{"title":"在射频诱导肠吻合术中减少组织热损伤的新型电极。","authors":"Xupo Xing, Chengli Song","doi":"10.1080/13645706.2023.2297774","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study aimed to design a novel electrode for reducing tissue thermal damage in radiofrequency-induced intestinal anastomosis.</p><p><strong>Material and methods: </strong>We developed and compared two electrodes (Ring electrode, and Plum electrode with reduced section of the middle fusion area by nearly 80% arising from novel structural design) by performing <i>ex-vivo</i> experiments and finite element analysis.</p><p><strong>Results: </strong>In contrast to the Ring electrode group, slightly higher mean strength is acquired with the tensile force and burst pressure results increasing from 9.7 ± 1.47 N, 84.0 ± 5.99 mmHg to 11.1 ± 1.71 N, 89.4 ± 6.60 mmHg, respectively, as well as a significant reduction in tissue thermal damage for the Plum electrode group, with compression pressure of 20 kPa, RF energy of 120 W and welding duration of 8 s applied to the target regions to achieve anastomosis. Besides, the novel structural design of the Plum electrode can counteract the tension generated by intestinal peristalsis and enhance the biomechanical strength of the anastomotic area. The histological observation showed that the fusion area of the two-layer intestinal tissue is tightly connected with decreased thickness.</p><p><strong>Conclusion: </strong>The novel electrode (Plum electrode) could reduce tissue thermal damage in radiofrequency-induced intestinal anastomosis.</p>","PeriodicalId":18537,"journal":{"name":"Minimally Invasive Therapy & Allied Technologies","volume":" ","pages":"80-89"},"PeriodicalIF":1.7000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A novel electrode for reducing tissue thermal damage in radiofrequency-induced intestinal anastomosis.\",\"authors\":\"Xupo Xing, Chengli Song\",\"doi\":\"10.1080/13645706.2023.2297774\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>This study aimed to design a novel electrode for reducing tissue thermal damage in radiofrequency-induced intestinal anastomosis.</p><p><strong>Material and methods: </strong>We developed and compared two electrodes (Ring electrode, and Plum electrode with reduced section of the middle fusion area by nearly 80% arising from novel structural design) by performing <i>ex-vivo</i> experiments and finite element analysis.</p><p><strong>Results: </strong>In contrast to the Ring electrode group, slightly higher mean strength is acquired with the tensile force and burst pressure results increasing from 9.7 ± 1.47 N, 84.0 ± 5.99 mmHg to 11.1 ± 1.71 N, 89.4 ± 6.60 mmHg, respectively, as well as a significant reduction in tissue thermal damage for the Plum electrode group, with compression pressure of 20 kPa, RF energy of 120 W and welding duration of 8 s applied to the target regions to achieve anastomosis. Besides, the novel structural design of the Plum electrode can counteract the tension generated by intestinal peristalsis and enhance the biomechanical strength of the anastomotic area. The histological observation showed that the fusion area of the two-layer intestinal tissue is tightly connected with decreased thickness.</p><p><strong>Conclusion: </strong>The novel electrode (Plum electrode) could reduce tissue thermal damage in radiofrequency-induced intestinal anastomosis.</p>\",\"PeriodicalId\":18537,\"journal\":{\"name\":\"Minimally Invasive Therapy & Allied Technologies\",\"volume\":\" \",\"pages\":\"80-89\"},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2024-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Minimally Invasive Therapy & Allied Technologies\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1080/13645706.2023.2297774\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/12/26 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"SURGERY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Minimally Invasive Therapy & Allied Technologies","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/13645706.2023.2297774","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/12/26 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"SURGERY","Score":null,"Total":0}
A novel electrode for reducing tissue thermal damage in radiofrequency-induced intestinal anastomosis.
Purpose: This study aimed to design a novel electrode for reducing tissue thermal damage in radiofrequency-induced intestinal anastomosis.
Material and methods: We developed and compared two electrodes (Ring electrode, and Plum electrode with reduced section of the middle fusion area by nearly 80% arising from novel structural design) by performing ex-vivo experiments and finite element analysis.
Results: In contrast to the Ring electrode group, slightly higher mean strength is acquired with the tensile force and burst pressure results increasing from 9.7 ± 1.47 N, 84.0 ± 5.99 mmHg to 11.1 ± 1.71 N, 89.4 ± 6.60 mmHg, respectively, as well as a significant reduction in tissue thermal damage for the Plum electrode group, with compression pressure of 20 kPa, RF energy of 120 W and welding duration of 8 s applied to the target regions to achieve anastomosis. Besides, the novel structural design of the Plum electrode can counteract the tension generated by intestinal peristalsis and enhance the biomechanical strength of the anastomotic area. The histological observation showed that the fusion area of the two-layer intestinal tissue is tightly connected with decreased thickness.
Conclusion: The novel electrode (Plum electrode) could reduce tissue thermal damage in radiofrequency-induced intestinal anastomosis.
期刊介绍:
Minimally Invasive Therapy and Allied Technologies (MITAT) is an international forum for endoscopic surgeons, interventional radiologists and industrial instrument manufacturers. It is the official journal of the Society for Medical Innovation and Technology (SMIT) whose membership includes representatives from a broad spectrum of medical specialities, instrument manufacturing and research. The journal brings the latest developments and innovations in minimally invasive therapy to its readers. What makes Minimally Invasive Therapy and Allied Technologies unique is that we publish one or two special issues each year, which are devoted to a specific theme. Key topics covered by the journal include: interventional radiology, endoscopic surgery, imaging technology, manipulators and robotics for surgery and education and training for MIS.