使用深度摄像头或压力垫进行视觉反馈平衡训练与功能性电刺激的可行性。

IF 2.9 4区 医学 Q3 ENGINEERING, BIOMEDICAL
Derrick Lim, William Pei, Jae W Lee, Kristin E Musselman, Kei Masani
{"title":"使用深度摄像头或压力垫进行视觉反馈平衡训练与功能性电刺激的可行性。","authors":"Derrick Lim, William Pei, Jae W Lee, Kristin E Musselman, Kei Masani","doi":"10.1186/s12938-023-01191-y","DOIUrl":null,"url":null,"abstract":"<p><p>Individuals with incomplete spinal-cord injury/disease are at an increased risk of falling due to their impaired ability to maintain balance. Our research group has developed a closed-loop visual-feedback balance training (VFBT) system coupled with functional electrical stimulation (FES) for rehabilitation of standing balance (FES + VFBT system); however, clinical usage of this system is limited by the use of force plates, which are expensive and not easily accessible. This study aimed to investigate the feasibility of a more affordable and accessible sensor such as a depth camera or pressure mat in place of the force plate. Ten able-bodied participants (7 males, 3 females) performed three sets of four different standing balance exercises using the FES + VFBT system with the force plate. A depth camera and pressure mat collected centre of mass and centre of pressure data passively, respectively. The depth camera showed higher Pearson's correlation (r > 98) and lower root mean squared error (RMSE < 10 mm) than the pressure mat (r > 0.82; RMSE < 4.5 mm) when compared with the force plate overall. Stimulation based on the depth camera showed lower RMSE than that based on the pressure mat relative to the FES + VFBT system. The depth camera shows potential as a replacement sensor to the force plate for providing feedback to the FES + VFBT system.</p>","PeriodicalId":8927,"journal":{"name":"BioMedical Engineering OnLine","volume":"23 1","pages":"19"},"PeriodicalIF":2.9000,"publicationDate":"2024-02-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10863251/pdf/","citationCount":"0","resultStr":"{\"title\":\"Feasibility of using a depth camera or pressure mat for visual feedback balance training with functional electrical stimulation.\",\"authors\":\"Derrick Lim, William Pei, Jae W Lee, Kristin E Musselman, Kei Masani\",\"doi\":\"10.1186/s12938-023-01191-y\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Individuals with incomplete spinal-cord injury/disease are at an increased risk of falling due to their impaired ability to maintain balance. Our research group has developed a closed-loop visual-feedback balance training (VFBT) system coupled with functional electrical stimulation (FES) for rehabilitation of standing balance (FES + VFBT system); however, clinical usage of this system is limited by the use of force plates, which are expensive and not easily accessible. This study aimed to investigate the feasibility of a more affordable and accessible sensor such as a depth camera or pressure mat in place of the force plate. Ten able-bodied participants (7 males, 3 females) performed three sets of four different standing balance exercises using the FES + VFBT system with the force plate. A depth camera and pressure mat collected centre of mass and centre of pressure data passively, respectively. The depth camera showed higher Pearson's correlation (r > 98) and lower root mean squared error (RMSE < 10 mm) than the pressure mat (r > 0.82; RMSE < 4.5 mm) when compared with the force plate overall. Stimulation based on the depth camera showed lower RMSE than that based on the pressure mat relative to the FES + VFBT system. The depth camera shows potential as a replacement sensor to the force plate for providing feedback to the FES + VFBT system.</p>\",\"PeriodicalId\":8927,\"journal\":{\"name\":\"BioMedical Engineering OnLine\",\"volume\":\"23 1\",\"pages\":\"19\"},\"PeriodicalIF\":2.9000,\"publicationDate\":\"2024-02-12\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10863251/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"BioMedical Engineering OnLine\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1186/s12938-023-01191-y\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"BioMedical Engineering OnLine","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1186/s12938-023-01191-y","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

摘要

患有不完全脊髓损伤/疾病的人由于保持平衡的能力受损,跌倒的风险会增加。我们的研究小组开发了一种闭环视觉反馈平衡训练(VFBT)系统,该系统与功能性电刺激(FES)相结合,用于站立平衡的康复(FES + VFBT 系统);然而,该系统的临床应用受到力板使用的限制,力板价格昂贵且不易获得。本研究旨在探讨使用深度摄像头或压力垫等更经济实惠、更易于获取的传感器来替代测力板的可行性。10 名健全的参与者(7 名男性,3 名女性)使用 FES + VFBT 系统和测力板进行了三组四种不同的站立平衡练习。深度摄像头和压力垫分别被动地收集质量中心和压力中心数据。深度照相机显示出较高的皮尔逊相关性(r > 98)和较低的均方根误差(RMSE 0.82;RMSE
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Feasibility of using a depth camera or pressure mat for visual feedback balance training with functional electrical stimulation.

Individuals with incomplete spinal-cord injury/disease are at an increased risk of falling due to their impaired ability to maintain balance. Our research group has developed a closed-loop visual-feedback balance training (VFBT) system coupled with functional electrical stimulation (FES) for rehabilitation of standing balance (FES + VFBT system); however, clinical usage of this system is limited by the use of force plates, which are expensive and not easily accessible. This study aimed to investigate the feasibility of a more affordable and accessible sensor such as a depth camera or pressure mat in place of the force plate. Ten able-bodied participants (7 males, 3 females) performed three sets of four different standing balance exercises using the FES + VFBT system with the force plate. A depth camera and pressure mat collected centre of mass and centre of pressure data passively, respectively. The depth camera showed higher Pearson's correlation (r > 98) and lower root mean squared error (RMSE < 10 mm) than the pressure mat (r > 0.82; RMSE < 4.5 mm) when compared with the force plate overall. Stimulation based on the depth camera showed lower RMSE than that based on the pressure mat relative to the FES + VFBT system. The depth camera shows potential as a replacement sensor to the force plate for providing feedback to the FES + VFBT system.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
BioMedical Engineering OnLine
BioMedical Engineering OnLine 工程技术-工程:生物医学
CiteScore
6.70
自引率
2.60%
发文量
79
审稿时长
1 months
期刊介绍: BioMedical Engineering OnLine is an open access, peer-reviewed journal that is dedicated to publishing research in all areas of biomedical engineering. BioMedical Engineering OnLine is aimed at readers and authors throughout the world, with an interest in using tools of the physical and data sciences and techniques in engineering to understand and solve problems in the biological and medical sciences. Topical areas include, but are not limited to: Bioinformatics- Bioinstrumentation- Biomechanics- Biomedical Devices & Instrumentation- Biomedical Signal Processing- Healthcare Information Systems- Human Dynamics- Neural Engineering- Rehabilitation Engineering- Biomaterials- Biomedical Imaging & Image Processing- BioMEMS and On-Chip Devices- Bio-Micro/Nano Technologies- Biomolecular Engineering- Biosensors- Cardiovascular Systems Engineering- Cellular Engineering- Clinical Engineering- Computational Biology- Drug Delivery Technologies- Modeling Methodologies- Nanomaterials and Nanotechnology in Biomedicine- Respiratory Systems Engineering- Robotics in Medicine- Systems and Synthetic Biology- Systems Biology- Telemedicine/Smartphone Applications in Medicine- Therapeutic Systems, Devices and Technologies- Tissue Engineering
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信