IODA-Shoulder,用于量化肩关节不稳定骨质流失的自动 3D 软件。验证研究

Q4 Medicine
Marco Maiotti , Saverio De Nigris , Stefano Tardioli , Massimo Cavaliere , Santo Barillà , Arianna Angelini , Guglielmo Torre , Marc Olivier Gauci
{"title":"IODA-Shoulder,用于量化肩关节不稳定骨质流失的自动 3D 软件。验证研究","authors":"Marco Maiotti ,&nbsp;Saverio De Nigris ,&nbsp;Stefano Tardioli ,&nbsp;Massimo Cavaliere ,&nbsp;Santo Barillà ,&nbsp;Arianna Angelini ,&nbsp;Guglielmo Torre ,&nbsp;Marc Olivier Gauci","doi":"10.1016/j.rcot.2024.01.001","DOIUrl":null,"url":null,"abstract":"<div><h3>Contexte</h3><p>Le calcul de la perte de substance osseuse glénoïdienne (PSOG) chez les patients souffrant d’instabilité de l’épaule présente une grande variabilité de méthodes et leur reproductibilité varie. L’objectif de cette étude était de décrire et de valider un nouveau logiciel semi-automatique développé pour améliorer le calcul de la PSOG en utilisant un modèle d’imagerie 3D (IODA-shoulder) et de comparer la méthode à la méthode PICO.</p></div><div><h3>Méthodes</h3><p>Un logiciel semi-automatique pour évaluer le PSOG a été développé et validé sur 7 spécimens frais congelés (scapula avec défect glénoïdien réalisé manuellement), en utilisant la méthode de déplacement de l’eau pour mesurer les volumes. Ensuite, le logiciel a été utilisé rétrospectivement sur 20 patients souffrant de luxations récidivantes de l’épaule. Les critères d’inclusion étaient : luxation unilatérale avec au moins deux épisodes de luxation. Les critères d’exclusion étaient : reconstruction osseuse de la glène, échec de la stabilisation de l’épaule, luxation bilatérale, arthrite de l’épaule. Des reconstructions 3D à partir de scanners des deux épaules étaient collectées pour chaque patient. Deux méthodes de détermination du PSOG étaient comparées. La méthode de la surface PICO et la nouvelle méthode <em>IODA-Shoulder</em>. Nous avons évalué la reproductibilité intra et inter-observateurs des deux méthodes avec le coefficient de corrélation intraclasse (CCI), l’analyse de Bland-Alman et le coefficient de corrélation de concordance de Lin (CCC).</p></div><div><h3>Résultats</h3><p>Nous n’avons pas trouvé de différence statistiquement significative entre les volumes moyens calculés avec les méthodes PICO et <em>IODA-Shoulder</em>, respectivement 914 vs. 815 mm<sup>3</sup>, <em>p</em> <!-->=<!--> <!-->0,155. L’analyse réalisée à l’aide de la méthode PICO traditionnelle a montré un taux de concordance plus faible entre quatre observateurs alors qu’une concordance plus élevée était retrouvée avec la méthode <em>IODA-Shoulder</em>, indépendamment de la taille et de la localisation du défect. La concordance CCI avec PICO était significativement plus faible qu’avec <em>IODA-Shoulder</em> (0,76 contre 0,97). Nous avons constaté une faible CCC avec PICO (de 0,65 à 0,81) et une CCC importante avec <em>IODA-Shoulder</em> (de 0,96 à 0,98).</p></div><div><h3>Discussion</h3><p>La fiabilité intra et inter-observateur de la méthode <em>IODA-Shoulder</em> est significativement meilleure que celle de la méthode PICO. L’évaluation de la PSOG par la méthode <em>IODA-Shoulder</em> permet de gagner du temps et d’éviter des variations inter et intra observateurs significatives principalement relatives à l’expérience de l’évaluateur.</p></div><div><h3>Niveau de preuve</h3><p>IV ; étude expérimentale.</p></div><div><h3>Background</h3><p>Glenoid Bone Loss (GBL) calculation in patients with shoulder instability has a wide variability in methods and their reliability. The purpose of this study was to describe and validate a new semi-automated software developed to improve GBL calculation using a 3D imaging modeling (IODA-shoulder) and to compare the method to the PICO area method.</p></div><div><h3>Patients and methods</h3><p>A semi-automated software to assess GBL was preliminary developed and validated on 7 fresh frozen specimens (scapulae with handmade glenoid defect), using water displacement method. Afterwards, the software was retrospectively deployed in 20 patients affected by recurrent shoulder dislocation. Inclusion criteria were: unilateral dislocation, minimum 2 dislocation episodes. Exclusion criteria were bone reconstruction of the glenoid, failed shoulder stabilization, bilateral dislocation, shoulder arthritis. Three-dimensional computed tomography images of bilateral shoulder were retrieved for each patient. Two methods to determine GBL were compared. The PICO surface area method and the new IODA method. We assessed the intra and inter-rater reliability of the two methods with the Intraclass Correlation Coefficient (ICC), the Bland-Alman analysis, and Lin's concordance correlation coefficient (CCC).</p></div><div><h3>Results</h3><p>we did not find a statistically significant difference between the mean volumes calculated with PICO and IODA methods, respectively 914 vs. 815 mm<sup>3</sup>, <em>P</em> <!-->=<!--> <!-->0.155. The analysis carried out by using the traditional PICO method showed a lower concordance rate among four observers than the higher concordance found using IODA method, regardless of the size and the location of the defect. The ICC agreement with PICO was significantly lower than with IODA (0.76 vs. 0.97). We found a poor CCC with PICO (from 0.65 to 0.81) and a substantial one with IODA (from 0.96 to 0.98).</p></div><div><h3>Discussion</h3><p>The intra and inter-rater reliability using IODA method is significantly better than PICO method. The assessment of GBL using IODA method is time saving, avoiding significant inter and intra observer variation, mainly due to individual skill and experience in the method.</p></div><div><h3>Level of evidence</h3><p>IV Experimental study.</p></div>","PeriodicalId":39565,"journal":{"name":"Revue de Chirurgie Orthopedique et Traumatologique","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"IODA-Shoulder, un logiciel 3D automatique pour quantifier la perte de substance osseuse dans l’instabilité de l’épaule. Étude de validation\",\"authors\":\"Marco Maiotti ,&nbsp;Saverio De Nigris ,&nbsp;Stefano Tardioli ,&nbsp;Massimo Cavaliere ,&nbsp;Santo Barillà ,&nbsp;Arianna Angelini ,&nbsp;Guglielmo Torre ,&nbsp;Marc Olivier Gauci\",\"doi\":\"10.1016/j.rcot.2024.01.001\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Contexte</h3><p>Le calcul de la perte de substance osseuse glénoïdienne (PSOG) chez les patients souffrant d’instabilité de l’épaule présente une grande variabilité de méthodes et leur reproductibilité varie. L’objectif de cette étude était de décrire et de valider un nouveau logiciel semi-automatique développé pour améliorer le calcul de la PSOG en utilisant un modèle d’imagerie 3D (IODA-shoulder) et de comparer la méthode à la méthode PICO.</p></div><div><h3>Méthodes</h3><p>Un logiciel semi-automatique pour évaluer le PSOG a été développé et validé sur 7 spécimens frais congelés (scapula avec défect glénoïdien réalisé manuellement), en utilisant la méthode de déplacement de l’eau pour mesurer les volumes. Ensuite, le logiciel a été utilisé rétrospectivement sur 20 patients souffrant de luxations récidivantes de l’épaule. Les critères d’inclusion étaient : luxation unilatérale avec au moins deux épisodes de luxation. Les critères d’exclusion étaient : reconstruction osseuse de la glène, échec de la stabilisation de l’épaule, luxation bilatérale, arthrite de l’épaule. Des reconstructions 3D à partir de scanners des deux épaules étaient collectées pour chaque patient. Deux méthodes de détermination du PSOG étaient comparées. La méthode de la surface PICO et la nouvelle méthode <em>IODA-Shoulder</em>. Nous avons évalué la reproductibilité intra et inter-observateurs des deux méthodes avec le coefficient de corrélation intraclasse (CCI), l’analyse de Bland-Alman et le coefficient de corrélation de concordance de Lin (CCC).</p></div><div><h3>Résultats</h3><p>Nous n’avons pas trouvé de différence statistiquement significative entre les volumes moyens calculés avec les méthodes PICO et <em>IODA-Shoulder</em>, respectivement 914 vs. 815 mm<sup>3</sup>, <em>p</em> <!-->=<!--> <!-->0,155. L’analyse réalisée à l’aide de la méthode PICO traditionnelle a montré un taux de concordance plus faible entre quatre observateurs alors qu’une concordance plus élevée était retrouvée avec la méthode <em>IODA-Shoulder</em>, indépendamment de la taille et de la localisation du défect. La concordance CCI avec PICO était significativement plus faible qu’avec <em>IODA-Shoulder</em> (0,76 contre 0,97). Nous avons constaté une faible CCC avec PICO (de 0,65 à 0,81) et une CCC importante avec <em>IODA-Shoulder</em> (de 0,96 à 0,98).</p></div><div><h3>Discussion</h3><p>La fiabilité intra et inter-observateur de la méthode <em>IODA-Shoulder</em> est significativement meilleure que celle de la méthode PICO. L’évaluation de la PSOG par la méthode <em>IODA-Shoulder</em> permet de gagner du temps et d’éviter des variations inter et intra observateurs significatives principalement relatives à l’expérience de l’évaluateur.</p></div><div><h3>Niveau de preuve</h3><p>IV ; étude expérimentale.</p></div><div><h3>Background</h3><p>Glenoid Bone Loss (GBL) calculation in patients with shoulder instability has a wide variability in methods and their reliability. The purpose of this study was to describe and validate a new semi-automated software developed to improve GBL calculation using a 3D imaging modeling (IODA-shoulder) and to compare the method to the PICO area method.</p></div><div><h3>Patients and methods</h3><p>A semi-automated software to assess GBL was preliminary developed and validated on 7 fresh frozen specimens (scapulae with handmade glenoid defect), using water displacement method. Afterwards, the software was retrospectively deployed in 20 patients affected by recurrent shoulder dislocation. Inclusion criteria were: unilateral dislocation, minimum 2 dislocation episodes. Exclusion criteria were bone reconstruction of the glenoid, failed shoulder stabilization, bilateral dislocation, shoulder arthritis. Three-dimensional computed tomography images of bilateral shoulder were retrieved for each patient. Two methods to determine GBL were compared. The PICO surface area method and the new IODA method. We assessed the intra and inter-rater reliability of the two methods with the Intraclass Correlation Coefficient (ICC), the Bland-Alman analysis, and Lin's concordance correlation coefficient (CCC).</p></div><div><h3>Results</h3><p>we did not find a statistically significant difference between the mean volumes calculated with PICO and IODA methods, respectively 914 vs. 815 mm<sup>3</sup>, <em>P</em> <!-->=<!--> <!-->0.155. The analysis carried out by using the traditional PICO method showed a lower concordance rate among four observers than the higher concordance found using IODA method, regardless of the size and the location of the defect. The ICC agreement with PICO was significantly lower than with IODA (0.76 vs. 0.97). We found a poor CCC with PICO (from 0.65 to 0.81) and a substantial one with IODA (from 0.96 to 0.98).</p></div><div><h3>Discussion</h3><p>The intra and inter-rater reliability using IODA method is significantly better than PICO method. The assessment of GBL using IODA method is time saving, avoiding significant inter and intra observer variation, mainly due to individual skill and experience in the method.</p></div><div><h3>Level of evidence</h3><p>IV Experimental study.</p></div>\",\"PeriodicalId\":39565,\"journal\":{\"name\":\"Revue de Chirurgie Orthopedique et Traumatologique\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-05-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Revue de Chirurgie Orthopedique et Traumatologique\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1877051724000017\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Revue de Chirurgie Orthopedique et Traumatologique","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1877051724000017","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

背景肩关节不稳定患者盂骨损失(GBL)的计算方法差异很大,其可重复性也不尽相同。本研究的目的是描述和验证一种新开发的半自动软件,该软件利用三维成像模型(IODA-shoulder)改进了PSOG的计算,并将该方法与PICO方法进行了比较。方法开发了评估PSOG的半自动软件,并在7个新鲜冷冻标本(人工制作盂缺损的肩胛骨)上进行了验证,使用水位移法测量体积。该软件随后被用于对20例复发性肩关节脱位患者进行回顾性分析。纳入标准为:至少发生过两次单侧脱位。排除标准为:盂骨重建、肩关节稳定失败、双侧脱位、肩关节炎。为每位患者收集了双肩扫描的三维重建数据。比较了两种确定 PSOG 的方法。PICO 表面法和新的 IODA 肩部法。我们使用类内相关系数 (ICC)、Bland-Alman 分析和 Lin 一致性相关系数 (CCC) 评估了这两种方法在观察者内部和观察者之间的可重复性。使用传统的 PICO 方法进行的分析表明,四名观察者之间的一致性较低,而使用 IODA 肩部方法,无论缺损的大小和位置如何,一致性都较高。PICO 的 CCI 一致性明显低于 IODA 肩法(0.76 对 0.97)。我们发现 PICO 的 CCC 较低(从 0.65 到 0.81),而 IODA-Shoulder 的 CCC 较高(从 0.96 到 0.98)。IODA-Shoulder法评估PSOG节省了时间,避免了观察者之间和观察者内部的显著差异,这些差异主要与评估者的经验有关。本研究的目的是描述和验证一种新开发的半自动软件,该软件利用三维成像建模(IODA-shoulder)改进了GBL的计算,并将该方法与PICO面积法进行了比较。患者和方法初步开发了一种评估GBL的半自动软件,并利用水位移法在7个新鲜冷冻标本(具有手工制作盂状缺损的肩胛骨)上进行了验证。随后,该软件在20例复发性肩关节脱位患者中进行了回顾性应用。纳入标准为:单侧脱位,至少两次脱位发作。排除标准为:盂骨重建、肩关节稳定失败、双侧脱位、肩关节炎。为每位患者采集双侧肩部的三维计算机断层扫描图像。比较了两种确定 GBL 的方法。PICO表面积法和新的IODA法。我们使用类内相关系数(ICC)、Bland-Alman 分析和 Lin 一致性相关系数(CCC)评估了这两种方法在评分者内部和评分者之间的可靠性。结果我们没有发现 PICO 和 IODA 方法计算出的平均体积(分别为 914 vs. 815 mm3,P = 0.155)有显著的统计学差异。使用传统的 PICO 方法进行的分析表明,无论缺损的大小和位置如何,四位观察者的一致率均低于使用 IODA 方法的一致率。PICO 的 ICC 一致性明显低于 IODA(0.76 对 0.97)。我们发现 PICO 的 CCC 较差(从 0.65 到 0.81),而 IODA 的 CCC 较高(从 0.96 到 0.98)。使用 IODA 法评估 GBL 节省了时间,避免了观察者之间和观察者内部的显著差异,这主要是由于个人技能和方法经验造成的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
IODA-Shoulder, un logiciel 3D automatique pour quantifier la perte de substance osseuse dans l’instabilité de l’épaule. Étude de validation

Contexte

Le calcul de la perte de substance osseuse glénoïdienne (PSOG) chez les patients souffrant d’instabilité de l’épaule présente une grande variabilité de méthodes et leur reproductibilité varie. L’objectif de cette étude était de décrire et de valider un nouveau logiciel semi-automatique développé pour améliorer le calcul de la PSOG en utilisant un modèle d’imagerie 3D (IODA-shoulder) et de comparer la méthode à la méthode PICO.

Méthodes

Un logiciel semi-automatique pour évaluer le PSOG a été développé et validé sur 7 spécimens frais congelés (scapula avec défect glénoïdien réalisé manuellement), en utilisant la méthode de déplacement de l’eau pour mesurer les volumes. Ensuite, le logiciel a été utilisé rétrospectivement sur 20 patients souffrant de luxations récidivantes de l’épaule. Les critères d’inclusion étaient : luxation unilatérale avec au moins deux épisodes de luxation. Les critères d’exclusion étaient : reconstruction osseuse de la glène, échec de la stabilisation de l’épaule, luxation bilatérale, arthrite de l’épaule. Des reconstructions 3D à partir de scanners des deux épaules étaient collectées pour chaque patient. Deux méthodes de détermination du PSOG étaient comparées. La méthode de la surface PICO et la nouvelle méthode IODA-Shoulder. Nous avons évalué la reproductibilité intra et inter-observateurs des deux méthodes avec le coefficient de corrélation intraclasse (CCI), l’analyse de Bland-Alman et le coefficient de corrélation de concordance de Lin (CCC).

Résultats

Nous n’avons pas trouvé de différence statistiquement significative entre les volumes moyens calculés avec les méthodes PICO et IODA-Shoulder, respectivement 914 vs. 815 mm3, p = 0,155. L’analyse réalisée à l’aide de la méthode PICO traditionnelle a montré un taux de concordance plus faible entre quatre observateurs alors qu’une concordance plus élevée était retrouvée avec la méthode IODA-Shoulder, indépendamment de la taille et de la localisation du défect. La concordance CCI avec PICO était significativement plus faible qu’avec IODA-Shoulder (0,76 contre 0,97). Nous avons constaté une faible CCC avec PICO (de 0,65 à 0,81) et une CCC importante avec IODA-Shoulder (de 0,96 à 0,98).

Discussion

La fiabilité intra et inter-observateur de la méthode IODA-Shoulder est significativement meilleure que celle de la méthode PICO. L’évaluation de la PSOG par la méthode IODA-Shoulder permet de gagner du temps et d’éviter des variations inter et intra observateurs significatives principalement relatives à l’expérience de l’évaluateur.

Niveau de preuve

IV ; étude expérimentale.

Background

Glenoid Bone Loss (GBL) calculation in patients with shoulder instability has a wide variability in methods and their reliability. The purpose of this study was to describe and validate a new semi-automated software developed to improve GBL calculation using a 3D imaging modeling (IODA-shoulder) and to compare the method to the PICO area method.

Patients and methods

A semi-automated software to assess GBL was preliminary developed and validated on 7 fresh frozen specimens (scapulae with handmade glenoid defect), using water displacement method. Afterwards, the software was retrospectively deployed in 20 patients affected by recurrent shoulder dislocation. Inclusion criteria were: unilateral dislocation, minimum 2 dislocation episodes. Exclusion criteria were bone reconstruction of the glenoid, failed shoulder stabilization, bilateral dislocation, shoulder arthritis. Three-dimensional computed tomography images of bilateral shoulder were retrieved for each patient. Two methods to determine GBL were compared. The PICO surface area method and the new IODA method. We assessed the intra and inter-rater reliability of the two methods with the Intraclass Correlation Coefficient (ICC), the Bland-Alman analysis, and Lin's concordance correlation coefficient (CCC).

Results

we did not find a statistically significant difference between the mean volumes calculated with PICO and IODA methods, respectively 914 vs. 815 mm3, P = 0.155. The analysis carried out by using the traditional PICO method showed a lower concordance rate among four observers than the higher concordance found using IODA method, regardless of the size and the location of the defect. The ICC agreement with PICO was significantly lower than with IODA (0.76 vs. 0.97). We found a poor CCC with PICO (from 0.65 to 0.81) and a substantial one with IODA (from 0.96 to 0.98).

Discussion

The intra and inter-rater reliability using IODA method is significantly better than PICO method. The assessment of GBL using IODA method is time saving, avoiding significant inter and intra observer variation, mainly due to individual skill and experience in the method.

Level of evidence

IV Experimental study.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
0.10
自引率
0.00%
发文量
301
期刊介绍: A 118 ans, la Revue de Chirurgie orthopédique franchit, en 2009, une étape décisive dans son développement afin de renforcer la diffusion et la notoriété des publications francophones auprès des praticiens et chercheurs non-francophones. Les auteurs ayant leurs racines dans la francophonie trouveront ainsi une chance supplémentaire de voir reconnus les qualités et le intérêt de leurs recherches par le plus grand nombre.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信