{"title":"测量两种护理点分析仪和一种台式分析仪对伊斯帕尼奥拉亚马逊鹦鹉(Amazona ventralis)乳酸测量结果的一致程度。","authors":"Alexandra K Mason, Jordan Padlo, Mark A Mitchell, J Jill Heatley, Thomas N Tully","doi":"10.1647/1082-6742-37.4.314","DOIUrl":null,"url":null,"abstract":"<p><p>Lactate is an important biochemistry analyte used in human and veterinary medicine to assess tissue perfusion and can be used as a prognostic indicator for certain disease conditions. Whereas lactate is commonly measured using \"patient-side\" handheld meters, these meters have not been validated for companion avian species. The purpose of this study was to measure the level of agreement between 2 commercially available point-of-care lactate meters and a laboratory benchtop blood analyzer in Hispaniolan Amazon parrots (<i>Amazona ventralis</i>). Blood samples were collected from 20 adult parrots at Louisiana State University by drawing 1.5 mL of blood from the right jugular vein. One drop of whole blood was used for the Lactate Plus analyzer and the remainder of the sample transferred into a lithium heparin microtainer. From the blood in the microtainer, 0.2 mL whole blood was analyzed using the epoc Blood Analysis System, and the remaining sample was centrifuged to obtain plasma that was immediately frozen at -80°C (-112°F) and submitted to the Texas A&M University Clinical Pathology Laboratory for analysis on the VITROS 4500 benchtop analyzer. Bland-Altman agreement plots and Passing-Bablok regression were used to measure the level of agreement between the methods. There was poor agreement between all 3 methods with mean percentage differences in lactate concentrations ≥22% (epoc and Lactate Plus: 33.6% [95% CI: 27-40]; epoc and VITROS 4500: 55% [95% CI:52-58]; VITROS 4500 and Lactate Plus: 22% [95% CI:16-28]). Based on these results, the point-of-care meters tested in this study are not interchangeable, and separate reference intervals were calculated for each method. Blood lactate concentrations may have more utility in tracing lactate trends over time in an individual rather than being able to utilize this information at 1 time point for disease diagnosis and prognosis.</p>","PeriodicalId":15102,"journal":{"name":"Journal of Avian Medicine and Surgery","volume":"37 4","pages":"314-320"},"PeriodicalIF":0.6000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Measuring the Level of Agreement for Lactate Measurements in Hispaniolan Amazon Parrots (<i>Amazona ventralis</i>) Among 2 Point-of-Care Analyzers and a Benchtop Analyzer.\",\"authors\":\"Alexandra K Mason, Jordan Padlo, Mark A Mitchell, J Jill Heatley, Thomas N Tully\",\"doi\":\"10.1647/1082-6742-37.4.314\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Lactate is an important biochemistry analyte used in human and veterinary medicine to assess tissue perfusion and can be used as a prognostic indicator for certain disease conditions. Whereas lactate is commonly measured using \\\"patient-side\\\" handheld meters, these meters have not been validated for companion avian species. The purpose of this study was to measure the level of agreement between 2 commercially available point-of-care lactate meters and a laboratory benchtop blood analyzer in Hispaniolan Amazon parrots (<i>Amazona ventralis</i>). Blood samples were collected from 20 adult parrots at Louisiana State University by drawing 1.5 mL of blood from the right jugular vein. One drop of whole blood was used for the Lactate Plus analyzer and the remainder of the sample transferred into a lithium heparin microtainer. From the blood in the microtainer, 0.2 mL whole blood was analyzed using the epoc Blood Analysis System, and the remaining sample was centrifuged to obtain plasma that was immediately frozen at -80°C (-112°F) and submitted to the Texas A&M University Clinical Pathology Laboratory for analysis on the VITROS 4500 benchtop analyzer. Bland-Altman agreement plots and Passing-Bablok regression were used to measure the level of agreement between the methods. There was poor agreement between all 3 methods with mean percentage differences in lactate concentrations ≥22% (epoc and Lactate Plus: 33.6% [95% CI: 27-40]; epoc and VITROS 4500: 55% [95% CI:52-58]; VITROS 4500 and Lactate Plus: 22% [95% CI:16-28]). Based on these results, the point-of-care meters tested in this study are not interchangeable, and separate reference intervals were calculated for each method. Blood lactate concentrations may have more utility in tracing lactate trends over time in an individual rather than being able to utilize this information at 1 time point for disease diagnosis and prognosis.</p>\",\"PeriodicalId\":15102,\"journal\":{\"name\":\"Journal of Avian Medicine and Surgery\",\"volume\":\"37 4\",\"pages\":\"314-320\"},\"PeriodicalIF\":0.6000,\"publicationDate\":\"2024-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Avian Medicine and Surgery\",\"FirstCategoryId\":\"97\",\"ListUrlMain\":\"https://doi.org/10.1647/1082-6742-37.4.314\",\"RegionNum\":4,\"RegionCategory\":\"农林科学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"VETERINARY SCIENCES\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Avian Medicine and Surgery","FirstCategoryId":"97","ListUrlMain":"https://doi.org/10.1647/1082-6742-37.4.314","RegionNum":4,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

摘要

乳酸盐是一种重要的生化分析物,在人类和兽医学中用于评估组织灌注,并可作为某些疾病的预后指标。乳酸盐通常使用 "患者端 "手持式测量仪进行测量,但这些测量仪尚未针对伴侣禽类物种进行过验证。本研究的目的是测量两种市场上销售的护理点乳酸测量仪与实验室台式血液分析仪在伊斯帕尼奥拉亚马逊鹦鹉(Amazona ventralis)中的一致性水平。我们从路易斯安那州立大学的 20 只成年鹦鹉的右颈静脉抽取了 1.5 mL 的血液样本。一滴全血用于 Lactate Plus 分析仪,其余样本转移到肝素锂微容器中。从微量容器中抽取 0.2 毫升全血,使用 epoc 血液分析系统进行分析,其余样本离心后获得血浆,立即冷冻至 -80°C (-112°F),并提交至德克萨斯农工大学临床病理实验室,使用 VITROS 4500 台式分析仪进行分析。使用 Bland-Altman 一致图和 Passing-Bablok 回归法来衡量方法之间的一致程度。所有 3 种方法之间的一致性都很差,乳酸浓度的平均百分比差异≥22%(EPOC 和 Lactate Plus:33.6% [95% CI:27-40];EPOC 和 VITROS 4500:55% [95% CI:52-58];VITROS 4500 和 Lactate Plus:22% [95% CI:16-28])。根据这些结果,本研究中测试的护理点测量仪不能互换,因此为每种方法分别计算了参考区间。血液乳酸浓度在追踪个体乳酸随时间变化的趋势方面可能更有用,而不是在一个时间点上利用这一信息进行疾病诊断和预后。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Measuring the Level of Agreement for Lactate Measurements in Hispaniolan Amazon Parrots (Amazona ventralis) Among 2 Point-of-Care Analyzers and a Benchtop Analyzer.

Lactate is an important biochemistry analyte used in human and veterinary medicine to assess tissue perfusion and can be used as a prognostic indicator for certain disease conditions. Whereas lactate is commonly measured using "patient-side" handheld meters, these meters have not been validated for companion avian species. The purpose of this study was to measure the level of agreement between 2 commercially available point-of-care lactate meters and a laboratory benchtop blood analyzer in Hispaniolan Amazon parrots (Amazona ventralis). Blood samples were collected from 20 adult parrots at Louisiana State University by drawing 1.5 mL of blood from the right jugular vein. One drop of whole blood was used for the Lactate Plus analyzer and the remainder of the sample transferred into a lithium heparin microtainer. From the blood in the microtainer, 0.2 mL whole blood was analyzed using the epoc Blood Analysis System, and the remaining sample was centrifuged to obtain plasma that was immediately frozen at -80°C (-112°F) and submitted to the Texas A&M University Clinical Pathology Laboratory for analysis on the VITROS 4500 benchtop analyzer. Bland-Altman agreement plots and Passing-Bablok regression were used to measure the level of agreement between the methods. There was poor agreement between all 3 methods with mean percentage differences in lactate concentrations ≥22% (epoc and Lactate Plus: 33.6% [95% CI: 27-40]; epoc and VITROS 4500: 55% [95% CI:52-58]; VITROS 4500 and Lactate Plus: 22% [95% CI:16-28]). Based on these results, the point-of-care meters tested in this study are not interchangeable, and separate reference intervals were calculated for each method. Blood lactate concentrations may have more utility in tracing lactate trends over time in an individual rather than being able to utilize this information at 1 time point for disease diagnosis and prognosis.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Avian Medicine and Surgery
Journal of Avian Medicine and Surgery 农林科学-兽医学
CiteScore
0.90
自引率
0.00%
发文量
52
审稿时长
>36 weeks
期刊介绍: The Journal of Avian Medicine and Surgery is an international journal of the medicine and surgery of both captive and wild birds. Published materials include scientific articles, case reports, editorials, abstracts, new research, and book reviews.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信