{"title":"利用 LC-MS/MS 对丙泊酚代谢物进行定量,为尿液药物监测提供长检测窗","authors":"Thomas G. Rosano ,&nbsp;Michelle Wood ,&nbsp;Kiley L. Scholz ,&nbsp;Kiera Whitely ,&nbsp;Nathaniel Kim ,&nbsp;Melissa Ehlers","doi":"10.1016/j.jmsacl.2024.04.001","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>Chromatographic methods for analysis of propofol and its metabolites have been widely used in pharmacokinetic studies of propofol distribution, metabolism, and clearance. Application of chromatographic methods is also needed in clinical and forensic laboratories for detecting and monitoring propofol misuse.</p></div><div><h3>Objective</h3><p>We report a method for sensitive analysis of propofol, propofol 1-glucuronide (PG), 4-hydroxypropofol 1-glucuronide (1-QG), 4-hydroxypropofol 4-glucuronide (4-QG) and 4-hydroxypropofol 4-sulfate (4-QS) in urine by LC–MS/MS analysis. The method employs a simple dilute-and-analyze sample preparation with stable isotope internal standardization.</p></div><div><h3>Results</h3><p>Validation studies demonstrate a linear calibration model (100–10,000 ng/mL), with dilution integrity verified for the extended range of concentrations experienced in propofol use. Criteria-based validation was achieved, including an average coefficient of variation of 6.5 % and a percent bias of −4.2 ng/mL. The method was evaluated in 12 surgical patients, with monitoring periods lasting up to 30 days following intravenous propofol administrations of 100–3000 mg on the day of surgery. While the concentration ratio of PG to 4-hydroxy propofol metabolite decreased significantly in the days following surgery, PG maintained the highest concentration in all specimens. Both PG and 1-QG were detectable throughout the monitoring periods, including in a patient monitored for 30 days. Lower concentrations were determined for 4-QG and 4-QS, with evidence of detection up to 20 days. Propofol was not detectable in any urine specimens, thereby proving ineffective for identifying drug use.</p></div><div><h3>Conclusion</h3><p>The validated method for quantifying propofol metabolites demonstrates its applicability for the sensitive detection of propofol misuse over a long window of drug-use detection.</p></div>","PeriodicalId":52406,"journal":{"name":"Journal of Mass Spectrometry and Advances in the Clinical Lab","volume":"32 ","pages":"Pages 60-67"},"PeriodicalIF":3.1000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2667145X24000191/pdfft?md5=18d77756ab2034ba58526495d0704b3b&pid=1-s2.0-S2667145X24000191-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Quantitation of propofol metabolites by LC–MS/MS demonstrating long detection window for urine drug monitoring\",\"authors\":\"Thomas G. Rosano ,&nbsp;Michelle Wood ,&nbsp;Kiley L. Scholz ,&nbsp;Kiera Whitely ,&nbsp;Nathaniel Kim ,&nbsp;Melissa Ehlers\",\"doi\":\"10.1016/j.jmsacl.2024.04.001\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Introduction</h3><p>Chromatographic methods for analysis of propofol and its metabolites have been widely used in pharmacokinetic studies of propofol distribution, metabolism, and clearance. Application of chromatographic methods is also needed in clinical and forensic laboratories for detecting and monitoring propofol misuse.</p></div><div><h3>Objective</h3><p>We report a method for sensitive analysis of propofol, propofol 1-glucuronide (PG), 4-hydroxypropofol 1-glucuronide (1-QG), 4-hydroxypropofol 4-glucuronide (4-QG) and 4-hydroxypropofol 4-sulfate (4-QS) in urine by LC–MS/MS analysis. The method employs a simple dilute-and-analyze sample preparation with stable isotope internal standardization.</p></div><div><h3>Results</h3><p>Validation studies demonstrate a linear calibration model (100–10,000 ng/mL), with dilution integrity verified for the extended range of concentrations experienced in propofol use. Criteria-based validation was achieved, including an average coefficient of variation of 6.5 % and a percent bias of −4.2 ng/mL. The method was evaluated in 12 surgical patients, with monitoring periods lasting up to 30 days following intravenous propofol administrations of 100–3000 mg on the day of surgery. While the concentration ratio of PG to 4-hydroxy propofol metabolite decreased significantly in the days following surgery, PG maintained the highest concentration in all specimens. Both PG and 1-QG were detectable throughout the monitoring periods, including in a patient monitored for 30 days. Lower concentrations were determined for 4-QG and 4-QS, with evidence of detection up to 20 days. Propofol was not detectable in any urine specimens, thereby proving ineffective for identifying drug use.</p></div><div><h3>Conclusion</h3><p>The validated method for quantifying propofol metabolites demonstrates its applicability for the sensitive detection of propofol misuse over a long window of drug-use detection.</p></div>\",\"PeriodicalId\":52406,\"journal\":{\"name\":\"Journal of Mass Spectrometry and Advances in the Clinical Lab\",\"volume\":\"32 \",\"pages\":\"Pages 60-67\"},\"PeriodicalIF\":3.1000,\"publicationDate\":\"2024-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2667145X24000191/pdfft?md5=18d77756ab2034ba58526495d0704b3b&pid=1-s2.0-S2667145X24000191-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Mass Spectrometry and Advances in the Clinical Lab\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2667145X24000191\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Mass Spectrometry and Advances in the Clinical Lab","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2667145X24000191","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

导言分析异丙酚及其代谢物的色谱法已广泛应用于异丙酚分布、代谢和清除的药代动力学研究。我们报告了一种利用 LC-MS/MS灵敏分析尿液中丙泊酚、丙泊酚 1-葡萄糖醛酸苷(PG)、4-羟基丙泊酚 1-葡萄糖醛酸苷(1-QG)、4-羟基丙泊酚 4-葡萄糖醛酸苷(4-QG)和 4-羟基丙泊酚 4-硫酸盐(4-QS)的方法。结果验证研究表明,该方法具有线性校准模型(100-10,000 ng/mL),在使用异丙酚的更大浓度范围内,稀释的完整性得到了验证。实现了基于标准的验证,包括 6.5 % 的平均变异系数和 -4.2 纳克/毫升的百分比偏差。该方法在 12 名手术患者中进行了评估,在手术当天静脉注射 100-3000 毫克异丙酚后进行了长达 30 天的监测。在手术后的几天内,PG 与 4-羟基异丙酚代谢物的浓度比显著下降,但 PG 在所有样本中都保持最高浓度。在整个监测期间,包括在一名接受了 30 天监测的患者体内,都能检测到 PG 和 1-QG。4-QG 和 4-QS 的浓度较低,有证据表明在 20 天内仍可检测到。结论经验证的异丙酚代谢物定量方法证明其适用于在较长的药物使用检测期内灵敏地检测异丙酚的滥用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Quantitation of propofol metabolites by LC–MS/MS demonstrating long detection window for urine drug monitoring

Introduction

Chromatographic methods for analysis of propofol and its metabolites have been widely used in pharmacokinetic studies of propofol distribution, metabolism, and clearance. Application of chromatographic methods is also needed in clinical and forensic laboratories for detecting and monitoring propofol misuse.

Objective

We report a method for sensitive analysis of propofol, propofol 1-glucuronide (PG), 4-hydroxypropofol 1-glucuronide (1-QG), 4-hydroxypropofol 4-glucuronide (4-QG) and 4-hydroxypropofol 4-sulfate (4-QS) in urine by LC–MS/MS analysis. The method employs a simple dilute-and-analyze sample preparation with stable isotope internal standardization.

Results

Validation studies demonstrate a linear calibration model (100–10,000 ng/mL), with dilution integrity verified for the extended range of concentrations experienced in propofol use. Criteria-based validation was achieved, including an average coefficient of variation of 6.5 % and a percent bias of −4.2 ng/mL. The method was evaluated in 12 surgical patients, with monitoring periods lasting up to 30 days following intravenous propofol administrations of 100–3000 mg on the day of surgery. While the concentration ratio of PG to 4-hydroxy propofol metabolite decreased significantly in the days following surgery, PG maintained the highest concentration in all specimens. Both PG and 1-QG were detectable throughout the monitoring periods, including in a patient monitored for 30 days. Lower concentrations were determined for 4-QG and 4-QS, with evidence of detection up to 20 days. Propofol was not detectable in any urine specimens, thereby proving ineffective for identifying drug use.

Conclusion

The validated method for quantifying propofol metabolites demonstrates its applicability for the sensitive detection of propofol misuse over a long window of drug-use detection.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Mass Spectrometry and Advances in the Clinical Lab
Journal of Mass Spectrometry and Advances in the Clinical Lab Health Professions-Medical Laboratory Technology
CiteScore
4.30
自引率
18.20%
发文量
41
审稿时长
81 days
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信