Oscar M Cañete, Mouhamad El-Haj-Ali, Michal Fereczkowski
{"title":"两种用于测量正常听力成年人失真产物耳声发射的临床设备的比较。","authors":"Oscar M Cañete, Mouhamad El-Haj-Ali, Michal Fereczkowski","doi":"10.7874/jao.2023.00269","DOIUrl":null,"url":null,"abstract":"<p><strong>Background and objectives: </strong>Otoacoustic emissions (OAEs) are low-intensity sounds generated by the cochlea and associated with the function of the outer hair cells. Since OAE measurements do not require active participation of a listener, OAEs are considered an objective measure of cochlear function. While distortion-product OAEs (DPOAEs) are commonly used in clinical practice, limited information is available on the performance of various clinical devices. This study compared two commercial clinical devices, Titan and Eclipse from Interacoustics, and collected normative data for DPOAEs in people with normal hearing. The data collection and analysis were focused on signal and noise amplitudes as well as signal-to-noise ratios (SNRs).</p><p><strong>Subjects and methods: </strong>Sixty-three participants with normal hearing (age 23.2±2.0 years) were included in the study. DPOAEs were measured at 31 frequencies ranging from 500 to 10,000 Hz.</p><p><strong>Results: </strong>DPOAE amplitude differed across frequencies. Additionally, a high number of unreliable responses were observed at the edge frequencies (<828 Hz and >6,072 Hz). There were no significant differences between the two devices in terms of DPOAE amplitudes, but the recorded-noise levels (amplitude) differed significantly. SNRs of the recorded OAEs were frequency dependent (higher frequencies showed larger SNRs), and significant differences in terms of SNR were found between the two devices.</p><p><strong>Conclusions: </strong>Despite the above-mentioned differences, the OAEs recorded with both devices met the pass criteria for the SNR (≥6 dB) consistently across frequencies, and thus the differences do not compromise the test outcomes. The frequency dependence of the OAE amplitudes and the corresponding SNRs may be relevant for clinical practice.</p>","PeriodicalId":44886,"journal":{"name":"Journal of Audiology and Otology","volume":null,"pages":null},"PeriodicalIF":1.1000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11065547/pdf/","citationCount":"0","resultStr":"{\"title\":\"Comparison of Two Clinical Devices for the Measurement of Distortion Product Otoacoustic Emissions in Normal-Hearing Adults.\",\"authors\":\"Oscar M Cañete, Mouhamad El-Haj-Ali, Michal Fereczkowski\",\"doi\":\"10.7874/jao.2023.00269\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background and objectives: </strong>Otoacoustic emissions (OAEs) are low-intensity sounds generated by the cochlea and associated with the function of the outer hair cells. Since OAE measurements do not require active participation of a listener, OAEs are considered an objective measure of cochlear function. While distortion-product OAEs (DPOAEs) are commonly used in clinical practice, limited information is available on the performance of various clinical devices. This study compared two commercial clinical devices, Titan and Eclipse from Interacoustics, and collected normative data for DPOAEs in people with normal hearing. The data collection and analysis were focused on signal and noise amplitudes as well as signal-to-noise ratios (SNRs).</p><p><strong>Subjects and methods: </strong>Sixty-three participants with normal hearing (age 23.2±2.0 years) were included in the study. DPOAEs were measured at 31 frequencies ranging from 500 to 10,000 Hz.</p><p><strong>Results: </strong>DPOAE amplitude differed across frequencies. Additionally, a high number of unreliable responses were observed at the edge frequencies (<828 Hz and >6,072 Hz). There were no significant differences between the two devices in terms of DPOAE amplitudes, but the recorded-noise levels (amplitude) differed significantly. SNRs of the recorded OAEs were frequency dependent (higher frequencies showed larger SNRs), and significant differences in terms of SNR were found between the two devices.</p><p><strong>Conclusions: </strong>Despite the above-mentioned differences, the OAEs recorded with both devices met the pass criteria for the SNR (≥6 dB) consistently across frequencies, and thus the differences do not compromise the test outcomes. The frequency dependence of the OAE amplitudes and the corresponding SNRs may be relevant for clinical practice.</p>\",\"PeriodicalId\":44886,\"journal\":{\"name\":\"Journal of Audiology and Otology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.1000,\"publicationDate\":\"2024-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11065547/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Audiology and Otology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.7874/jao.2023.00269\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/2/22 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"OTORHINOLARYNGOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Audiology and Otology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.7874/jao.2023.00269","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/22 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"OTORHINOLARYNGOLOGY","Score":null,"Total":0}
Comparison of Two Clinical Devices for the Measurement of Distortion Product Otoacoustic Emissions in Normal-Hearing Adults.
Background and objectives: Otoacoustic emissions (OAEs) are low-intensity sounds generated by the cochlea and associated with the function of the outer hair cells. Since OAE measurements do not require active participation of a listener, OAEs are considered an objective measure of cochlear function. While distortion-product OAEs (DPOAEs) are commonly used in clinical practice, limited information is available on the performance of various clinical devices. This study compared two commercial clinical devices, Titan and Eclipse from Interacoustics, and collected normative data for DPOAEs in people with normal hearing. The data collection and analysis were focused on signal and noise amplitudes as well as signal-to-noise ratios (SNRs).
Subjects and methods: Sixty-three participants with normal hearing (age 23.2±2.0 years) were included in the study. DPOAEs were measured at 31 frequencies ranging from 500 to 10,000 Hz.
Results: DPOAE amplitude differed across frequencies. Additionally, a high number of unreliable responses were observed at the edge frequencies (<828 Hz and >6,072 Hz). There were no significant differences between the two devices in terms of DPOAE amplitudes, but the recorded-noise levels (amplitude) differed significantly. SNRs of the recorded OAEs were frequency dependent (higher frequencies showed larger SNRs), and significant differences in terms of SNR were found between the two devices.
Conclusions: Despite the above-mentioned differences, the OAEs recorded with both devices met the pass criteria for the SNR (≥6 dB) consistently across frequencies, and thus the differences do not compromise the test outcomes. The frequency dependence of the OAE amplitudes and the corresponding SNRs may be relevant for clinical practice.
期刊介绍:
Journal of Audiology and Otology (JAO) (formerly known as Korean Journal of Audiology) aims to publish the most advanced findings for all aspects of the auditory and vestibular system and diseases of the ear using state-of-the-art techniques and analyses. The journal covers recent trends related to the topics of audiology, otology, and neurotology conducted by professionals, with the goal of providing better possible treatment to people of all ages, from infants to the elderly, who suffer from auditory and/or vestibular disorders and thus, improving their quality of life. This journal encourages the submission of review papers about current professional issues, research papers presenting a scientific base and clinical application, and case papers with unique reports or clinical trials. We also invite letters to the editor and papers related to the manufacture and distribution of medical devices. This journal provides integrated views from otologists, audiologists, and other healthcare practitioners, offering readers high quality scientific and clinical information. This peer-reviewed and open access journal has been the official journal of the Korean Audiological Society since 1997 and of both the Korean Audiological Society and the Korean Otological Society since 2017. It is published in English four times a year in January, April, July, and October.