{"title":"基于光学相干断层扫描的新型动态实时可视化 25 kHz 扫源生物测量仪的评估。","authors":"Jinjin Yu, Xuanqiao Lin, Xiaomin Huang, Zhenyu Xu, Rui Ning, Kexin Li, Giacomo Savini, Domenico Schiano-Lomoriello, Xingtao Zhou, Jinhai Huang","doi":"10.1186/s40662-024-00377-2","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>To evaluate the intraobserver repeatability and interobserver reproducibility of a newly developed dynamic real-time visualization 25 kHz swept-source optical coherence tomography (SS-OCT) based biometer (ZW-30, TowardPi Medical Technology Ltd, China) and compare its agreement with another SS-OCT based biometer (IOLMaster 700, Carl Zeiss Meditec AG, Jena, Germany).</p><p><strong>Methods: </strong>Eighty-two healthy right eyes were enrolled in this prospective observational study. Measurements were repeated for three times using the ZW-30 and IOLMaster 700 in a random order. Obtained parameters included axial length (AL), central corneal thickness (CCT), aqueous depth (AQD), anterior chamber depth (ACD), lens thickness (LT), mean keratometry (Km), astigmatism magnitude (AST), vector J<sub>0</sub>, vector J<sub>45</sub>, and corneal diameter (CD). The within-subject standard deviation (Sw), test-retest (TRT) variability, coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were adopted to assess the intraobserver repeatability and interobserver reproducibility. The double-angle plot was also used to display the distribution of AST. To estimate agreement, Bland-Altman plots were used.</p><p><strong>Results: </strong>For the intraobserver repeatability and interobserver reproducibility, the Sw, TRT and CoV for all parameters were low. Meanwhile, the ICC values were all close to 1.000, except for the J<sub>45</sub> (ICC = 0.887 for the intraobserver repeatability). The double-angle plot showed that the distribution of AST measured by these two devices was similar. For agreement, the Bland-Altman plots showed narrow 95% limits of agreements (LoAs) for AL, CCT, AQD, ACD, LT, Km AST, J<sub>0</sub>, J<sub>45</sub>, and CD (- 0.02 mm to 0.02 mm, - 7.49 μm to 8.08 μm, - 0.07 mm to 0.04 mm, - 0.07 mm to 0.04 mm, - 0.07 mm to 0.08 mm, - 0.16 D to 0.30 D, - 0.30 D to 0.29 D, - 0.16 D to 0.16 D, - 0.23 D to 0.13 D, and - 0.39 mm to 0.10 mm, respectively).</p><p><strong>Conclusions: </strong>The newly dynamic real-time visualization biometer exhibited excellent intraobserver repeatability and interobserver reproducibility. The two devices both based on the SS-OCT principle had similar ocular parameters measurement values and can be interchanged in clinical practice.</p>","PeriodicalId":12194,"journal":{"name":"Eye and Vision","volume":"11 1","pages":"9"},"PeriodicalIF":4.1000,"publicationDate":"2024-03-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10910812/pdf/","citationCount":"0","resultStr":"{\"title\":\"Evaluation of a new dynamic real-time visualization 25 kHz swept-source optical coherence tomography based biometer.\",\"authors\":\"Jinjin Yu, Xuanqiao Lin, Xiaomin Huang, Zhenyu Xu, Rui Ning, Kexin Li, Giacomo Savini, Domenico Schiano-Lomoriello, Xingtao Zhou, Jinhai Huang\",\"doi\":\"10.1186/s40662-024-00377-2\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>To evaluate the intraobserver repeatability and interobserver reproducibility of a newly developed dynamic real-time visualization 25 kHz swept-source optical coherence tomography (SS-OCT) based biometer (ZW-30, TowardPi Medical Technology Ltd, China) and compare its agreement with another SS-OCT based biometer (IOLMaster 700, Carl Zeiss Meditec AG, Jena, Germany).</p><p><strong>Methods: </strong>Eighty-two healthy right eyes were enrolled in this prospective observational study. Measurements were repeated for three times using the ZW-30 and IOLMaster 700 in a random order. Obtained parameters included axial length (AL), central corneal thickness (CCT), aqueous depth (AQD), anterior chamber depth (ACD), lens thickness (LT), mean keratometry (Km), astigmatism magnitude (AST), vector J<sub>0</sub>, vector J<sub>45</sub>, and corneal diameter (CD). The within-subject standard deviation (Sw), test-retest (TRT) variability, coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were adopted to assess the intraobserver repeatability and interobserver reproducibility. The double-angle plot was also used to display the distribution of AST. To estimate agreement, Bland-Altman plots were used.</p><p><strong>Results: </strong>For the intraobserver repeatability and interobserver reproducibility, the Sw, TRT and CoV for all parameters were low. Meanwhile, the ICC values were all close to 1.000, except for the J<sub>45</sub> (ICC = 0.887 for the intraobserver repeatability). The double-angle plot showed that the distribution of AST measured by these two devices was similar. For agreement, the Bland-Altman plots showed narrow 95% limits of agreements (LoAs) for AL, CCT, AQD, ACD, LT, Km AST, J<sub>0</sub>, J<sub>45</sub>, and CD (- 0.02 mm to 0.02 mm, - 7.49 μm to 8.08 μm, - 0.07 mm to 0.04 mm, - 0.07 mm to 0.04 mm, - 0.07 mm to 0.08 mm, - 0.16 D to 0.30 D, - 0.30 D to 0.29 D, - 0.16 D to 0.16 D, - 0.23 D to 0.13 D, and - 0.39 mm to 0.10 mm, respectively).</p><p><strong>Conclusions: </strong>The newly dynamic real-time visualization biometer exhibited excellent intraobserver repeatability and interobserver reproducibility. The two devices both based on the SS-OCT principle had similar ocular parameters measurement values and can be interchanged in clinical practice.</p>\",\"PeriodicalId\":12194,\"journal\":{\"name\":\"Eye and Vision\",\"volume\":\"11 1\",\"pages\":\"9\"},\"PeriodicalIF\":4.1000,\"publicationDate\":\"2024-03-04\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10910812/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Eye and Vision\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1186/s40662-024-00377-2\",\"RegionNum\":1,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Eye and Vision","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40662-024-00377-2","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
摘要
背景:目的:评估一种新开发的基于动态实时可视化25 kHz扫源光学相干断层扫描(SS-OCT)的生物测量仪(ZW-30,TowardPi Medical Technology Ltd,中国)的观察者内重复性和观察者间重复性,并将其与另一种基于SS-OCT的生物测量仪(IOLMaster 700,Carl Zeiss Meditec AG,Jena,德国)进行比较:方法:82 只健康右眼被纳入这项前瞻性观察研究。使用 ZW-30 和 IOLMaster 700 按随机顺序重复测量三次。获得的参数包括轴长(AL)、中央角膜厚度(CCT)、水深(AQD)、前房深度(ACD)、晶状体厚度(LT)、平均角膜度数(Km)、散光幅度(AST)、矢量 J0、矢量 J45 和角膜直径(CD)。采用受试者内标准偏差(Sw)、重复测试(TRT)变异性、变异系数(CoV)和类内相关系数(ICC)来评估观察者内重复性和观察者间重复性。双角图也用于显示 AST 的分布。为了估计一致性,采用了布兰-阿尔特曼图:结果:在观察者内重复性和观察者间重复性方面,所有参数的 Sw、TRT 和 CoV 均较低。同时,除 J45(观察者内重复性 ICC = 0.887)外,其他参数的 ICC 值均接近 1.000。双角图显示,这两种仪器测量的 AST 分布相似。在一致性方面,Bland-Altman 图显示 AL、CCT、AQD、ACD、LT、Km AST、J0、J45 和 CD 的 95% 一致性限 (LoAs) 很窄(- 0.02 mm 至 0.02 mm、- 7.49 μm 至 8.08 μm、- 0.07 mm 至 0.04 mm,- 0.07 mm 至 0.04 mm,- 0.07 mm 至 0.08 mm,- 0.16 D 至 0.30 D,- 0.30 D 至 0.29 D,- 0.16 D 至 0.16 D,- 0.23 D 至 0.13 D,以及- 0.39 mm 至 0.10 mm).结论:结论:新的动态实时可视生物测量仪具有极佳的观察者内重复性和观察者间重复性。两种基于 SS-OCT 原理的设备具有相似的眼部参数测量值,可在临床实践中互换使用。
Evaluation of a new dynamic real-time visualization 25 kHz swept-source optical coherence tomography based biometer.
Background: To evaluate the intraobserver repeatability and interobserver reproducibility of a newly developed dynamic real-time visualization 25 kHz swept-source optical coherence tomography (SS-OCT) based biometer (ZW-30, TowardPi Medical Technology Ltd, China) and compare its agreement with another SS-OCT based biometer (IOLMaster 700, Carl Zeiss Meditec AG, Jena, Germany).
Methods: Eighty-two healthy right eyes were enrolled in this prospective observational study. Measurements were repeated for three times using the ZW-30 and IOLMaster 700 in a random order. Obtained parameters included axial length (AL), central corneal thickness (CCT), aqueous depth (AQD), anterior chamber depth (ACD), lens thickness (LT), mean keratometry (Km), astigmatism magnitude (AST), vector J0, vector J45, and corneal diameter (CD). The within-subject standard deviation (Sw), test-retest (TRT) variability, coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were adopted to assess the intraobserver repeatability and interobserver reproducibility. The double-angle plot was also used to display the distribution of AST. To estimate agreement, Bland-Altman plots were used.
Results: For the intraobserver repeatability and interobserver reproducibility, the Sw, TRT and CoV for all parameters were low. Meanwhile, the ICC values were all close to 1.000, except for the J45 (ICC = 0.887 for the intraobserver repeatability). The double-angle plot showed that the distribution of AST measured by these two devices was similar. For agreement, the Bland-Altman plots showed narrow 95% limits of agreements (LoAs) for AL, CCT, AQD, ACD, LT, Km AST, J0, J45, and CD (- 0.02 mm to 0.02 mm, - 7.49 μm to 8.08 μm, - 0.07 mm to 0.04 mm, - 0.07 mm to 0.04 mm, - 0.07 mm to 0.08 mm, - 0.16 D to 0.30 D, - 0.30 D to 0.29 D, - 0.16 D to 0.16 D, - 0.23 D to 0.13 D, and - 0.39 mm to 0.10 mm, respectively).
Conclusions: The newly dynamic real-time visualization biometer exhibited excellent intraobserver repeatability and interobserver reproducibility. The two devices both based on the SS-OCT principle had similar ocular parameters measurement values and can be interchanged in clinical practice.
期刊介绍:
Eye and Vision is an open access, peer-reviewed journal for ophthalmologists and visual science specialists. It welcomes research articles, reviews, methodologies, commentaries, case reports, perspectives and short reports encompassing all aspects of eye and vision. Topics of interest include but are not limited to: current developments of theoretical, experimental and clinical investigations in ophthalmology, optometry and vision science which focus on novel and high-impact findings on central issues pertaining to biology, pathophysiology and etiology of eye diseases as well as advances in diagnostic techniques, surgical treatment, instrument updates, the latest drug findings, results of clinical trials and research findings. It aims to provide ophthalmologists and visual science specialists with the latest developments in theoretical, experimental and clinical investigations in eye and vision.