{"title":"膝关节计算机 X 射线摄影中使用反散射网格的影响","authors":"Luísa Vargas Cassol, Nataly Nogueira Favarin, Felipe de Bail, Edméia Lopes Ramai Buss, Laura Pizarro Trojahn Nogueira, Jéssica Fetzer da Costa Rosa, Thiago Victorino Claus","doi":"10.9734/psij/2024/v28i4834","DOIUrl":null,"url":null,"abstract":"Aims: This experimental study investigated the effect of using an anti-scatter grid in computerized knee radiography (CR) on image quality (IQ) and patient surface radiation dose (Equivalent Surface Air Kerma – Ka,e), measured with an ionization chamber.\nPlace and Duration of Study: The experimental study was conducted between February 2024 and April 2024, in the radiodiagnosis laboratory belonging to the Medical Physics and Radiology Technology courses at the Franciscan University (UFN) in the city of Santa Maria, Rio Grande do Sul.\nMethodology: Utilizing a semi-anatomical knee phantom to simulate clinical examination conditions, ten images were acquired, with five obtained using technique 1 (70 kV, 200mA, and 20mAs) and another five with technique 2 (70 kVp, 200mA, and 5 mAs), with and without an anti-scatter grid, respectively. The phantom images were digitized in a CR system and quantified using a publicly available automatic analyzer software based on histograms and regions of interest (ROI), defined by signal and noise. The obtained results were used to calculate the signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and radiographic contrast (RC), considered IQ descriptors.\nResults: As a selection criterion, the percentage deviation (D%) was chosen, considering technique 1 as the reference concerning technique 2. It was observed that technique 1 showed an SNR 1.20%, RC 3.86%, and Ka,e 73.68% higher than technique 2; on the other hand, technique 2 indicated a CNR 4.76% higher compared to technique 1.\nConclusion:  It is concluded that technique 2 without an anti-scatter grid may be preferable when considering the principle of optimization, where the dose is significantly reduced without a significant loss in IQ descriptors.","PeriodicalId":124795,"journal":{"name":"Physical Science International Journal","volume":"1 3","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-05-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"The Influence of Anti-scatter Grid Usage for Knee Computerized Radiography\",\"authors\":\"Luísa Vargas Cassol, Nataly Nogueira Favarin, Felipe de Bail, Edméia Lopes Ramai Buss, Laura Pizarro Trojahn Nogueira, Jéssica Fetzer da Costa Rosa, Thiago Victorino Claus\",\"doi\":\"10.9734/psij/2024/v28i4834\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Aims: This experimental study investigated the effect of using an anti-scatter grid in computerized knee radiography (CR) on image quality (IQ) and patient surface radiation dose (Equivalent Surface Air Kerma – Ka,e), measured with an ionization chamber.\\nPlace and Duration of Study: The experimental study was conducted between February 2024 and April 2024, in the radiodiagnosis laboratory belonging to the Medical Physics and Radiology Technology courses at the Franciscan University (UFN) in the city of Santa Maria, Rio Grande do Sul.\\nMethodology: Utilizing a semi-anatomical knee phantom to simulate clinical examination conditions, ten images were acquired, with five obtained using technique 1 (70 kV, 200mA, and 20mAs) and another five with technique 2 (70 kVp, 200mA, and 5 mAs), with and without an anti-scatter grid, respectively. The phantom images were digitized in a CR system and quantified using a publicly available automatic analyzer software based on histograms and regions of interest (ROI), defined by signal and noise. The obtained results were used to calculate the signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and radiographic contrast (RC), considered IQ descriptors.\\nResults: As a selection criterion, the percentage deviation (D%) was chosen, considering technique 1 as the reference concerning technique 2. It was observed that technique 1 showed an SNR 1.20%, RC 3.86%, and Ka,e 73.68% higher than technique 2; on the other hand, technique 2 indicated a CNR 4.76% higher compared to technique 1.\\nConclusion:  It is concluded that technique 2 without an anti-scatter grid may be preferable when considering the principle of optimization, where the dose is significantly reduced without a significant loss in IQ descriptors.\",\"PeriodicalId\":124795,\"journal\":{\"name\":\"Physical Science International Journal\",\"volume\":\"1 3\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-05-13\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Physical Science International Journal\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.9734/psij/2024/v28i4834\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physical Science International Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.9734/psij/2024/v28i4834","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:本实验研究调查了在计算机化膝关节放射摄影(CR)中使用反散射网格对图像质量(IQ)和患者表面辐射剂量(等效表面空气Kerma - Ka,e)的影响:实验研究于 2024 年 2 月至 2024 年 4 月期间在位于南里奥格兰德州圣玛丽亚市的方济各大学(UFN)医学物理和放射技术课程所属的放射诊断实验室进行:利用半解剖膝关节模型模拟临床检查条件,采集了十幅图像,其中五幅使用技术 1(70 kV、200mA 和 20mAs),另外五幅使用技术 2(70 kVp、200mA 和 5mAs),分别使用和不使用反散射网格。幻影图像在 CR 系统中进行数字化处理,并使用公开的自动分析软件根据直方图和感兴趣区(ROI)(由信号和噪声定义)进行量化。获得的结果用于计算信噪比(SNR)、对比度-噪声比(CNR)和放射对比度(RC),这些都被视为智商描述符:将技术 1 作为技术 2 的参照,选择百分比偏差 (D%) 作为选择标准。结果显示,技术 1 的信噪比为 1.20%,RC 为 3.86%,Ka,e 为 73.68%;另一方面,技术 2 的 CNR 比技术 1 高 4.76%: 结论:考虑到优化原则,不带反散射栅格的技术 2 更为可取,在不明显降低 IQ 描述指标的情况下,剂量会显著降低。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
The Influence of Anti-scatter Grid Usage for Knee Computerized Radiography
Aims: This experimental study investigated the effect of using an anti-scatter grid in computerized knee radiography (CR) on image quality (IQ) and patient surface radiation dose (Equivalent Surface Air Kerma – Ka,e), measured with an ionization chamber. Place and Duration of Study: The experimental study was conducted between February 2024 and April 2024, in the radiodiagnosis laboratory belonging to the Medical Physics and Radiology Technology courses at the Franciscan University (UFN) in the city of Santa Maria, Rio Grande do Sul. Methodology: Utilizing a semi-anatomical knee phantom to simulate clinical examination conditions, ten images were acquired, with five obtained using technique 1 (70 kV, 200mA, and 20mAs) and another five with technique 2 (70 kVp, 200mA, and 5 mAs), with and without an anti-scatter grid, respectively. The phantom images were digitized in a CR system and quantified using a publicly available automatic analyzer software based on histograms and regions of interest (ROI), defined by signal and noise. The obtained results were used to calculate the signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and radiographic contrast (RC), considered IQ descriptors. Results: As a selection criterion, the percentage deviation (D%) was chosen, considering technique 1 as the reference concerning technique 2. It was observed that technique 1 showed an SNR 1.20%, RC 3.86%, and Ka,e 73.68% higher than technique 2; on the other hand, technique 2 indicated a CNR 4.76% higher compared to technique 1. Conclusion:  It is concluded that technique 2 without an anti-scatter grid may be preferable when considering the principle of optimization, where the dose is significantly reduced without a significant loss in IQ descriptors.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信