Depeng Liu, Gang Li, Shuyuan Wang, Zixuan Liu, Yanzhou Wang, Laura Connolly, David E. Usevitch, Guofeng Shen, Kevin Cleary, Iulian Iordachita
{"title":"用于腰椎注射的磁共振条件机器人:开发和初步验证","authors":"Depeng Liu, Gang Li, Shuyuan Wang, Zixuan Liu, Yanzhou Wang, Laura Connolly, David E. Usevitch, Guofeng Shen, Kevin Cleary, Iulian Iordachita","doi":"10.1002/rcs.2618","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Purpose</h3>\n \n <p>This work presents the design and preliminary validation of a Magnetic Resonance (MR) conditional robot for lumbar injection for the treatment of lower back pain.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>This is a 4-degree-of-freedom (DOF) robot that is 200 × 230 × 130 mm3 in volume and has a mass of 0.8 kg. Its lightweight and compact features allow it to be directly affixed to patient's back, establishing a rigid connection, thus reducing positional errors caused by patient movements during treatment.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>To validate the positioning accuracy of the needle by the robot, an electromagnetic (EM) tracking system and a needle with an EM sensor embedded in the tip were used for the free space evaluation with position accuracy of 0.88 ± 0.46 mm and phantom mock insertions using the Loop-X CBCT scanner with target position accuracy of 3.62 ± 0.92 mm.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>Preliminary experiments demonstrated that the proposed robot showed improvements and benefits in its rotation range, flexible needle adjustment, and sensor protection compared with previous and existing systems, offering broader clinical applications.</p>\n </section>\n </div>","PeriodicalId":50311,"journal":{"name":"International Journal of Medical Robotics and Computer Assisted Surgery","volume":null,"pages":null},"PeriodicalIF":2.3000,"publicationDate":"2023-12-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"An magnetic resonance conditional robot for lumbar spinal injection: Development and preliminary validation\",\"authors\":\"Depeng Liu, Gang Li, Shuyuan Wang, Zixuan Liu, Yanzhou Wang, Laura Connolly, David E. Usevitch, Guofeng Shen, Kevin Cleary, Iulian Iordachita\",\"doi\":\"10.1002/rcs.2618\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Purpose</h3>\\n \\n <p>This work presents the design and preliminary validation of a Magnetic Resonance (MR) conditional robot for lumbar injection for the treatment of lower back pain.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>This is a 4-degree-of-freedom (DOF) robot that is 200 × 230 × 130 mm3 in volume and has a mass of 0.8 kg. Its lightweight and compact features allow it to be directly affixed to patient's back, establishing a rigid connection, thus reducing positional errors caused by patient movements during treatment.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>To validate the positioning accuracy of the needle by the robot, an electromagnetic (EM) tracking system and a needle with an EM sensor embedded in the tip were used for the free space evaluation with position accuracy of 0.88 ± 0.46 mm and phantom mock insertions using the Loop-X CBCT scanner with target position accuracy of 3.62 ± 0.92 mm.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusion</h3>\\n \\n <p>Preliminary experiments demonstrated that the proposed robot showed improvements and benefits in its rotation range, flexible needle adjustment, and sensor protection compared with previous and existing systems, offering broader clinical applications.</p>\\n </section>\\n </div>\",\"PeriodicalId\":50311,\"journal\":{\"name\":\"International Journal of Medical Robotics and Computer Assisted Surgery\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.3000,\"publicationDate\":\"2023-12-28\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International Journal of Medical Robotics and Computer Assisted Surgery\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1002/rcs.2618\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"SURGERY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Medical Robotics and Computer Assisted Surgery","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/rcs.2618","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"SURGERY","Score":null,"Total":0}
An magnetic resonance conditional robot for lumbar spinal injection: Development and preliminary validation
Purpose
This work presents the design and preliminary validation of a Magnetic Resonance (MR) conditional robot for lumbar injection for the treatment of lower back pain.
Methods
This is a 4-degree-of-freedom (DOF) robot that is 200 × 230 × 130 mm3 in volume and has a mass of 0.8 kg. Its lightweight and compact features allow it to be directly affixed to patient's back, establishing a rigid connection, thus reducing positional errors caused by patient movements during treatment.
Results
To validate the positioning accuracy of the needle by the robot, an electromagnetic (EM) tracking system and a needle with an EM sensor embedded in the tip were used for the free space evaluation with position accuracy of 0.88 ± 0.46 mm and phantom mock insertions using the Loop-X CBCT scanner with target position accuracy of 3.62 ± 0.92 mm.
Conclusion
Preliminary experiments demonstrated that the proposed robot showed improvements and benefits in its rotation range, flexible needle adjustment, and sensor protection compared with previous and existing systems, offering broader clinical applications.
期刊介绍:
The International Journal of Medical Robotics and Computer Assisted Surgery provides a cross-disciplinary platform for presenting the latest developments in robotics and computer assisted technologies for medical applications. The journal publishes cutting-edge papers and expert reviews, complemented by commentaries, correspondence and conference highlights that stimulate discussion and exchange of ideas. Areas of interest include robotic surgery aids and systems, operative planning tools, medical imaging and visualisation, simulation and navigation, virtual reality, intuitive command and control systems, haptics and sensor technologies. In addition to research and surgical planning studies, the journal welcomes papers detailing clinical trials and applications of computer-assisted workflows and robotic systems in neurosurgery, urology, paediatric, orthopaedic, craniofacial, cardiovascular, thoraco-abdominal, musculoskeletal and visceral surgery. Articles providing critical analysis of clinical trials, assessment of the benefits and risks of the application of these technologies, commenting on ease of use, or addressing surgical education and training issues are also encouraged. The journal aims to foster a community that encompasses medical practitioners, researchers, and engineers and computer scientists developing robotic systems and computational tools in academic and commercial environments, with the intention of promoting and developing these exciting areas of medical technology.