{"title":"SINO 机器人结合 Angio Render 技术辅助立体脑电图电极植入治疗耐药性癫痫的价值。","authors":"Yihai Dai, Rifeng Jiang, Jingyi Zhang, Zhe Qian, Zhen Chen, Songsheng Shi, Shiwei Song","doi":"10.1055/a-2299-7781","DOIUrl":null,"url":null,"abstract":"BACKGROUND AND OBJECTIVE\nStereo-electroencephalography (SEEG) electrodes are implanted using a variety of stereotactic technologies to treat refractory epilepsy. The value of SINO-robot for SEEG electrode implantation is rarely reported. The aim of the current study was to assess the value of SINO-robot in conjunction with Angio Render technology, in SEEG electrode implantation. We also assess its efficacy by examining factors such as localization error, operation time, and complications.\n\n\nMETHODS\nBetween June 2018 and October 2020, we retrospectively reviewed 58 patients who underwent SEEG implantation to resect or ablate their epileptogenic zone (EZ) while minimizing the risk of hemorrhage. SINO-robot combined with Angio Render technology-assisted SEEG electrode implantation was used to visualize each patient' blood vessel in a 3D plane. The 3D view functionality was used to increase the safety and accuracy of the implantation, and reducing the risk of hemorrhage by avoiding said blood vessel.\n\n\nRESULTS\nIn this study, 634 SEEG electrodes were implanted in 58 patients. The mean 10.92(range 5- 18) leads per patient. The mean entry point localization error (EPLE) was 0.94 ± 0.23 mm (range: 0.39- 1.63 mm), and the mean target point localization error (TPLE) was 1.49 ± 0.37 mm (range: 0.80-2.78 mm). The mean operating time per lead (MOTPL) was 6. 18 ± 1.80 min (range: 3.02- 14.61 min). And the mean depth of electrodes was 56.96± 3.62 mm (range:27.23-124.85 mm). At a follow-up of at least one year, totally 81.57% (47/58) of patients achieved an Engel class I of seizure freedom. There were 2 patients with asymptomatic brain hematomas following SEEG placement, and no late complications or mortality in this cohort.\n\n\nCONCLUSIONS\nSINO-robot in conjunction with Angio Render technology assist, in SEEG electrode implantation is safe and accurate in mitigating the risk of intracranial hemorrhage in patients suffering from drug-resistant epilepsy.","PeriodicalId":16544,"journal":{"name":"Journal of neurological surgery. Part A, Central European neurosurgery","volume":null,"pages":null},"PeriodicalIF":0.9000,"publicationDate":"2024-04-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"The value of SINO robot combined with Angio Render technology-assisted stereo-electroencephalography electrode implantation for the treatment of drug-resistant epilepsy.\",\"authors\":\"Yihai Dai, Rifeng Jiang, Jingyi Zhang, Zhe Qian, Zhen Chen, Songsheng Shi, Shiwei Song\",\"doi\":\"10.1055/a-2299-7781\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"BACKGROUND AND OBJECTIVE\\nStereo-electroencephalography (SEEG) electrodes are implanted using a variety of stereotactic technologies to treat refractory epilepsy. The value of SINO-robot for SEEG electrode implantation is rarely reported. The aim of the current study was to assess the value of SINO-robot in conjunction with Angio Render technology, in SEEG electrode implantation. We also assess its efficacy by examining factors such as localization error, operation time, and complications.\\n\\n\\nMETHODS\\nBetween June 2018 and October 2020, we retrospectively reviewed 58 patients who underwent SEEG implantation to resect or ablate their epileptogenic zone (EZ) while minimizing the risk of hemorrhage. SINO-robot combined with Angio Render technology-assisted SEEG electrode implantation was used to visualize each patient' blood vessel in a 3D plane. The 3D view functionality was used to increase the safety and accuracy of the implantation, and reducing the risk of hemorrhage by avoiding said blood vessel.\\n\\n\\nRESULTS\\nIn this study, 634 SEEG electrodes were implanted in 58 patients. The mean 10.92(range 5- 18) leads per patient. The mean entry point localization error (EPLE) was 0.94 ± 0.23 mm (range: 0.39- 1.63 mm), and the mean target point localization error (TPLE) was 1.49 ± 0.37 mm (range: 0.80-2.78 mm). The mean operating time per lead (MOTPL) was 6. 18 ± 1.80 min (range: 3.02- 14.61 min). And the mean depth of electrodes was 56.96± 3.62 mm (range:27.23-124.85 mm). At a follow-up of at least one year, totally 81.57% (47/58) of patients achieved an Engel class I of seizure freedom. There were 2 patients with asymptomatic brain hematomas following SEEG placement, and no late complications or mortality in this cohort.\\n\\n\\nCONCLUSIONS\\nSINO-robot in conjunction with Angio Render technology assist, in SEEG electrode implantation is safe and accurate in mitigating the risk of intracranial hemorrhage in patients suffering from drug-resistant epilepsy.\",\"PeriodicalId\":16544,\"journal\":{\"name\":\"Journal of neurological surgery. Part A, Central European neurosurgery\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.9000,\"publicationDate\":\"2024-04-04\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of neurological surgery. Part A, Central European neurosurgery\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1055/a-2299-7781\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"CLINICAL NEUROLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of neurological surgery. Part A, Central European neurosurgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1055/a-2299-7781","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
The value of SINO robot combined with Angio Render technology-assisted stereo-electroencephalography electrode implantation for the treatment of drug-resistant epilepsy.
BACKGROUND AND OBJECTIVE
Stereo-electroencephalography (SEEG) electrodes are implanted using a variety of stereotactic technologies to treat refractory epilepsy. The value of SINO-robot for SEEG electrode implantation is rarely reported. The aim of the current study was to assess the value of SINO-robot in conjunction with Angio Render technology, in SEEG electrode implantation. We also assess its efficacy by examining factors such as localization error, operation time, and complications.
METHODS
Between June 2018 and October 2020, we retrospectively reviewed 58 patients who underwent SEEG implantation to resect or ablate their epileptogenic zone (EZ) while minimizing the risk of hemorrhage. SINO-robot combined with Angio Render technology-assisted SEEG electrode implantation was used to visualize each patient' blood vessel in a 3D plane. The 3D view functionality was used to increase the safety and accuracy of the implantation, and reducing the risk of hemorrhage by avoiding said blood vessel.
RESULTS
In this study, 634 SEEG electrodes were implanted in 58 patients. The mean 10.92(range 5- 18) leads per patient. The mean entry point localization error (EPLE) was 0.94 ± 0.23 mm (range: 0.39- 1.63 mm), and the mean target point localization error (TPLE) was 1.49 ± 0.37 mm (range: 0.80-2.78 mm). The mean operating time per lead (MOTPL) was 6. 18 ± 1.80 min (range: 3.02- 14.61 min). And the mean depth of electrodes was 56.96± 3.62 mm (range:27.23-124.85 mm). At a follow-up of at least one year, totally 81.57% (47/58) of patients achieved an Engel class I of seizure freedom. There were 2 patients with asymptomatic brain hematomas following SEEG placement, and no late complications or mortality in this cohort.
CONCLUSIONS
SINO-robot in conjunction with Angio Render technology assist, in SEEG electrode implantation is safe and accurate in mitigating the risk of intracranial hemorrhage in patients suffering from drug-resistant epilepsy.
期刊介绍:
The Journal of Neurological Surgery Part A: Central European Neurosurgery (JNLS A) is a major publication from the world''s leading publisher in neurosurgery. JNLS A currently serves as the official organ of several national neurosurgery societies.
JNLS A is a peer-reviewed journal publishing original research, review articles, and technical notes covering all aspects of neurological surgery. The focus of JNLS A includes microsurgery as well as the latest minimally invasive techniques, such as stereotactic-guided surgery, endoscopy, and endovascular procedures. JNLS A covers purely neurosurgical topics.