窄口种植体口内骨折是否与疲劳机制有关?对从一名患者身上取回的两个失败种植体固定装置的彻底取回分析

IF 1.7 Q3 DENTISTRY, ORAL SURGERY & MEDICINE
Youssef S. Al Jabbari , Raymond Fournelle , Arif Sufyan , Spiros Zinelis
{"title":"窄口种植体口内骨折是否与疲劳机制有关?对从一名患者身上取回的两个失败种植体固定装置的彻底取回分析","authors":"Youssef S. Al Jabbari ,&nbsp;Raymond Fournelle ,&nbsp;Arif Sufyan ,&nbsp;Spiros Zinelis","doi":"10.1016/j.sdentj.2024.02.006","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><p>This study aimed to perform a thorough failure analysis of two fractured narrow dental implants after medium-term in vivo use.</p></div><div><h3>Materials and methods</h3><p>The top parts of two fractured Narrow Dental Implant (NDI) fixtures were retrieved from two different locations at two different times from the same patient. The NDI-specimen-1 was 12-months in service while the NDI-specimen-2 was 17-months in service. In both cases, the top parts of the fractured NDI fixtures that were attached to prosthetic components were retrieved and subjected to thorough, non-destructive and destructive testing.</p></div><div><h3>Results</h3><p>Light Microscopy <strong>(</strong>LM) and Scanning Electron Microscopy (SEM) revealed that both the retrieved fractured NDIs failed because of fatigue, characterized by beach and ratchet marks. Macroscopic examination revealed that fatigue cracks initiated at the internal thread surfaces of the implants and propagated around them until final fracture. Both samples fractured near the end of the retaining screw and followed the root of the internal thread. Optical and SEM analyses revealed a uniform distribution of irregularly shaped grains (diameter = 2 to 5 μm). X -ray Energy Dispersive Spectroscopy (EDS) analysis showed that the NDI-specimen-1 was made using Ti-14%Zr with a Vickers Hardens (HV) of 288 ± 5.</p></div><div><h3>Conclusion</h3><p>Since the fracture occurred by a fatigue; thus, an increase in fatigue resistance will be beneficial for the longevity of NDI.</p></div>","PeriodicalId":47246,"journal":{"name":"Saudi Dental Journal","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2024-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S1013905224000580/pdfft?md5=2e7a751f955d5a7b36eb513d57c5ae80&pid=1-s2.0-S1013905224000580-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Is fatigue mechanism implicated in intraoral fracture of narrow dental implants? A thorough retrieval analysis of two failed implant fixtures retrieved from a single patient\",\"authors\":\"Youssef S. Al Jabbari ,&nbsp;Raymond Fournelle ,&nbsp;Arif Sufyan ,&nbsp;Spiros Zinelis\",\"doi\":\"10.1016/j.sdentj.2024.02.006\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Purpose</h3><p>This study aimed to perform a thorough failure analysis of two fractured narrow dental implants after medium-term in vivo use.</p></div><div><h3>Materials and methods</h3><p>The top parts of two fractured Narrow Dental Implant (NDI) fixtures were retrieved from two different locations at two different times from the same patient. The NDI-specimen-1 was 12-months in service while the NDI-specimen-2 was 17-months in service. In both cases, the top parts of the fractured NDI fixtures that were attached to prosthetic components were retrieved and subjected to thorough, non-destructive and destructive testing.</p></div><div><h3>Results</h3><p>Light Microscopy <strong>(</strong>LM) and Scanning Electron Microscopy (SEM) revealed that both the retrieved fractured NDIs failed because of fatigue, characterized by beach and ratchet marks. Macroscopic examination revealed that fatigue cracks initiated at the internal thread surfaces of the implants and propagated around them until final fracture. Both samples fractured near the end of the retaining screw and followed the root of the internal thread. Optical and SEM analyses revealed a uniform distribution of irregularly shaped grains (diameter = 2 to 5 μm). X -ray Energy Dispersive Spectroscopy (EDS) analysis showed that the NDI-specimen-1 was made using Ti-14%Zr with a Vickers Hardens (HV) of 288 ± 5.</p></div><div><h3>Conclusion</h3><p>Since the fracture occurred by a fatigue; thus, an increase in fatigue resistance will be beneficial for the longevity of NDI.</p></div>\",\"PeriodicalId\":47246,\"journal\":{\"name\":\"Saudi Dental Journal\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2024-05-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S1013905224000580/pdfft?md5=2e7a751f955d5a7b36eb513d57c5ae80&pid=1-s2.0-S1013905224000580-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Saudi Dental Journal\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1013905224000580\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Saudi Dental Journal","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1013905224000580","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

摘要

材料和方法在同一患者的两个不同时间,从两个不同位置取回了两个断裂的窄牙种植体(Narrow Dental Implant,NDI)的顶部。NDI 样本-1 使用了 12 个月,而 NDI 样本-2 使用了 17 个月。结果光学显微镜(LM)和扫描电子显微镜(SEM)显示,取回的两个断裂的 NDI 都因疲劳而失效,其特征是有沙滩和棘轮痕迹。宏观检查显示,疲劳裂纹始于植入物的内螺纹表面,并在其周围扩展,直至最终断裂。两个样品都是在固定螺钉末端附近断裂,并沿着内螺纹根部断裂。光学和扫描电镜分析表明,不规则形状的晶粒(直径 = 2 至 5 μm)分布均匀。X 射线能量色散光谱(EDS)分析表明,NDI-试样-1 是用 Ti-14%Zr 制成的,维氏硬度(HV)为 288 ± 5。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Is fatigue mechanism implicated in intraoral fracture of narrow dental implants? A thorough retrieval analysis of two failed implant fixtures retrieved from a single patient

Purpose

This study aimed to perform a thorough failure analysis of two fractured narrow dental implants after medium-term in vivo use.

Materials and methods

The top parts of two fractured Narrow Dental Implant (NDI) fixtures were retrieved from two different locations at two different times from the same patient. The NDI-specimen-1 was 12-months in service while the NDI-specimen-2 was 17-months in service. In both cases, the top parts of the fractured NDI fixtures that were attached to prosthetic components were retrieved and subjected to thorough, non-destructive and destructive testing.

Results

Light Microscopy (LM) and Scanning Electron Microscopy (SEM) revealed that both the retrieved fractured NDIs failed because of fatigue, characterized by beach and ratchet marks. Macroscopic examination revealed that fatigue cracks initiated at the internal thread surfaces of the implants and propagated around them until final fracture. Both samples fractured near the end of the retaining screw and followed the root of the internal thread. Optical and SEM analyses revealed a uniform distribution of irregularly shaped grains (diameter = 2 to 5 μm). X -ray Energy Dispersive Spectroscopy (EDS) analysis showed that the NDI-specimen-1 was made using Ti-14%Zr with a Vickers Hardens (HV) of 288 ± 5.

Conclusion

Since the fracture occurred by a fatigue; thus, an increase in fatigue resistance will be beneficial for the longevity of NDI.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Saudi Dental Journal
Saudi Dental Journal DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
3.60
自引率
0.00%
发文量
86
审稿时长
22 weeks
期刊介绍: Saudi Dental Journal is an English language, peer-reviewed scholarly publication in the area of dentistry. Saudi Dental Journal publishes original research and reviews on, but not limited to: • dental disease • clinical trials • dental equipment • new and experimental techniques • epidemiology and oral health • restorative dentistry • periodontology • endodontology • prosthodontics • paediatric dentistry • orthodontics and dental education Saudi Dental Journal is the official publication of the Saudi Dental Society and is published by King Saud University in collaboration with Elsevier and is edited by an international group of eminent researchers.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信