Komal Arora, Kerrie Plant, Pam Dawson, Robert Cowan
{"title":"降低电刺激频率对 Nucleus® 人工耳蜗植入者听力表现的影响","authors":"Komal Arora, Kerrie Plant, Pam Dawson, Robert Cowan","doi":"10.1080/14992027.2024.2314620","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To evaluate whether a 500 pulses per second per channel (pps/ch) rate would provide non-inferior hearing performance compared to the 900 pps/ch rate in the Advanced Combination Encoder (ACE™) sound coding strategy.</p><p><strong>Design: </strong>A repeated measures single-subject design was employed, wherein each subject served as their own control. All except one subject used 900 pps/ch at enrolment. After three weeks of using the alternative rate program, both programs were loaded into the sound processor for two more weeks of take-home use. Subjective performance, preference, words in quiet, sentences in babble, music quality, and fundamental frequency (F0) discrimination were assessed using a balanced design.</p><p><strong>Study sample: </strong>Data from 18 subjects were analysed, with complete datasets available for 17 subjects.</p><p><strong>Results: </strong>Non-inferior performance on all clinical measures was shown for the lower rate program. Subjects' preference ratings were comparable for the programs, with 53% reporting no difference overall. When a preference was expressed, the 900 pps/ch condition was preferred more often.</p><p><strong>Conclusion: </strong>Reducing the stimulation rate from 900 pps/ch to 500 pps/ch did not compromise the hearing outcomes evaluated in this study. A lower pulse rate in future cochlear implants could reduce power consumption, allowing for smaller batteries and processors.</p>","PeriodicalId":13759,"journal":{"name":"International Journal of Audiology","volume":" ","pages":"173-182"},"PeriodicalIF":1.8000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Effect of reducing electrical stimulation rate on hearing performance of Nucleus<sup>®</sup> cochlear implant recipients.\",\"authors\":\"Komal Arora, Kerrie Plant, Pam Dawson, Robert Cowan\",\"doi\":\"10.1080/14992027.2024.2314620\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To evaluate whether a 500 pulses per second per channel (pps/ch) rate would provide non-inferior hearing performance compared to the 900 pps/ch rate in the Advanced Combination Encoder (ACE™) sound coding strategy.</p><p><strong>Design: </strong>A repeated measures single-subject design was employed, wherein each subject served as their own control. All except one subject used 900 pps/ch at enrolment. After three weeks of using the alternative rate program, both programs were loaded into the sound processor for two more weeks of take-home use. Subjective performance, preference, words in quiet, sentences in babble, music quality, and fundamental frequency (F0) discrimination were assessed using a balanced design.</p><p><strong>Study sample: </strong>Data from 18 subjects were analysed, with complete datasets available for 17 subjects.</p><p><strong>Results: </strong>Non-inferior performance on all clinical measures was shown for the lower rate program. Subjects' preference ratings were comparable for the programs, with 53% reporting no difference overall. When a preference was expressed, the 900 pps/ch condition was preferred more often.</p><p><strong>Conclusion: </strong>Reducing the stimulation rate from 900 pps/ch to 500 pps/ch did not compromise the hearing outcomes evaluated in this study. A lower pulse rate in future cochlear implants could reduce power consumption, allowing for smaller batteries and processors.</p>\",\"PeriodicalId\":13759,\"journal\":{\"name\":\"International Journal of Audiology\",\"volume\":\" \",\"pages\":\"173-182\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2025-02-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International Journal of Audiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1080/14992027.2024.2314620\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/2/29 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"AUDIOLOGY & SPEECH-LANGUAGE PATHOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Audiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/14992027.2024.2314620","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/29 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"AUDIOLOGY & SPEECH-LANGUAGE PATHOLOGY","Score":null,"Total":0}
Effect of reducing electrical stimulation rate on hearing performance of Nucleus® cochlear implant recipients.
Objective: To evaluate whether a 500 pulses per second per channel (pps/ch) rate would provide non-inferior hearing performance compared to the 900 pps/ch rate in the Advanced Combination Encoder (ACE™) sound coding strategy.
Design: A repeated measures single-subject design was employed, wherein each subject served as their own control. All except one subject used 900 pps/ch at enrolment. After three weeks of using the alternative rate program, both programs were loaded into the sound processor for two more weeks of take-home use. Subjective performance, preference, words in quiet, sentences in babble, music quality, and fundamental frequency (F0) discrimination were assessed using a balanced design.
Study sample: Data from 18 subjects were analysed, with complete datasets available for 17 subjects.
Results: Non-inferior performance on all clinical measures was shown for the lower rate program. Subjects' preference ratings were comparable for the programs, with 53% reporting no difference overall. When a preference was expressed, the 900 pps/ch condition was preferred more often.
Conclusion: Reducing the stimulation rate from 900 pps/ch to 500 pps/ch did not compromise the hearing outcomes evaluated in this study. A lower pulse rate in future cochlear implants could reduce power consumption, allowing for smaller batteries and processors.
期刊介绍:
International Journal of Audiology is committed to furthering development of a scientifically robust evidence base for audiology. The journal is published by the British Society of Audiology, the International Society of Audiology and the Nordic Audiological Society.