Alejandro D. Fernandez Schrunder;Yu-Kai Huang;Saul Rodriguez;Ana Rusu
{"title":"基于中频采样和伪双路径 SC 带通 ΔΣ ADC 的 EIM 生物阻抗光谱接口。","authors":"Alejandro D. Fernandez Schrunder;Yu-Kai Huang;Saul Rodriguez;Ana Rusu","doi":"10.1109/TBCAS.2024.3370399","DOIUrl":null,"url":null,"abstract":"This paper presents a low-noise bioimpedance (bio-Z) spectroscopy interface for electrical impedance myography (EIM) over the 1 kHz to 2 MHz frequency range. The proposed interface employs a sinusoidal signal generator based on direct-digital-synthesis (DDS) to improve the accuracy of the bio-Z reading, and a quadrature low-intermediate frequency (IF) readout to achieve a good noise-to-power efficiency and the required data throughput to detect muscle contractions. The readout is able to measure baseline and time-varying bio-Z by employing robust and power-efficient low-gain IAs and sixth-order single-bit bandpass (BP) \n<inline-formula><tex-math>$\\mathbf{\\Delta\\Sigma}$</tex-math></inline-formula>\n ADCs. The proposed bio-Z spectroscopy interface is implemented in a 180 nm CMOS process, consumes 344.3 \n<inline-formula><tex-math>$-$</tex-math></inline-formula>\n 479.3 \n<inline-formula><tex-math>$\\mathbf{\\mu}$</tex-math></inline-formula>\nW, and occupies 5.4 mm\n<sup>2</sup>\n area. Measurement results show 0.7 m\n<inline-formula><tex-math>$\\mathbf{\\Omega}$</tex-math></inline-formula>\n/\n<inline-formula><tex-math>$\\sqrt{\\text{Hz}}$</tex-math></inline-formula>\n sensitivity at 15.625 kHz, 105.8 dB SNR within 4 Hz bandwidth, and a 146.5 dB figure-of-merit. Additionally, recording of EIM in time and frequency domain during contractions of the bicep brachii muscle demonstrates the potential of the proposed bio-Z interface for wearable EIM systems.","PeriodicalId":94031,"journal":{"name":"IEEE transactions on biomedical circuits and systems","volume":"18 5","pages":"1001-1013"},"PeriodicalIF":0.0000,"publicationDate":"2024-02-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=10445383","citationCount":"0","resultStr":"{\"title\":\"A Bioimpedance Spectroscopy Interface for EIM Based on IF-Sampling and Pseudo 2-Path SC Bandpass $\\\\Delta\\\\Sigma$ ADC\",\"authors\":\"Alejandro D. Fernandez Schrunder;Yu-Kai Huang;Saul Rodriguez;Ana Rusu\",\"doi\":\"10.1109/TBCAS.2024.3370399\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This paper presents a low-noise bioimpedance (bio-Z) spectroscopy interface for electrical impedance myography (EIM) over the 1 kHz to 2 MHz frequency range. The proposed interface employs a sinusoidal signal generator based on direct-digital-synthesis (DDS) to improve the accuracy of the bio-Z reading, and a quadrature low-intermediate frequency (IF) readout to achieve a good noise-to-power efficiency and the required data throughput to detect muscle contractions. The readout is able to measure baseline and time-varying bio-Z by employing robust and power-efficient low-gain IAs and sixth-order single-bit bandpass (BP) \\n<inline-formula><tex-math>$\\\\mathbf{\\\\Delta\\\\Sigma}$</tex-math></inline-formula>\\n ADCs. The proposed bio-Z spectroscopy interface is implemented in a 180 nm CMOS process, consumes 344.3 \\n<inline-formula><tex-math>$-$</tex-math></inline-formula>\\n 479.3 \\n<inline-formula><tex-math>$\\\\mathbf{\\\\mu}$</tex-math></inline-formula>\\nW, and occupies 5.4 mm\\n<sup>2</sup>\\n area. Measurement results show 0.7 m\\n<inline-formula><tex-math>$\\\\mathbf{\\\\Omega}$</tex-math></inline-formula>\\n/\\n<inline-formula><tex-math>$\\\\sqrt{\\\\text{Hz}}$</tex-math></inline-formula>\\n sensitivity at 15.625 kHz, 105.8 dB SNR within 4 Hz bandwidth, and a 146.5 dB figure-of-merit. Additionally, recording of EIM in time and frequency domain during contractions of the bicep brachii muscle demonstrates the potential of the proposed bio-Z interface for wearable EIM systems.\",\"PeriodicalId\":94031,\"journal\":{\"name\":\"IEEE transactions on biomedical circuits and systems\",\"volume\":\"18 5\",\"pages\":\"1001-1013\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-02-26\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=10445383\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"IEEE transactions on biomedical circuits and systems\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://ieeexplore.ieee.org/document/10445383/\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE transactions on biomedical circuits and systems","FirstCategoryId":"1085","ListUrlMain":"https://ieeexplore.ieee.org/document/10445383/","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
A Bioimpedance Spectroscopy Interface for EIM Based on IF-Sampling and Pseudo 2-Path SC Bandpass $\Delta\Sigma$ ADC
This paper presents a low-noise bioimpedance (bio-Z) spectroscopy interface for electrical impedance myography (EIM) over the 1 kHz to 2 MHz frequency range. The proposed interface employs a sinusoidal signal generator based on direct-digital-synthesis (DDS) to improve the accuracy of the bio-Z reading, and a quadrature low-intermediate frequency (IF) readout to achieve a good noise-to-power efficiency and the required data throughput to detect muscle contractions. The readout is able to measure baseline and time-varying bio-Z by employing robust and power-efficient low-gain IAs and sixth-order single-bit bandpass (BP)
$\mathbf{\Delta\Sigma}$
ADCs. The proposed bio-Z spectroscopy interface is implemented in a 180 nm CMOS process, consumes 344.3
$-$
479.3
$\mathbf{\mu}$
W, and occupies 5.4 mm
2
area. Measurement results show 0.7 m
$\mathbf{\Omega}$
/
$\sqrt{\text{Hz}}$