experimental medicine\",\"volume\":\"8 8\",\"pages\":\"14472-8\"},\"PeriodicalIF\":0.2000,\"publicationDate\":\"2015-08-15\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4613122/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International journal of clinical and experimental medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2015/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q4\",\"JCRName\":\"MEDICINE, RESEARCH & EXPERIMENTAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International journal of clinical and experimental medicine","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2015/1/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"MEDICINE, RESEARCH & EXPERIMENTAL","Score":null,"Total":0}
Retaining of PTCA guide wire in the left ventricular lead and subsequent application of epicardial electrode when CRT-D implantation in a patient with severe heart failure and persistent left superior vena cava: a case report.
Objective: One patient with severe heart failure (LV 92 mm, EF 28%) was treated by cardiac resynchronization therapy (CRT).
Method: During the operation, it was found that double superior vena cava coexisted, and selective coronary venography cannot clearly show every branch. It was difficult to push ventriculus sinister electrode to sideward vein, so the electrode was released to far end of frontal septal branch along great cardiac vein.
Result: However, because of insufficient braced force of ventriculus sinister electrode, 0.014 PTCA guide wire was detained in the electrode. 2 years later, two spots of PTCA guide wire retained in ventriculus sinister electrode broke in atrium dextrum, so the implantation of epicardial electrode was conducted.
Conclusion: After the operation, heart failure was relieved. After 43 months, the battery of pacemaker depleted, so the pacemaker was changed. The effect since follow-up visit was good, LV decreased to 86 mm, EF increased to 32%, and SPWMD time limit shortened from 147 ms to 45 ms. The therapeutic experience of this patient indicated that the effect of detaining PTCA guide wire to enhance braced force in implantation of ventriculus sinister is unreliable and inappropriate to be advocated.