比较 Thames Medical CAT+ 多普勒和 SunTech Vet 20 血压计设备对清醒猫咪的无创血压测量。

IF 1.9 2区 农林科学 Q2 VETERINARY SCIENCES
Clara Casas, Charlotte Dye
{"title":"比较 Thames Medical CAT+ 多普勒和 SunTech Vet 20 血压计设备对清醒猫咪的无创血压测量。","authors":"Clara Casas, Charlotte Dye","doi":"10.1177/1098612X231216350","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>A comparative assessment of systolic blood pressure (BP) measurement agreement and precision in two commonly used non-invasive BP devices was carried out in conscious cats.</p><p><strong>Methods: </strong>Systolic BP measurements were obtained from 50 conscious cats as part of their clinical investigations. All measurements were taken by the same operator and were performed according to the American College of Veterinary Internal Medicine (ACVIM) consensus guidelines. The same cuff location and cuff size were used for paired measurements. The order of device use was randomised, and an arousal score was assigned during each procedure. Precision was assessed using standard deviation and coefficient of variance comparisons. Agreement was assessed using graphical and statistical comparisons of derived 'delta mean' and 'mean of means' data.</p><p><strong>Results: </strong>A total of 50 cats aged between 3 months and 15 years were enrolled (29 domestic shorthair, seven domestic longhair, three British Shorthair, two Ragdoll, two Sphynx, two Persian, one Siamese, one Burmese, one Russian Blue, one Maine Coon and one Oriental) with a body weight in the range of 1.2-6.2 kg. BP measurements were in the range of 95-179 mmHg. Oscillometric measurements were associated with lower arousal scores. The Doppler device demonstrated superior repeatability precision. Body weight, sex, cuff size, cuff location, arousal score and the order in which the devices were used had no significant effect on precision or agreement. Correlation between the two devices was 0.0837 and agreement was considered clinically acceptable (<10 mmHg) in 32/50 (64%) cats. Of the 18 cats with suboptimal agreement, the oscillometric mean BP was higher in 14 (78%) cats.</p><p><strong>Conclusions and relevance: </strong>Based on this study sample, suboptimal agreement between paired CAT+ Doppler and SunTech Vet20 oscillometric BP measurements in 36% of conscious cats suggests that these devices should not be used interchangeably. Compared with Doppler, oscillometric measurements were consistently higher, particularly at higher blood pressures. Further studies are required to assess which device has superior accuracy in conscious cats.</p>","PeriodicalId":15851,"journal":{"name":"Journal of Feline Medicine and Surgery","volume":"26 2","pages":"1098612X231216350"},"PeriodicalIF":1.9000,"publicationDate":"2024-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10911307/pdf/","citationCount":"0","resultStr":"{\"title\":\"Comparison of Thames Medical CAT+ Doppler and SunTech Vet 20 oscillometric devices for non-invasive blood pressure measurement in conscious cats.\",\"authors\":\"Clara Casas, Charlotte Dye\",\"doi\":\"10.1177/1098612X231216350\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objectives: </strong>A comparative assessment of systolic blood pressure (BP) measurement agreement and precision in two commonly used non-invasive BP devices was carried out in conscious cats.</p><p><strong>Methods: </strong>Systolic BP measurements were obtained from 50 conscious cats as part of their clinical investigations. All measurements were taken by the same operator and were performed according to the American College of Veterinary Internal Medicine (ACVIM) consensus guidelines. The same cuff location and cuff size were used for paired measurements. The order of device use was randomised, and an arousal score was assigned during each procedure. Precision was assessed using standard deviation and coefficient of variance comparisons. Agreement was assessed using graphical and statistical comparisons of derived 'delta mean' and 'mean of means' data.</p><p><strong>Results: </strong>A total of 50 cats aged between 3 months and 15 years were enrolled (29 domestic shorthair, seven domestic longhair, three British Shorthair, two Ragdoll, two Sphynx, two Persian, one Siamese, one Burmese, one Russian Blue, one Maine Coon and one Oriental) with a body weight in the range of 1.2-6.2 kg. BP measurements were in the range of 95-179 mmHg. Oscillometric measurements were associated with lower arousal scores. The Doppler device demonstrated superior repeatability precision. Body weight, sex, cuff size, cuff location, arousal score and the order in which the devices were used had no significant effect on precision or agreement. Correlation between the two devices was 0.0837 and agreement was considered clinically acceptable (<10 mmHg) in 32/50 (64%) cats. Of the 18 cats with suboptimal agreement, the oscillometric mean BP was higher in 14 (78%) cats.</p><p><strong>Conclusions and relevance: </strong>Based on this study sample, suboptimal agreement between paired CAT+ Doppler and SunTech Vet20 oscillometric BP measurements in 36% of conscious cats suggests that these devices should not be used interchangeably. Compared with Doppler, oscillometric measurements were consistently higher, particularly at higher blood pressures. Further studies are required to assess which device has superior accuracy in conscious cats.</p>\",\"PeriodicalId\":15851,\"journal\":{\"name\":\"Journal of Feline Medicine and Surgery\",\"volume\":\"26 2\",\"pages\":\"1098612X231216350\"},\"PeriodicalIF\":1.9000,\"publicationDate\":\"2024-02-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10911307/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Feline Medicine and Surgery\",\"FirstCategoryId\":\"97\",\"ListUrlMain\":\"https://doi.org/10.1177/1098612X231216350\",\"RegionNum\":2,\"RegionCategory\":\"农林科学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"VETERINARY SCIENCES\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Feline Medicine and Surgery","FirstCategoryId":"97","ListUrlMain":"https://doi.org/10.1177/1098612X231216350","RegionNum":2,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

摘要

目的在有意识的猫身上对两种常用无创血压设备的收缩压测量一致性和精确性进行比较评估:方法:作为临床检查的一部分,对 50 只意识清醒的猫进行了收缩压测量。所有测量均由同一操作者进行,并按照美国兽医内科学院(ACVIM)的共识指南执行。成对测量使用相同的袖带位置和袖带尺寸。设备的使用顺序是随机的,并在每个过程中分配唤醒评分。精确度采用标准偏差和方差系数比较法进行评估。通过对得出的 "delta 平均值 "和 "平均值的平均值 "数据进行图形和统计比较来评估一致性:共有 50 只年龄在 3 个月到 15 岁之间的猫(29 只家养短毛猫、7 只家养长毛猫、3 只英国短毛猫、2 只布娃娃猫、2 只斯芬克斯猫、2 只波斯猫、1 只暹罗猫、1 只缅甸猫、1 只俄罗斯蓝猫、1 只缅因浣熊猫和 1 只东方猫)参加了测量,体重在 1.2-6.2 公斤之间。血压测量值在 95-179 mmHg 之间。振荡测量与较低的唤醒评分有关。多普勒设备的重复性精度更高。体重、性别、袖带尺寸、袖带位置、唤醒评分和设备使用顺序对精确度或一致性没有显著影响。两种设备之间的相关性为 0.0837,一致性在临床上是可以接受的(结论和相关性:根据这项研究的样本,在 36% 的清醒猫中,CAT+ 多普勒和 SunTech Vet20 血压示波测量法之间的一致性不佳,这表明这些设备不应互换使用。与多普勒测量仪相比,示波测量仪的测量结果始终较高,尤其是在血压较高的情况下。还需要进一步研究,以评估哪种设备在有意识的猫中具有更高的准确性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Comparison of Thames Medical CAT+ Doppler and SunTech Vet 20 oscillometric devices for non-invasive blood pressure measurement in conscious cats.

Objectives: A comparative assessment of systolic blood pressure (BP) measurement agreement and precision in two commonly used non-invasive BP devices was carried out in conscious cats.

Methods: Systolic BP measurements were obtained from 50 conscious cats as part of their clinical investigations. All measurements were taken by the same operator and were performed according to the American College of Veterinary Internal Medicine (ACVIM) consensus guidelines. The same cuff location and cuff size were used for paired measurements. The order of device use was randomised, and an arousal score was assigned during each procedure. Precision was assessed using standard deviation and coefficient of variance comparisons. Agreement was assessed using graphical and statistical comparisons of derived 'delta mean' and 'mean of means' data.

Results: A total of 50 cats aged between 3 months and 15 years were enrolled (29 domestic shorthair, seven domestic longhair, three British Shorthair, two Ragdoll, two Sphynx, two Persian, one Siamese, one Burmese, one Russian Blue, one Maine Coon and one Oriental) with a body weight in the range of 1.2-6.2 kg. BP measurements were in the range of 95-179 mmHg. Oscillometric measurements were associated with lower arousal scores. The Doppler device demonstrated superior repeatability precision. Body weight, sex, cuff size, cuff location, arousal score and the order in which the devices were used had no significant effect on precision or agreement. Correlation between the two devices was 0.0837 and agreement was considered clinically acceptable (<10 mmHg) in 32/50 (64%) cats. Of the 18 cats with suboptimal agreement, the oscillometric mean BP was higher in 14 (78%) cats.

Conclusions and relevance: Based on this study sample, suboptimal agreement between paired CAT+ Doppler and SunTech Vet20 oscillometric BP measurements in 36% of conscious cats suggests that these devices should not be used interchangeably. Compared with Doppler, oscillometric measurements were consistently higher, particularly at higher blood pressures. Further studies are required to assess which device has superior accuracy in conscious cats.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
3.90
自引率
17.60%
发文量
254
审稿时长
8-16 weeks
期刊介绍: JFMS is an international, peer-reviewed journal aimed at both practitioners and researchers with an interest in the clinical veterinary healthcare of domestic cats. The journal is published monthly in two formats: ‘Classic’ editions containing high-quality original papers on all aspects of feline medicine and surgery, including basic research relevant to clinical practice; and dedicated ‘Clinical Practice’ editions primarily containing opinionated review articles providing state-of-the-art information for feline clinicians, along with other relevant articles such as consensus guidelines.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信