{"title":"氟化钠和血清分离管中葡萄糖浓度的比较和稳定性研究","authors":"Mustapha Dibbasey, Solomon Umukoro, Abdoulie Bojang","doi":"10.1016/j.plabm.2024.e00360","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>Sodium fluoride/potassium oxalate (NaF/KOx) tubes has been regarded as the gold-standard tubes for glucose analysis. Even though their ineffectiveness in immediately inhibiting glycolysis has been reported in several studies especially in the first 1–4h, they are still used in our clinical biochemistry laboratory for glucose measurement. However, in its absence, only serum separator tubes are employed for glucose measurement. We aim to determine whether serum separator tubes (SSTs) can replace NaF/KOx tubes for laboratory measurement of blood glucose and to assess the stability of glucose concentrations for 3 days period.</p></div><div><h3>Methods and findings</h3><p>NaF/KOx tube type was the reference method while SSTs type was the candidate method for glucose measurement. A total of 50 paired samples collected separately in NaF/KOx tubes and SSTs from healthy adult participants in the Gambia Adults Reference Intervals Study (GARIS) project were used as the project sample size. Following blood collection and separation, the glucose concentration was measured within 2 h, and at 24h, 42h and 72h time-points. Our data analysis showed no significant difference in the mean glucose concentrations between the reference tube and candidate tube types (Mean difference = 0.06 mmol/L; P = 0.38) recorded in the different timepoints. Using growth trajectory and mixed effects model, the study data further showed no significant change in the glucose concentrations (p = 0.25) for three days period.</p></div><div><h3>Conclusions</h3><p>The study confirms that SSTs can produce similar glucose results when employed in the absence of NaF/KOx tubes. Besides, the glucose concentrations were stable in both tubes for three days when the samples were separated within 2 h and refrigerated in 2–8°C.</p></div>","PeriodicalId":20421,"journal":{"name":"Practical Laboratory Medicine","volume":"39 ","pages":"Article e00360"},"PeriodicalIF":1.7000,"publicationDate":"2024-01-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2352551724000064/pdfft?md5=853bc903207429646d72e91070a938b5&pid=1-s2.0-S2352551724000064-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Comparative and stability study of glucose concentrations measured in both sodium fluoride and serum separator tubes\",\"authors\":\"Mustapha Dibbasey, Solomon Umukoro, Abdoulie Bojang\",\"doi\":\"10.1016/j.plabm.2024.e00360\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Introduction</h3><p>Sodium fluoride/potassium oxalate (NaF/KOx) tubes has been regarded as the gold-standard tubes for glucose analysis. Even though their ineffectiveness in immediately inhibiting glycolysis has been reported in several studies especially in the first 1–4h, they are still used in our clinical biochemistry laboratory for glucose measurement. However, in its absence, only serum separator tubes are employed for glucose measurement. We aim to determine whether serum separator tubes (SSTs) can replace NaF/KOx tubes for laboratory measurement of blood glucose and to assess the stability of glucose concentrations for 3 days period.</p></div><div><h3>Methods and findings</h3><p>NaF/KOx tube type was the reference method while SSTs type was the candidate method for glucose measurement. A total of 50 paired samples collected separately in NaF/KOx tubes and SSTs from healthy adult participants in the Gambia Adults Reference Intervals Study (GARIS) project were used as the project sample size. Following blood collection and separation, the glucose concentration was measured within 2 h, and at 24h, 42h and 72h time-points. Our data analysis showed no significant difference in the mean glucose concentrations between the reference tube and candidate tube types (Mean difference = 0.06 mmol/L; P = 0.38) recorded in the different timepoints. Using growth trajectory and mixed effects model, the study data further showed no significant change in the glucose concentrations (p = 0.25) for three days period.</p></div><div><h3>Conclusions</h3><p>The study confirms that SSTs can produce similar glucose results when employed in the absence of NaF/KOx tubes. Besides, the glucose concentrations were stable in both tubes for three days when the samples were separated within 2 h and refrigerated in 2–8°C.</p></div>\",\"PeriodicalId\":20421,\"journal\":{\"name\":\"Practical Laboratory Medicine\",\"volume\":\"39 \",\"pages\":\"Article e00360\"},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2024-01-23\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2352551724000064/pdfft?md5=853bc903207429646d72e91070a938b5&pid=1-s2.0-S2352551724000064-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Practical Laboratory Medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2352551724000064\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Practical Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2352551724000064","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
Comparative and stability study of glucose concentrations measured in both sodium fluoride and serum separator tubes
Introduction
Sodium fluoride/potassium oxalate (NaF/KOx) tubes has been regarded as the gold-standard tubes for glucose analysis. Even though their ineffectiveness in immediately inhibiting glycolysis has been reported in several studies especially in the first 1–4h, they are still used in our clinical biochemistry laboratory for glucose measurement. However, in its absence, only serum separator tubes are employed for glucose measurement. We aim to determine whether serum separator tubes (SSTs) can replace NaF/KOx tubes for laboratory measurement of blood glucose and to assess the stability of glucose concentrations for 3 days period.
Methods and findings
NaF/KOx tube type was the reference method while SSTs type was the candidate method for glucose measurement. A total of 50 paired samples collected separately in NaF/KOx tubes and SSTs from healthy adult participants in the Gambia Adults Reference Intervals Study (GARIS) project were used as the project sample size. Following blood collection and separation, the glucose concentration was measured within 2 h, and at 24h, 42h and 72h time-points. Our data analysis showed no significant difference in the mean glucose concentrations between the reference tube and candidate tube types (Mean difference = 0.06 mmol/L; P = 0.38) recorded in the different timepoints. Using growth trajectory and mixed effects model, the study data further showed no significant change in the glucose concentrations (p = 0.25) for three days period.
Conclusions
The study confirms that SSTs can produce similar glucose results when employed in the absence of NaF/KOx tubes. Besides, the glucose concentrations were stable in both tubes for three days when the samples were separated within 2 h and refrigerated in 2–8°C.
期刊介绍:
Practical Laboratory Medicine is a high-quality, peer-reviewed, international open-access journal publishing original research, new methods and critical evaluations, case reports and short papers in the fields of clinical chemistry and laboratory medicine. The objective of the journal is to provide practical information of immediate relevance to workers in clinical laboratories. The primary scope of the journal covers clinical chemistry, hematology, molecular biology and genetics relevant to laboratory medicine, microbiology, immunology, therapeutic drug monitoring and toxicology, laboratory management and informatics. We welcome papers which describe critical evaluations of biomarkers and their role in the diagnosis and treatment of clinically significant disease, validation of commercial and in-house IVD methods, method comparisons, interference reports, the development of new reagents and reference materials, reference range studies and regulatory compliance reports. Manuscripts describing the development of new methods applicable to laboratory medicine (including point-of-care testing) are particularly encouraged, even if preliminary or small scale.