使用大口径 088 颅内入路支持导管输送大型颅内器械:125 例使用 TracStar LDP 的病例系列。

IF 4.5 1区 医学 Q1 NEUROIMAGING
Jessica K Campos, David A Zarrin, Benjamen M Meyer, Muhammad Waqas Khan, Fahad J Laghari, Jonathan C Collard de Beaufort, Gizal Amin, Narlin B Beaty, Matthew T Bender, Shuichi Suzuki, Geoffrey P Colby, Li-Mei Lin, Alexander L Coon
{"title":"使用大口径 088 颅内入路支持导管输送大型颅内器械:125 例使用 TracStar LDP 的病例系列。","authors":"Jessica K Campos, David A Zarrin, Benjamen M Meyer, Muhammad Waqas Khan, Fahad J Laghari, Jonathan C Collard de Beaufort, Gizal Amin, Narlin B Beaty, Matthew T Bender, Shuichi Suzuki, Geoffrey P Colby, Li-Mei Lin, Alexander L Coon","doi":"10.1136/jnis-2023-021054","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The delivery of neuroendovascular devices requires a robust proximal access platform. This demand has previously been met with a 6Fr long sheath (8Fr guide) that is placed in the proximal internal carotid artery (ICA) or vertebral artery segments. We share our experience with the first 0.088 inch 8Fr guide catheter designed for direct intracranial access.</p><p><strong>Methods: </strong>We retrospectively reviewed a prospectively maintained IRB-approved institutional database of the senior authors to identify all cases where the TracStar Large Distal Platform (LDP) was positioned within the intracranial vasculature, defined as within or distal to the petrous ICA, vertebral artery (V3) segments, or transverse sinus. Technical success was defined as safe placement of the TracStar LDP within or distal to the described distal vessel segments with subsequent complication-free device implantation.</p><p><strong>Results: </strong>Over the 41-month study period from January 2020 to June 2023, 125 consecutive cases were identified in whom the TracStar LDP was navigated into the intracranial vasculature for triaxial delivery of large devices, 0.027 inch microcatheter and greater, for aneurysm treatment (n=108, 86%), intracranial angioplasty/stenting (n=15, 12%), and venous sinus stenting (n=2, 1.6%). All cases used a direct select catheter technique for initial guide placement (no exchange). Posterior circulation treatments occurred in 14.4% (n=18) of cases. Technical success was achieved in 100% of cases. No vessel dissections occurred in any cases.</p><p><strong>Conclusion: </strong>The TracStar LDP is an 0.088 inch 8Fr guide catheter that can establish direct intracranial access with an acceptable safety profile. This can be achieved in a wide range of neurointerventional cases with a high rate of technical success.</p>","PeriodicalId":16411,"journal":{"name":"Journal of NeuroInterventional Surgery","volume":" ","pages":"1228-1231"},"PeriodicalIF":4.5000,"publicationDate":"2024-11-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Use of a large-bore 088 intracranial access support catheter for delivery of large intracranial devices: case series with the TracStar LDP in 125 cases.\",\"authors\":\"Jessica K Campos, David A Zarrin, Benjamen M Meyer, Muhammad Waqas Khan, Fahad J Laghari, Jonathan C Collard de Beaufort, Gizal Amin, Narlin B Beaty, Matthew T Bender, Shuichi Suzuki, Geoffrey P Colby, Li-Mei Lin, Alexander L Coon\",\"doi\":\"10.1136/jnis-2023-021054\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>The delivery of neuroendovascular devices requires a robust proximal access platform. This demand has previously been met with a 6Fr long sheath (8Fr guide) that is placed in the proximal internal carotid artery (ICA) or vertebral artery segments. We share our experience with the first 0.088 inch 8Fr guide catheter designed for direct intracranial access.</p><p><strong>Methods: </strong>We retrospectively reviewed a prospectively maintained IRB-approved institutional database of the senior authors to identify all cases where the TracStar Large Distal Platform (LDP) was positioned within the intracranial vasculature, defined as within or distal to the petrous ICA, vertebral artery (V3) segments, or transverse sinus. Technical success was defined as safe placement of the TracStar LDP within or distal to the described distal vessel segments with subsequent complication-free device implantation.</p><p><strong>Results: </strong>Over the 41-month study period from January 2020 to June 2023, 125 consecutive cases were identified in whom the TracStar LDP was navigated into the intracranial vasculature for triaxial delivery of large devices, 0.027 inch microcatheter and greater, for aneurysm treatment (n=108, 86%), intracranial angioplasty/stenting (n=15, 12%), and venous sinus stenting (n=2, 1.6%). All cases used a direct select catheter technique for initial guide placement (no exchange). Posterior circulation treatments occurred in 14.4% (n=18) of cases. Technical success was achieved in 100% of cases. No vessel dissections occurred in any cases.</p><p><strong>Conclusion: </strong>The TracStar LDP is an 0.088 inch 8Fr guide catheter that can establish direct intracranial access with an acceptable safety profile. This can be achieved in a wide range of neurointerventional cases with a high rate of technical success.</p>\",\"PeriodicalId\":16411,\"journal\":{\"name\":\"Journal of NeuroInterventional Surgery\",\"volume\":\" \",\"pages\":\"1228-1231\"},\"PeriodicalIF\":4.5000,\"publicationDate\":\"2024-11-22\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of NeuroInterventional Surgery\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1136/jnis-2023-021054\",\"RegionNum\":1,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"NEUROIMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of NeuroInterventional Surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1136/jnis-2023-021054","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"NEUROIMAGING","Score":null,"Total":0}
引用次数: 0

摘要

背景:神经内血管器械的输送需要一个强大的近端接入平台。以前,人们使用 6Fr 长的鞘管(8Fr 导引导管)来满足这一需求,这种鞘管被放置在颈内动脉 (ICA) 或椎动脉近端。我们将与大家分享首个专为颅内直接入路设计的 0.088 英寸 8Fr 导引导管的使用经验:我们回顾性地查看了资深作者的经 IRB 批准的前瞻性机构数据库,以确定将 TracStar 大远端平台 (LDP) 置入颅内血管的所有病例,颅内血管的定义是指瓣状 ICA、椎动脉 (V3) 段或横窦内或远端。技术成功的定义是将 TracStar LDP 安全置入所述远端血管内或远端,且随后的设备植入无并发症:结果:在 2020 年 1 月至 2023 年 6 月的 41 个月研究期间,确定了 125 个连续病例,在这些病例中,TracStar LDP 被导航至颅内血管,用于三轴输送大型装置(0.027 英寸微导管及以上),以治疗动脉瘤(108 例,占 86%)、颅内血管成形术/支架植入术(15 例,占 12%)和静脉窦支架植入术(2 例,占 1.6%)。所有病例均采用直接选择导管技术进行初始导引置入(无交换)。14.4%(18 例)的病例进行了后循环治疗。100%的病例都取得了技术成功。所有病例均未发生血管离断:TracStar LDP 是一种 0.088 英寸的 8Fr 导引导管,可以建立直接的颅内入路,其安全性是可以接受的。该导管适用于多种神经介入病例,技术成功率高。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Use of a large-bore 088 intracranial access support catheter for delivery of large intracranial devices: case series with the TracStar LDP in 125 cases.

Background: The delivery of neuroendovascular devices requires a robust proximal access platform. This demand has previously been met with a 6Fr long sheath (8Fr guide) that is placed in the proximal internal carotid artery (ICA) or vertebral artery segments. We share our experience with the first 0.088 inch 8Fr guide catheter designed for direct intracranial access.

Methods: We retrospectively reviewed a prospectively maintained IRB-approved institutional database of the senior authors to identify all cases where the TracStar Large Distal Platform (LDP) was positioned within the intracranial vasculature, defined as within or distal to the petrous ICA, vertebral artery (V3) segments, or transverse sinus. Technical success was defined as safe placement of the TracStar LDP within or distal to the described distal vessel segments with subsequent complication-free device implantation.

Results: Over the 41-month study period from January 2020 to June 2023, 125 consecutive cases were identified in whom the TracStar LDP was navigated into the intracranial vasculature for triaxial delivery of large devices, 0.027 inch microcatheter and greater, for aneurysm treatment (n=108, 86%), intracranial angioplasty/stenting (n=15, 12%), and venous sinus stenting (n=2, 1.6%). All cases used a direct select catheter technique for initial guide placement (no exchange). Posterior circulation treatments occurred in 14.4% (n=18) of cases. Technical success was achieved in 100% of cases. No vessel dissections occurred in any cases.

Conclusion: The TracStar LDP is an 0.088 inch 8Fr guide catheter that can establish direct intracranial access with an acceptable safety profile. This can be achieved in a wide range of neurointerventional cases with a high rate of technical success.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
9.50
自引率
14.60%
发文量
291
审稿时长
4-8 weeks
期刊介绍: The Journal of NeuroInterventional Surgery (JNIS) is a leading peer review journal for scientific research and literature pertaining to the field of neurointerventional surgery. The journal launch follows growing professional interest in neurointerventional techniques for the treatment of a range of neurological and vascular problems including stroke, aneurysms, brain tumors, and spinal compression.The journal is owned by SNIS and is also the official journal of the Interventional Chapter of the Australian and New Zealand Society of Neuroradiology (ANZSNR), the Canadian Interventional Neuro Group, the Hong Kong Neurological Society (HKNS) and the Neuroradiological Society of Taiwan.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信