评价球囊可膨胀血管支架的试验方法。径向强度和刚度的测量

Q4 Engineering
Christoph Brandt-Wunderlich, Wolfram Schmidt, Alexander Thiesen, Klaus-Peter Schmitz, Falk Reinhardt, Michael Stiehm, Stefan Siewert
{"title":"评价球囊可膨胀血管支架的试验方法。径向强度和刚度的测量","authors":"Christoph Brandt-Wunderlich, Wolfram Schmidt, Alexander Thiesen, Klaus-Peter Schmitz, Falk Reinhardt, Michael Stiehm, Stefan Siewert","doi":"10.1515/cdbme-2023-1104","DOIUrl":null,"url":null,"abstract":"Abstract The radial strength of balloon expandable stents represents a key property for a successful recanalization of sclerotic blood vessels. This study focuses on the most commonly used method for investigation of radial strength and radial stiffness via segmented head test setup. A custom made user software was used for evaluation of the radial force curves considering requirements of international standards such as ASTM F3067. Contributing factors during measurement such as friction and test setup deformation as well as the single cycle and multi cycle approach were addressed and discussed.","PeriodicalId":10739,"journal":{"name":"Current Directions in Biomedical Engineering","volume":"58 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2023-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Test methods for evaluation of balloon expandable vascular stents – measurement of radial strength and stiffness\",\"authors\":\"Christoph Brandt-Wunderlich, Wolfram Schmidt, Alexander Thiesen, Klaus-Peter Schmitz, Falk Reinhardt, Michael Stiehm, Stefan Siewert\",\"doi\":\"10.1515/cdbme-2023-1104\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Abstract The radial strength of balloon expandable stents represents a key property for a successful recanalization of sclerotic blood vessels. This study focuses on the most commonly used method for investigation of radial strength and radial stiffness via segmented head test setup. A custom made user software was used for evaluation of the radial force curves considering requirements of international standards such as ASTM F3067. Contributing factors during measurement such as friction and test setup deformation as well as the single cycle and multi cycle approach were addressed and discussed.\",\"PeriodicalId\":10739,\"journal\":{\"name\":\"Current Directions in Biomedical Engineering\",\"volume\":\"58 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Current Directions in Biomedical Engineering\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1515/cdbme-2023-1104\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Engineering\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Current Directions in Biomedical Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1515/cdbme-2023-1104","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Engineering","Score":null,"Total":0}
引用次数: 0

摘要

球囊可膨胀支架的径向强度是硬化血管再通成功的关键特性。本研究的重点是通过分段头部试验装置研究最常用的径向强度和径向刚度的方法。考虑到ASTM F3067等国际标准的要求,使用定制的用户软件来评估径向力曲线。分析和讨论了测量过程中的影响因素,如摩擦和试验装置变形,以及单循环和多循环方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Test methods for evaluation of balloon expandable vascular stents – measurement of radial strength and stiffness
Abstract The radial strength of balloon expandable stents represents a key property for a successful recanalization of sclerotic blood vessels. This study focuses on the most commonly used method for investigation of radial strength and radial stiffness via segmented head test setup. A custom made user software was used for evaluation of the radial force curves considering requirements of international standards such as ASTM F3067. Contributing factors during measurement such as friction and test setup deformation as well as the single cycle and multi cycle approach were addressed and discussed.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Current Directions in Biomedical Engineering
Current Directions in Biomedical Engineering Engineering-Biomedical Engineering
CiteScore
0.90
自引率
0.00%
发文量
239
审稿时长
14 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信