三维超声胎儿面部测量的准确性和可重复性:纵向研究

IF 2.2 3区 医学 Q2 OBSTETRICS & GYNECOLOGY
Nerea González-Aranceta , Antonia Alomar , Ricardo Rubio , Silvia Maya-Enero , Antonio Payá , Gemma Piella , Federico Sukno
{"title":"三维超声胎儿面部测量的准确性和可重复性:纵向研究","authors":"Nerea González-Aranceta ,&nbsp;Antonia Alomar ,&nbsp;Ricardo Rubio ,&nbsp;Silvia Maya-Enero ,&nbsp;Antonio Payá ,&nbsp;Gemma Piella ,&nbsp;Federico Sukno","doi":"10.1016/j.earlhumdev.2024.106021","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><p>Fetal face measurements in prenatal ultrasound can aid in identifying craniofacial abnormalities in the developing fetus. However, the accuracy and reliability of ultrasound measurements can be affected by factors such as fetal position, image quality, and the sonographer's expertise. This study assesses the accuracy and reliability of fetal facial measurements in prenatal ultrasound. Additionally, the temporal evolution of measurements is studied, comparing prenatal and postnatal measurements.</p></div><div><h3>Methods</h3><p>Three different experts located up to 23 facial landmarks in 49 prenatal 3D ultrasound scans from normal Caucasian fetuses at weeks 20, 26, and 35 of gestation. Intra- and inter-observer variability was obtained. Postnatal facial measurements were also obtained at 15 days and 1 month postpartum.</p></div><div><h3>Results</h3><p>Most facial landmarks exhibited low errors, with overall intra- and inter-observer errors of 1.01 mm and 1.60 mm, respectively. Landmarks on the nose were found to be the most reliable, while the most challenging ones were those located on the ears and eyes. Overall, scans obtained at 26 weeks of gestation presented the best trade-off between observer variability and landmark visibility. The temporal evolution of the measurements revealed that the lower face area had the highest rate of growth throughout the latest stages of pregnancy.</p></div><div><h3>Conclusions</h3><p>Craniofacial landmarks can be evaluated using 3D fetal ultrasound, especially those located on the nose, mouth, and chin. Despite its limitations, this study provides valuable insights into prenatal and postnatal biometric changes over time, which could aid in developing predictive models for postnatal measurements based on prenatal data.</p></div>","PeriodicalId":11435,"journal":{"name":"Early human development","volume":"193 ","pages":"Article 106021"},"PeriodicalIF":2.2000,"publicationDate":"2024-04-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Accuracy and repeatability of fetal facial measurements in 3D ultrasound: A longitudinal study\",\"authors\":\"Nerea González-Aranceta ,&nbsp;Antonia Alomar ,&nbsp;Ricardo Rubio ,&nbsp;Silvia Maya-Enero ,&nbsp;Antonio Payá ,&nbsp;Gemma Piella ,&nbsp;Federico Sukno\",\"doi\":\"10.1016/j.earlhumdev.2024.106021\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objective</h3><p>Fetal face measurements in prenatal ultrasound can aid in identifying craniofacial abnormalities in the developing fetus. However, the accuracy and reliability of ultrasound measurements can be affected by factors such as fetal position, image quality, and the sonographer's expertise. This study assesses the accuracy and reliability of fetal facial measurements in prenatal ultrasound. Additionally, the temporal evolution of measurements is studied, comparing prenatal and postnatal measurements.</p></div><div><h3>Methods</h3><p>Three different experts located up to 23 facial landmarks in 49 prenatal 3D ultrasound scans from normal Caucasian fetuses at weeks 20, 26, and 35 of gestation. Intra- and inter-observer variability was obtained. Postnatal facial measurements were also obtained at 15 days and 1 month postpartum.</p></div><div><h3>Results</h3><p>Most facial landmarks exhibited low errors, with overall intra- and inter-observer errors of 1.01 mm and 1.60 mm, respectively. Landmarks on the nose were found to be the most reliable, while the most challenging ones were those located on the ears and eyes. Overall, scans obtained at 26 weeks of gestation presented the best trade-off between observer variability and landmark visibility. The temporal evolution of the measurements revealed that the lower face area had the highest rate of growth throughout the latest stages of pregnancy.</p></div><div><h3>Conclusions</h3><p>Craniofacial landmarks can be evaluated using 3D fetal ultrasound, especially those located on the nose, mouth, and chin. Despite its limitations, this study provides valuable insights into prenatal and postnatal biometric changes over time, which could aid in developing predictive models for postnatal measurements based on prenatal data.</p></div>\",\"PeriodicalId\":11435,\"journal\":{\"name\":\"Early human development\",\"volume\":\"193 \",\"pages\":\"Article 106021\"},\"PeriodicalIF\":2.2000,\"publicationDate\":\"2024-04-26\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Early human development\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0378378224000902\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"OBSTETRICS & GYNECOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Early human development","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0378378224000902","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OBSTETRICS & GYNECOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

产前超声胎儿面部测量有助于识别发育中胎儿的颅面畸形。然而,超声测量的准确性和可靠性会受到胎位、图像质量和超声技师的专业知识等因素的影响。本研究评估了产前超声测量胎儿面部的准确性和可靠性。方法三位不同的专家在妊娠 20 周、26 周和 35 周时对 49 个正常高加索胎儿的产前三维超声扫描结果进行定位,最多可找到 23 个面部地标。结果显示了观察者内部和观察者之间的差异性。结果大多数面部地标显示出较低的误差,观察者内部和观察者之间的总体误差分别为 1.01 毫米和 1.60 毫米。鼻子上的地标最可靠,而耳朵和眼睛上的地标最难测量。总体而言,在妊娠 26 周时获得的扫描结果在观察者的可变性和地标可见度之间实现了最佳平衡。测量结果的时间变化显示,在妊娠晚期,面部下部的生长速度最快。结论颅面部地标可通过三维胎儿超声进行评估,尤其是位于鼻子、嘴巴和下巴的地标。尽管存在局限性,但这项研究为了解产前和产后生物特征随时间的变化提供了宝贵的见解,有助于根据产前数据开发产后测量的预测模型。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Accuracy and repeatability of fetal facial measurements in 3D ultrasound: A longitudinal study

Objective

Fetal face measurements in prenatal ultrasound can aid in identifying craniofacial abnormalities in the developing fetus. However, the accuracy and reliability of ultrasound measurements can be affected by factors such as fetal position, image quality, and the sonographer's expertise. This study assesses the accuracy and reliability of fetal facial measurements in prenatal ultrasound. Additionally, the temporal evolution of measurements is studied, comparing prenatal and postnatal measurements.

Methods

Three different experts located up to 23 facial landmarks in 49 prenatal 3D ultrasound scans from normal Caucasian fetuses at weeks 20, 26, and 35 of gestation. Intra- and inter-observer variability was obtained. Postnatal facial measurements were also obtained at 15 days and 1 month postpartum.

Results

Most facial landmarks exhibited low errors, with overall intra- and inter-observer errors of 1.01 mm and 1.60 mm, respectively. Landmarks on the nose were found to be the most reliable, while the most challenging ones were those located on the ears and eyes. Overall, scans obtained at 26 weeks of gestation presented the best trade-off between observer variability and landmark visibility. The temporal evolution of the measurements revealed that the lower face area had the highest rate of growth throughout the latest stages of pregnancy.

Conclusions

Craniofacial landmarks can be evaluated using 3D fetal ultrasound, especially those located on the nose, mouth, and chin. Despite its limitations, this study provides valuable insights into prenatal and postnatal biometric changes over time, which could aid in developing predictive models for postnatal measurements based on prenatal data.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Early human development
Early human development 医学-妇产科学
CiteScore
4.40
自引率
4.00%
发文量
100
审稿时长
46 days
期刊介绍: Established as an authoritative, highly cited voice on early human development, Early Human Development provides a unique opportunity for researchers and clinicians to bridge the communication gap between disciplines. Creating a forum for the productive exchange of ideas concerning early human growth and development, the journal publishes original research and clinical papers with particular emphasis on the continuum between fetal life and the perinatal period; aspects of postnatal growth influenced by early events; and the safeguarding of the quality of human survival. The first comprehensive and interdisciplinary journal in this area of growing importance, Early Human Development offers pertinent contributions to the following subject areas: Fetology; perinatology; pediatrics; growth and development; obstetrics; reproduction and fertility; epidemiology; behavioural sciences; nutrition and metabolism; teratology; neurology; brain biology; developmental psychology and screening.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信