Sietske Bergsma, Margo van Gent, Alida J Dam-Vervloet, Martine C M Lagerweij, Egbert-Jan van der Wouden, Ingrid M Nijholt, Martijn F Boomsma, Lieke Poot
{"title":"图像引导下的点剪切波弹性成像:一种有效可靠的肝纤维化分期技术。","authors":"Sietske Bergsma, Margo van Gent, Alida J Dam-Vervloet, Martine C M Lagerweij, Egbert-Jan van der Wouden, Ingrid M Nijholt, Martijn F Boomsma, Lieke Poot","doi":"10.1007/s40477-023-00854-9","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Despite progressive implementation of image-guided point-shear wave elastography (pSWE) in guidelines as an alternative to transient elastography for the staging of fibrotic liver disease, pSWE is not widely adopted in clinical workflow. More information on reliability and validity of pSWE systems is needed. Therefore, we performed a phantom study to evaluate the validity and reliability of pSWE with ultrasound systems.</p><p><strong>Methods: </strong>Validity and reliability of pSWE measurements from three ultrasound systems were evaluated. Measurements were performed on an elasticity phantom with reference elasticities of 7 ± 1 (low) (median ± interquartile range (IQR)), 14 ± 2 (medium) and 26 ± 3 (high) kPa. Measurements were repeated in tenfold for each reference at 2, 3 and 4 cm depth. Results were considered valid when median elasticity ± IQR was between the uncertainty limits (IQR) for each reference elasticity value and reliable when IQR/median < 0.30.</p><p><strong>Results: </strong>pSWE with the systems provided valid results for all reference elasticities and focal depths, except for overestimation of high reference elasticity at 2 and 4 cm depth for one system (41.5 ± 4.3 and 39.0 ± 1.2 kPa, respectively). Measurements were reliable with a maximum IQR/median of 0.13, well below the guideline of IQR/median < 0.30.</p><p><strong>Discussion: </strong>The results support the use of pSWE as an alternative to invasive or non-image guided noninvasive techniques for liver fibrotic staging.</p><p><strong>Conclusions: </strong>pSWE with ultrasound systems from different vendors is valid and reliable and can therefore be implemented to optimize clinical workflow by performing imaging and elastography simultaneously.</p>","PeriodicalId":51528,"journal":{"name":"Journal of Ultrasound","volume":null,"pages":null},"PeriodicalIF":1.3000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11178714/pdf/","citationCount":"0","resultStr":"{\"title\":\"Image-guided point-shear-wave elastography: a valid and reliable technique for liver fibrotic staging.\",\"authors\":\"Sietske Bergsma, Margo van Gent, Alida J Dam-Vervloet, Martine C M Lagerweij, Egbert-Jan van der Wouden, Ingrid M Nijholt, Martijn F Boomsma, Lieke Poot\",\"doi\":\"10.1007/s40477-023-00854-9\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>Despite progressive implementation of image-guided point-shear wave elastography (pSWE) in guidelines as an alternative to transient elastography for the staging of fibrotic liver disease, pSWE is not widely adopted in clinical workflow. More information on reliability and validity of pSWE systems is needed. Therefore, we performed a phantom study to evaluate the validity and reliability of pSWE with ultrasound systems.</p><p><strong>Methods: </strong>Validity and reliability of pSWE measurements from three ultrasound systems were evaluated. Measurements were performed on an elasticity phantom with reference elasticities of 7 ± 1 (low) (median ± interquartile range (IQR)), 14 ± 2 (medium) and 26 ± 3 (high) kPa. Measurements were repeated in tenfold for each reference at 2, 3 and 4 cm depth. Results were considered valid when median elasticity ± IQR was between the uncertainty limits (IQR) for each reference elasticity value and reliable when IQR/median < 0.30.</p><p><strong>Results: </strong>pSWE with the systems provided valid results for all reference elasticities and focal depths, except for overestimation of high reference elasticity at 2 and 4 cm depth for one system (41.5 ± 4.3 and 39.0 ± 1.2 kPa, respectively). Measurements were reliable with a maximum IQR/median of 0.13, well below the guideline of IQR/median < 0.30.</p><p><strong>Discussion: </strong>The results support the use of pSWE as an alternative to invasive or non-image guided noninvasive techniques for liver fibrotic staging.</p><p><strong>Conclusions: </strong>pSWE with ultrasound systems from different vendors is valid and reliable and can therefore be implemented to optimize clinical workflow by performing imaging and elastography simultaneously.</p>\",\"PeriodicalId\":51528,\"journal\":{\"name\":\"Journal of Ultrasound\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11178714/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Ultrasound\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1007/s40477-023-00854-9\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/1/28 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Ultrasound","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s40477-023-00854-9","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/28 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Image-guided point-shear-wave elastography: a valid and reliable technique for liver fibrotic staging.
Purpose: Despite progressive implementation of image-guided point-shear wave elastography (pSWE) in guidelines as an alternative to transient elastography for the staging of fibrotic liver disease, pSWE is not widely adopted in clinical workflow. More information on reliability and validity of pSWE systems is needed. Therefore, we performed a phantom study to evaluate the validity and reliability of pSWE with ultrasound systems.
Methods: Validity and reliability of pSWE measurements from three ultrasound systems were evaluated. Measurements were performed on an elasticity phantom with reference elasticities of 7 ± 1 (low) (median ± interquartile range (IQR)), 14 ± 2 (medium) and 26 ± 3 (high) kPa. Measurements were repeated in tenfold for each reference at 2, 3 and 4 cm depth. Results were considered valid when median elasticity ± IQR was between the uncertainty limits (IQR) for each reference elasticity value and reliable when IQR/median < 0.30.
Results: pSWE with the systems provided valid results for all reference elasticities and focal depths, except for overestimation of high reference elasticity at 2 and 4 cm depth for one system (41.5 ± 4.3 and 39.0 ± 1.2 kPa, respectively). Measurements were reliable with a maximum IQR/median of 0.13, well below the guideline of IQR/median < 0.30.
Discussion: The results support the use of pSWE as an alternative to invasive or non-image guided noninvasive techniques for liver fibrotic staging.
Conclusions: pSWE with ultrasound systems from different vendors is valid and reliable and can therefore be implemented to optimize clinical workflow by performing imaging and elastography simultaneously.
期刊介绍:
The Journal of Ultrasound is the official journal of the Italian Society for Ultrasound in Medicine and Biology (SIUMB). The journal publishes original contributions (research and review articles, case reports, technical reports and letters to the editor) on significant advances in clinical diagnostic, interventional and therapeutic applications, clinical techniques, the physics, engineering and technology of ultrasound in medicine and biology, and in cross-sectional diagnostic imaging. The official language of Journal of Ultrasound is English.