{"title":"根据通用标准对 ANDON KD-595 自动上臂式血压计进行验证。","authors":"Kui Liu, Dan Wu, Sen Bing, Lin-Yi Li","doi":"10.1097/MBP.0000000000000686","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To validate the ANDON KD-595 automated upper-arm blood pressure monitor for clinical use and self-measurement blood pressure measurement according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard.</p><p><strong>Methods: </strong>Same left-arm blood pressure was sequentially measured in 90 qualified adult participants and compared with a standard mercury sphygmomanometer. A total of 270 comparison pairs were obtained and analyzed according to the universal standard.</p><p><strong>Results: </strong>For the validation Criterion 1 of the universal standard, the mean ± SD of the differences between the test device and reference blood pressure readings was 0.96 ± 5.35 and 0.82 ± 5.08 mmHg for SBP and DBP, respectively. For Criterion 2, the SDs of the averaged blood pressure differences between the test device and reference blood pressure per subject were 4.84 and 4.64 mmHg (with maximum allowed SDs of 6.87 and 6.89 mmHg) for SBP and DBP, respectively.</p><p><strong>Conclusion: </strong>The ANDON KD-595 automated upper-arm blood pressure monitor passed all the validation requirements according to the AAMI/ESH/ISO Universal Standard and can be recommended for clinical use and self-measurement blood pressure measurement in the general population.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":" ","pages":"103-106"},"PeriodicalIF":1.2000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Validation of the ANDON KD-595 automated upper-arm blood pressure monitor according to the universal standard.\",\"authors\":\"Kui Liu, Dan Wu, Sen Bing, Lin-Yi Li\",\"doi\":\"10.1097/MBP.0000000000000686\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To validate the ANDON KD-595 automated upper-arm blood pressure monitor for clinical use and self-measurement blood pressure measurement according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard.</p><p><strong>Methods: </strong>Same left-arm blood pressure was sequentially measured in 90 qualified adult participants and compared with a standard mercury sphygmomanometer. A total of 270 comparison pairs were obtained and analyzed according to the universal standard.</p><p><strong>Results: </strong>For the validation Criterion 1 of the universal standard, the mean ± SD of the differences between the test device and reference blood pressure readings was 0.96 ± 5.35 and 0.82 ± 5.08 mmHg for SBP and DBP, respectively. For Criterion 2, the SDs of the averaged blood pressure differences between the test device and reference blood pressure per subject were 4.84 and 4.64 mmHg (with maximum allowed SDs of 6.87 and 6.89 mmHg) for SBP and DBP, respectively.</p><p><strong>Conclusion: </strong>The ANDON KD-595 automated upper-arm blood pressure monitor passed all the validation requirements according to the AAMI/ESH/ISO Universal Standard and can be recommended for clinical use and self-measurement blood pressure measurement in the general population.</p>\",\"PeriodicalId\":8950,\"journal\":{\"name\":\"Blood Pressure Monitoring\",\"volume\":\" \",\"pages\":\"103-106\"},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2024-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Blood Pressure Monitoring\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MBP.0000000000000686\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/12/14 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"PERIPHERAL VASCULAR DISEASE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000686","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/12/14 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
Validation of the ANDON KD-595 automated upper-arm blood pressure monitor according to the universal standard.
Objective: To validate the ANDON KD-595 automated upper-arm blood pressure monitor for clinical use and self-measurement blood pressure measurement according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard.
Methods: Same left-arm blood pressure was sequentially measured in 90 qualified adult participants and compared with a standard mercury sphygmomanometer. A total of 270 comparison pairs were obtained and analyzed according to the universal standard.
Results: For the validation Criterion 1 of the universal standard, the mean ± SD of the differences between the test device and reference blood pressure readings was 0.96 ± 5.35 and 0.82 ± 5.08 mmHg for SBP and DBP, respectively. For Criterion 2, the SDs of the averaged blood pressure differences between the test device and reference blood pressure per subject were 4.84 and 4.64 mmHg (with maximum allowed SDs of 6.87 and 6.89 mmHg) for SBP and DBP, respectively.
Conclusion: The ANDON KD-595 automated upper-arm blood pressure monitor passed all the validation requirements according to the AAMI/ESH/ISO Universal Standard and can be recommended for clinical use and self-measurement blood pressure measurement in the general population.
期刊介绍:
Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research.
This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine.
Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.