专利蓝干扰了前哨淋巴结患者的脂血指数测量。

Keiichi Nakano, Masanori Seimiya, Kojiro Yamazaki, Keiko Yasuda, N. Yamashita, Hideki Goto, Takanori Teshima
{"title":"专利蓝干扰了前哨淋巴结患者的脂血指数测量。","authors":"Keiichi Nakano, Masanori Seimiya, Kojiro Yamazaki, Keiko Yasuda, N. Yamashita, Hideki Goto, Takanori Teshima","doi":"10.1093/labmed/lmae028","DOIUrl":null,"url":null,"abstract":"Lipids interfere with absorbance measurements conducted using colorimetric methods. To monitor lipemia, some systems measure absorbance using an analyzer. This report describes a novel case of interference with the lipemia index without lipemia. A 64-year-old woman with giant basal cell carcinoma underwent resection and sentinel lymph node biopsy. The patient had been subcutaneously injected with patent blue during sentinel lymph node resection. After surgery, her serum and urine were yellow-green, and the lipemia index, calculated by measuring absorbance at 658 nm (main wavelength) and 694 nm (secondary wavelength) using a JCA-BM8040 chemistry analyzer, was high. The absorbance spectrum of the patient's serum and patent blue solution were compared to determine the cause of the high lipemia index. The patient's serum and the patent blue solution showed absorption at wavelengths between 540 and 698 nm. Moreover, the absorbance was concentration-dependent for patent blue. These results thus indicated that the patient's serum contained patent blue. Here, we report a case wherein patent blue affected the lipemia index. Thus, it must be noted that patent blue injection may yield inaccurate results when evaluating lipemia index.","PeriodicalId":17951,"journal":{"name":"Laboratory medicine","volume":"68 9","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-04-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Patent blue interferes with the measurement of lipemia index in a patient with sentinel lymph node.\",\"authors\":\"Keiichi Nakano, Masanori Seimiya, Kojiro Yamazaki, Keiko Yasuda, N. Yamashita, Hideki Goto, Takanori Teshima\",\"doi\":\"10.1093/labmed/lmae028\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Lipids interfere with absorbance measurements conducted using colorimetric methods. To monitor lipemia, some systems measure absorbance using an analyzer. This report describes a novel case of interference with the lipemia index without lipemia. A 64-year-old woman with giant basal cell carcinoma underwent resection and sentinel lymph node biopsy. The patient had been subcutaneously injected with patent blue during sentinel lymph node resection. After surgery, her serum and urine were yellow-green, and the lipemia index, calculated by measuring absorbance at 658 nm (main wavelength) and 694 nm (secondary wavelength) using a JCA-BM8040 chemistry analyzer, was high. The absorbance spectrum of the patient's serum and patent blue solution were compared to determine the cause of the high lipemia index. The patient's serum and the patent blue solution showed absorption at wavelengths between 540 and 698 nm. Moreover, the absorbance was concentration-dependent for patent blue. These results thus indicated that the patient's serum contained patent blue. Here, we report a case wherein patent blue affected the lipemia index. Thus, it must be noted that patent blue injection may yield inaccurate results when evaluating lipemia index.\",\"PeriodicalId\":17951,\"journal\":{\"name\":\"Laboratory medicine\",\"volume\":\"68 9\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-04-15\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Laboratory medicine\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1093/labmed/lmae028\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Laboratory medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/labmed/lmae028","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

血脂会干扰使用比色法进行的吸光度测量。为了监测脂血,一些系统使用分析仪测量吸光度。本报告描述了一例没有脂血却干扰脂血指数的新病例。一名患有巨大基底细胞癌的 64 岁女性接受了切除术和前哨淋巴结活检。在前哨淋巴结切除术中,患者皮下注射了专利蓝。术后,她的血清和尿液呈黄绿色,使用 JCA-BM8040 化学分析仪在 658 纳米(主波长)和 694 纳米(副波长)处测量吸光度计算得出的脂血指数很高。通过比较患者血清和专利蓝溶液的吸光度光谱,确定导致高脂血症指数的原因。患者血清和专利蓝溶液在波长 540 至 698 纳米之间有吸收。此外,漆蓝的吸收率与浓度有关。这些结果表明患者的血清中含有专利蓝。在此,我们报告了一个影响脂血指数的病例。因此,在评估脂血指数时必须注意,注射专利蓝可能会产生不准确的结果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Patent blue interferes with the measurement of lipemia index in a patient with sentinel lymph node.
Lipids interfere with absorbance measurements conducted using colorimetric methods. To monitor lipemia, some systems measure absorbance using an analyzer. This report describes a novel case of interference with the lipemia index without lipemia. A 64-year-old woman with giant basal cell carcinoma underwent resection and sentinel lymph node biopsy. The patient had been subcutaneously injected with patent blue during sentinel lymph node resection. After surgery, her serum and urine were yellow-green, and the lipemia index, calculated by measuring absorbance at 658 nm (main wavelength) and 694 nm (secondary wavelength) using a JCA-BM8040 chemistry analyzer, was high. The absorbance spectrum of the patient's serum and patent blue solution were compared to determine the cause of the high lipemia index. The patient's serum and the patent blue solution showed absorption at wavelengths between 540 and 698 nm. Moreover, the absorbance was concentration-dependent for patent blue. These results thus indicated that the patient's serum contained patent blue. Here, we report a case wherein patent blue affected the lipemia index. Thus, it must be noted that patent blue injection may yield inaccurate results when evaluating lipemia index.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信