{"title":"KOROT P3 精确自动听诊式血压测量仪的准确性,适用于手臂特大的专业人士。","authors":"Claudio Fania, Antonella Giletto, Alice Niero, Paolo Palatini","doi":"10.1097/MBP.0000000000000696","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To determine the accuracy of the KOROT P3 Accurate (previously InBody BPBIO480KV) monitor, an automated auscultatory blood pressure (BP) measuring device developed for professional use, in people with extra-large arms according to the ISO81060-2 2018 protocol.</p><p><strong>Methods: </strong>The KOROT P3 Accurate was tested in 37 subjects with upper-arm circumference ranging from >42 to 53 cm using a mercury sphygmomanometer coupled to a 20 × 40 cm tronco-conical cuff as the reference standard.</p><p><strong>Results: </strong>The mean BP difference between the device and the observers' reference measurements was 1.2 ± 2.0 mmHg for systolic BP and 1.0 ± 2.0 mmHg for diastolic BP. These data were in agreement with criterion 1 of the protocol standard requirements (≤5 ± 8 mmHg). Also criterion 2 was satisfied being the standard deviations ± 1.7 mmHg for systolic BP and ± 1.6 for diastolic BP, well below the maximum values required by the protocol (±6.84/6.87 mmHg). Scatterplots of device-reference systolic and diastolic BP differences showed similar accuracy across the range of participants' BP, arm circumference and upper-arm slant angle.</p><p><strong>Conclusions: </strong>These data show that the KOROT P3 Accurate monitor satisfied the ISO 81060-2:2018 standard requirements in a special population of people with extra-large arms ranging from >42 to 53 cm.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Accuracy of the KOROT P3 Accurate automated auscultatory blood pressure measuring device for professional use in people with extra-large arms.\",\"authors\":\"Claudio Fania, Antonella Giletto, Alice Niero, Paolo Palatini\",\"doi\":\"10.1097/MBP.0000000000000696\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To determine the accuracy of the KOROT P3 Accurate (previously InBody BPBIO480KV) monitor, an automated auscultatory blood pressure (BP) measuring device developed for professional use, in people with extra-large arms according to the ISO81060-2 2018 protocol.</p><p><strong>Methods: </strong>The KOROT P3 Accurate was tested in 37 subjects with upper-arm circumference ranging from >42 to 53 cm using a mercury sphygmomanometer coupled to a 20 × 40 cm tronco-conical cuff as the reference standard.</p><p><strong>Results: </strong>The mean BP difference between the device and the observers' reference measurements was 1.2 ± 2.0 mmHg for systolic BP and 1.0 ± 2.0 mmHg for diastolic BP. These data were in agreement with criterion 1 of the protocol standard requirements (≤5 ± 8 mmHg). Also criterion 2 was satisfied being the standard deviations ± 1.7 mmHg for systolic BP and ± 1.6 for diastolic BP, well below the maximum values required by the protocol (±6.84/6.87 mmHg). Scatterplots of device-reference systolic and diastolic BP differences showed similar accuracy across the range of participants' BP, arm circumference and upper-arm slant angle.</p><p><strong>Conclusions: </strong>These data show that the KOROT P3 Accurate monitor satisfied the ISO 81060-2:2018 standard requirements in a special population of people with extra-large arms ranging from >42 to 53 cm.</p>\",\"PeriodicalId\":8950,\"journal\":{\"name\":\"Blood Pressure Monitoring\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2024-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Blood Pressure Monitoring\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MBP.0000000000000696\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/2/23 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"PERIPHERAL VASCULAR DISEASE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000696","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/23 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

摘要

目的根据 ISO81060-2 2018 协议,确定 KOROT P3 Accurate(前身为 InBody BPBIO480KV)监测仪(一款专为专业用途开发的自动听诊式血压(BP)测量设备)在超大臂人群中的准确性:使用水银血压计和 20 × 40 厘米锥形袖带作为参考标准,对 37 名上臂周长大于 42 厘米至 53 厘米的受试者进行了 KOROT P3 Accurate 测试:设备与观察者参考测量值之间的平均血压差为:收缩压 1.2 ± 2.0 mmHg,舒张压 1.0 ± 2.0 mmHg。这些数据符合方案标准要求的标准 1(≤5 ± 8 mmHg)。此外,标准 2 也符合,收缩压的标准偏差为 ± 1.7 mmHg,舒张压的标准偏差为 ± 1.6 mmHg,远低于方案要求的最大值(±6.84/6.87 mmHg)。设备参考收缩压和舒张压差异的散点图显示,在参与者的血压、臂围和上臂斜角范围内,其准确性相似:这些数据表明,KOROT P3 Accurate 监护仪符合 ISO 81060-2:2018 标准的要求,适用于臂围大于 42 厘米至 53 厘米的特大臂人群。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Accuracy of the KOROT P3 Accurate automated auscultatory blood pressure measuring device for professional use in people with extra-large arms.

Objective: To determine the accuracy of the KOROT P3 Accurate (previously InBody BPBIO480KV) monitor, an automated auscultatory blood pressure (BP) measuring device developed for professional use, in people with extra-large arms according to the ISO81060-2 2018 protocol.

Methods: The KOROT P3 Accurate was tested in 37 subjects with upper-arm circumference ranging from >42 to 53 cm using a mercury sphygmomanometer coupled to a 20 × 40 cm tronco-conical cuff as the reference standard.

Results: The mean BP difference between the device and the observers' reference measurements was 1.2 ± 2.0 mmHg for systolic BP and 1.0 ± 2.0 mmHg for diastolic BP. These data were in agreement with criterion 1 of the protocol standard requirements (≤5 ± 8 mmHg). Also criterion 2 was satisfied being the standard deviations ± 1.7 mmHg for systolic BP and ± 1.6 for diastolic BP, well below the maximum values required by the protocol (±6.84/6.87 mmHg). Scatterplots of device-reference systolic and diastolic BP differences showed similar accuracy across the range of participants' BP, arm circumference and upper-arm slant angle.

Conclusions: These data show that the KOROT P3 Accurate monitor satisfied the ISO 81060-2:2018 standard requirements in a special population of people with extra-large arms ranging from >42 to 53 cm.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Blood Pressure Monitoring
Blood Pressure Monitoring 医学-外周血管病
CiteScore
2.00
自引率
7.70%
发文量
110
审稿时长
>12 weeks
期刊介绍: Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research. This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine. Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信