{"title":"根据 ISO 标准 81060-2:2018,在成人和儿童人群中对 Mindray VS9 生命体征监护仪进行验证。","authors":"Na Guo, Yihan Zhang, Weiqiang Chen, Hexian Zhong, Liping Li, Hanbin Xie, Wenxiu Zhu, Jun Liu, Shangrong Li","doi":"10.1097/MBP.0000000000000704","DOIUrl":null,"url":null,"abstract":"<p><p>We aimed to validate the accuracy of the Mindray VS9 Vital Signs Monitor, which features the Mindray TrueBP inflation algorithm for oscillometric blood pressure (BP) measurement, to check if it complies with the International Organization for Standardization Standard (ISO 81060-2:2018) in a combined adult and pediatric population. A total of 86 participants, including both adult and pediatric subjects, were recruited. The distribution of their ages, gender, BPs and limb sizes all complied with the requirement of the ISO standard. The inflation and deflation algorithms were validated independently using the same-arm sequential BP measurement method. For each subject, the BP was first determined by two independent observers using a mercury sphygmomanometer (R1). The BP of the subject was then determined by the third observer using the test equipment (T1). Then, using a mercury sphygmomanometer, two independent observers were asked to determine the subject's BP (R2) again. R1-T1-R2 were considered a valid pair of data. This cycle continued until 3 pairs of valid data were achieved. We collected 258 pairs of valid BP data for the validation of the inflation and deflation algorithms respectively. For validation Criterion 1, the mean ± SD of the differences between the readings obtained from the test device and reference BP was 0.0 ± 6.6/-1.8 ± 7.1 mmHg (systolic/diastolic) when the deflation algorithm was used, and 2.4 ± 6.3/ 0.3 ± 6.9 mmHg (systolic/diastolic) when the inflation algorithm was used. For validation Criterion 2, the SD of the averaged BP differences between the test device and the reference BP per subject was 5.35/6.33 mmHg (systolic/diastolic) when the deflation algorithm was used, and 5.17/5.75 mmHg (systolic/diastolic) when the inflation algorithm was used. The VS9 Vital Signs Monitor fulfilled all the criteria in the ISO Standard. Moreover, the inflation algorithm had a shorter Measure Time (by 7-21 s) and lower maximum inflation pressure (by 9.7-22 mmHg). The VS9 Vital Signs Monitor fulfilled all the requirements of the ISO Standard (ISO 81060-2:2018) in a combined adult and pediatric population and is recommended for clinical use.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11198959/pdf/","citationCount":"0","resultStr":"{\"title\":\"Validation of the Mindray VS9 Vital Signs Monitor in a combined adult and pediatric population according to ISO Standard 81060-2:2018.\",\"authors\":\"Na Guo, Yihan Zhang, Weiqiang Chen, Hexian Zhong, Liping Li, Hanbin Xie, Wenxiu Zhu, Jun Liu, Shangrong Li\",\"doi\":\"10.1097/MBP.0000000000000704\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>We aimed to validate the accuracy of the Mindray VS9 Vital Signs Monitor, which features the Mindray TrueBP inflation algorithm for oscillometric blood pressure (BP) measurement, to check if it complies with the International Organization for Standardization Standard (ISO 81060-2:2018) in a combined adult and pediatric population. A total of 86 participants, including both adult and pediatric subjects, were recruited. The distribution of their ages, gender, BPs and limb sizes all complied with the requirement of the ISO standard. The inflation and deflation algorithms were validated independently using the same-arm sequential BP measurement method. For each subject, the BP was first determined by two independent observers using a mercury sphygmomanometer (R1). The BP of the subject was then determined by the third observer using the test equipment (T1). Then, using a mercury sphygmomanometer, two independent observers were asked to determine the subject's BP (R2) again. R1-T1-R2 were considered a valid pair of data. This cycle continued until 3 pairs of valid data were achieved. We collected 258 pairs of valid BP data for the validation of the inflation and deflation algorithms respectively. For validation Criterion 1, the mean ± SD of the differences between the readings obtained from the test device and reference BP was 0.0 ± 6.6/-1.8 ± 7.1 mmHg (systolic/diastolic) when the deflation algorithm was used, and 2.4 ± 6.3/ 0.3 ± 6.9 mmHg (systolic/diastolic) when the inflation algorithm was used. For validation Criterion 2, the SD of the averaged BP differences between the test device and the reference BP per subject was 5.35/6.33 mmHg (systolic/diastolic) when the deflation algorithm was used, and 5.17/5.75 mmHg (systolic/diastolic) when the inflation algorithm was used. The VS9 Vital Signs Monitor fulfilled all the criteria in the ISO Standard. Moreover, the inflation algorithm had a shorter Measure Time (by 7-21 s) and lower maximum inflation pressure (by 9.7-22 mmHg). The VS9 Vital Signs Monitor fulfilled all the requirements of the ISO Standard (ISO 81060-2:2018) in a combined adult and pediatric population and is recommended for clinical use.</p>\",\"PeriodicalId\":8950,\"journal\":{\"name\":\"Blood Pressure Monitoring\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2024-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11198959/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Blood Pressure Monitoring\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MBP.0000000000000704\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/3/25 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"PERIPHERAL VASCULAR DISEASE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000704","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/3/25 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
Validation of the Mindray VS9 Vital Signs Monitor in a combined adult and pediatric population according to ISO Standard 81060-2:2018.
We aimed to validate the accuracy of the Mindray VS9 Vital Signs Monitor, which features the Mindray TrueBP inflation algorithm for oscillometric blood pressure (BP) measurement, to check if it complies with the International Organization for Standardization Standard (ISO 81060-2:2018) in a combined adult and pediatric population. A total of 86 participants, including both adult and pediatric subjects, were recruited. The distribution of their ages, gender, BPs and limb sizes all complied with the requirement of the ISO standard. The inflation and deflation algorithms were validated independently using the same-arm sequential BP measurement method. For each subject, the BP was first determined by two independent observers using a mercury sphygmomanometer (R1). The BP of the subject was then determined by the third observer using the test equipment (T1). Then, using a mercury sphygmomanometer, two independent observers were asked to determine the subject's BP (R2) again. R1-T1-R2 were considered a valid pair of data. This cycle continued until 3 pairs of valid data were achieved. We collected 258 pairs of valid BP data for the validation of the inflation and deflation algorithms respectively. For validation Criterion 1, the mean ± SD of the differences between the readings obtained from the test device and reference BP was 0.0 ± 6.6/-1.8 ± 7.1 mmHg (systolic/diastolic) when the deflation algorithm was used, and 2.4 ± 6.3/ 0.3 ± 6.9 mmHg (systolic/diastolic) when the inflation algorithm was used. For validation Criterion 2, the SD of the averaged BP differences between the test device and the reference BP per subject was 5.35/6.33 mmHg (systolic/diastolic) when the deflation algorithm was used, and 5.17/5.75 mmHg (systolic/diastolic) when the inflation algorithm was used. The VS9 Vital Signs Monitor fulfilled all the criteria in the ISO Standard. Moreover, the inflation algorithm had a shorter Measure Time (by 7-21 s) and lower maximum inflation pressure (by 9.7-22 mmHg). The VS9 Vital Signs Monitor fulfilled all the requirements of the ISO Standard (ISO 81060-2:2018) in a combined adult and pediatric population and is recommended for clinical use.
期刊介绍:
Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research.
This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine.
Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.