低成本 PM10 浓度测量设备与使用各种校正函数的参考方法的等效性

T. Owczarek, M. Rogulski, P. Czechowski, A. Badyda, Ernest Czermański
{"title":"低成本 PM10 浓度测量设备与使用各种校正函数的参考方法的等效性","authors":"T. Owczarek, M. Rogulski, P. Czechowski, A. Badyda, Ernest Czermański","doi":"10.34659/eis.2023.87.4.627","DOIUrl":null,"url":null,"abstract":"The aim of the study was to build a corrective model that can be used in the analysed devices and to assess the impact of such a model on the values of the measured concentrations. The novelty of this study is the test of equivalence with the equivalent reference method for hourly data. The study used hourly data of PM10 concentrations measured in a chosen city in Poland. Data was collected from two PM10 sensors and a reference device placed in close proximity. In addition, air temperature, humidity and wind speed were also measured. Among the tested models, a linear model was selected that used primary measurements of PM10, temperature, air velocity, and humidity as the most accurate approximation of the actual PM10 concentration level. The results of the analysis showed that it is possible to build mathematical models that effectively convert PM10 concentration data from tested low-cost electronic measuring devices to concentrations obtained by the reference method.","PeriodicalId":509109,"journal":{"name":"Economics and Environment","volume":"54 41","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-01-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Equivalence of low-cost PM10 concentration measuring devices with a reference method using various correction functions\",\"authors\":\"T. Owczarek, M. Rogulski, P. Czechowski, A. Badyda, Ernest Czermański\",\"doi\":\"10.34659/eis.2023.87.4.627\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The aim of the study was to build a corrective model that can be used in the analysed devices and to assess the impact of such a model on the values of the measured concentrations. The novelty of this study is the test of equivalence with the equivalent reference method for hourly data. The study used hourly data of PM10 concentrations measured in a chosen city in Poland. Data was collected from two PM10 sensors and a reference device placed in close proximity. In addition, air temperature, humidity and wind speed were also measured. Among the tested models, a linear model was selected that used primary measurements of PM10, temperature, air velocity, and humidity as the most accurate approximation of the actual PM10 concentration level. The results of the analysis showed that it is possible to build mathematical models that effectively convert PM10 concentration data from tested low-cost electronic measuring devices to concentrations obtained by the reference method.\",\"PeriodicalId\":509109,\"journal\":{\"name\":\"Economics and Environment\",\"volume\":\"54 41\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-01-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Economics and Environment\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.34659/eis.2023.87.4.627\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Economics and Environment","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.34659/eis.2023.87.4.627","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

研究的目的是建立一个可用于分析设备的修正模型,并评估该模型对测量浓度值的影响。这项研究的新颖之处在于测试每小时数据与等效参考方法的等效性。研究使用了在波兰一个选定城市测量到的 PM10 浓度小时数据。数据收集自两个 PM10 传感器和一个放置在附近的参考装置。此外,还测量了空气温度、湿度和风速。在测试的模型中,选择了一个线性模型,该模型使用 PM10、温度、空气流速和湿度的主要测量值作为实际 PM10 浓度水平的最准确近似值。分析结果表明,可以建立数学模型,将测试的低成本电子测量设备的 PM10 浓度数据有效地转换为参考方法获得的浓度。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Equivalence of low-cost PM10 concentration measuring devices with a reference method using various correction functions
The aim of the study was to build a corrective model that can be used in the analysed devices and to assess the impact of such a model on the values of the measured concentrations. The novelty of this study is the test of equivalence with the equivalent reference method for hourly data. The study used hourly data of PM10 concentrations measured in a chosen city in Poland. Data was collected from two PM10 sensors and a reference device placed in close proximity. In addition, air temperature, humidity and wind speed were also measured. Among the tested models, a linear model was selected that used primary measurements of PM10, temperature, air velocity, and humidity as the most accurate approximation of the actual PM10 concentration level. The results of the analysis showed that it is possible to build mathematical models that effectively convert PM10 concentration data from tested low-cost electronic measuring devices to concentrations obtained by the reference method.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信