Zeyang Li, Stuart Pollard, Gemma Smith, Subodh Deshmukh, Ziyun Ding
{"title":"骨折愈合过程中组合孔锁定加压钢板的生物力学分析:螺钉配置的数值研究","authors":"Zeyang Li, Stuart Pollard, Gemma Smith, Subodh Deshmukh, Ziyun Ding","doi":"10.1177/09544119241229157","DOIUrl":null,"url":null,"abstract":"<p><p>Locking compression plates (LCPs) have become a widely used option for treating femur bone fractures. However, the optimal screw configuration with combi-holes remains a subject of debate. The study aims to create a time-dependent finite element (FE) model to assess the impacts of different screw configurations on LCP fixation stiffness and healing efficiency across four healing stages during a complete fracture healing process. To simulate the healing process, we integrated a time-dependent callus formation mechanism into a FE model of the LCP with combi-holes. Three screw configuration parameters, namely working length, screw number, and screw position, were investigated. Increasing the working length negatively affected axial stiffness and healing efficiency (<i>p</i> < 0.001), while screw number or position had no significant impact (<i>p</i> > 0.01). The time-dependent model displayed a moderate correlation with the conventional time-independent model for axial stiffness and healing efficiency (ρ ≥ 0.733, <i>p</i> ≤ 0.025). The highest healing efficiency (95.2%) was observed in screw configuration C125 during the 4-8-week period. The results provide insights into managing fractures using LCPs with combi-holes over an extended duration. Under axial compressive loading conditions, the use of the C125 screw configuration can enhance callus formation during the 4-12-week period for transverse fractures. When employing the C12345 configuration, it becomes crucial to avoid overconstraint during the 4-8-week period.</p>","PeriodicalId":20666,"journal":{"name":"Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine","volume":" ","pages":"313-323"},"PeriodicalIF":1.7000,"publicationDate":"2024-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10941711/pdf/","citationCount":"0","resultStr":"{\"title\":\"Biomechanical analysis of combi-hole locking compression plate during fracture healing: A numerical study of screw configuration.\",\"authors\":\"Zeyang Li, Stuart Pollard, Gemma Smith, Subodh Deshmukh, Ziyun Ding\",\"doi\":\"10.1177/09544119241229157\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Locking compression plates (LCPs) have become a widely used option for treating femur bone fractures. However, the optimal screw configuration with combi-holes remains a subject of debate. The study aims to create a time-dependent finite element (FE) model to assess the impacts of different screw configurations on LCP fixation stiffness and healing efficiency across four healing stages during a complete fracture healing process. To simulate the healing process, we integrated a time-dependent callus formation mechanism into a FE model of the LCP with combi-holes. Three screw configuration parameters, namely working length, screw number, and screw position, were investigated. Increasing the working length negatively affected axial stiffness and healing efficiency (<i>p</i> < 0.001), while screw number or position had no significant impact (<i>p</i> > 0.01). The time-dependent model displayed a moderate correlation with the conventional time-independent model for axial stiffness and healing efficiency (ρ ≥ 0.733, <i>p</i> ≤ 0.025). The highest healing efficiency (95.2%) was observed in screw configuration C125 during the 4-8-week period. The results provide insights into managing fractures using LCPs with combi-holes over an extended duration. Under axial compressive loading conditions, the use of the C125 screw configuration can enhance callus formation during the 4-12-week period for transverse fractures. When employing the C12345 configuration, it becomes crucial to avoid overconstraint during the 4-8-week period.</p>\",\"PeriodicalId\":20666,\"journal\":{\"name\":\"Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine\",\"volume\":\" \",\"pages\":\"313-323\"},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2024-03-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10941711/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1177/09544119241229157\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/2/19 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1177/09544119241229157","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/19 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
Biomechanical analysis of combi-hole locking compression plate during fracture healing: A numerical study of screw configuration.
Locking compression plates (LCPs) have become a widely used option for treating femur bone fractures. However, the optimal screw configuration with combi-holes remains a subject of debate. The study aims to create a time-dependent finite element (FE) model to assess the impacts of different screw configurations on LCP fixation stiffness and healing efficiency across four healing stages during a complete fracture healing process. To simulate the healing process, we integrated a time-dependent callus formation mechanism into a FE model of the LCP with combi-holes. Three screw configuration parameters, namely working length, screw number, and screw position, were investigated. Increasing the working length negatively affected axial stiffness and healing efficiency (p < 0.001), while screw number or position had no significant impact (p > 0.01). The time-dependent model displayed a moderate correlation with the conventional time-independent model for axial stiffness and healing efficiency (ρ ≥ 0.733, p ≤ 0.025). The highest healing efficiency (95.2%) was observed in screw configuration C125 during the 4-8-week period. The results provide insights into managing fractures using LCPs with combi-holes over an extended duration. Under axial compressive loading conditions, the use of the C125 screw configuration can enhance callus formation during the 4-12-week period for transverse fractures. When employing the C12345 configuration, it becomes crucial to avoid overconstraint during the 4-8-week period.
期刊介绍:
The Journal of Engineering in Medicine is an interdisciplinary journal encompassing all aspects of engineering in medicine. The Journal is a vital tool for maintaining an understanding of the newest techniques and research in medical engineering.