{"title":"相同抛光方法对传统、CAD/CAM铣削和3D打印义齿基托材料表面粗糙度影响的比较","authors":"DEMİRKOL, Didem\n , TUĞUT, Faik\n ","doi":"10.7126/cumudj.1317851","DOIUrl":null,"url":null,"abstract":"Objective: The aim of this study was to evaluate the effect of both the same polishing method and those with and without thermal aging on the surface roughness of conventional, CAD/CAM milling and 3D printing denture base materials.
 Materials and Methods: A total of 30 round shaped specimens were obtained by 3 different methods (n=10): Conventional, CAD/CAM milling and 3D-Printing. After applying the same polishing technique to all groups, surface roughness values were measured. Profilometer device was used for surface roughness measurement. Then, after the thermal aging of all samples, surface roughness values were measured and the roughness values between no-thermocycling and thermocycling were compared. Tukey, Mann Whitney U and Kruskal Wallis tests were used statistically. P values of ≤ 0.05 were considered significant. 
 Results: As a result of the same polishing process, there was a difference in surface roughness in all groups. While the highest surface roughness values were seen in 3D-printing, the lowest roughness value was seen in the CAD/CAM milling and was statistically significant (p<0.05). Thermocycling did not show a statistically significant difference in surface roughness (p>0.05).
 Conclusions: The same polishing process caused different surface roughness values in the denture base materials obtained with different methods, and the lowest surface roughness value was seen in the CAD/CAM milling.","PeriodicalId":10781,"journal":{"name":"Cumhuriyet Dental Journal","volume":"68 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2023-09-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Comparison of the Effect of the Same Polishing Method on the Surface Roughness of Conventional, CAD/CAM Milling and 3D Printing Denture Base Materials\",\"authors\":\"DEMİRKOL, Didem\\n , TUĞUT, Faik\\n \",\"doi\":\"10.7126/cumudj.1317851\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective: The aim of this study was to evaluate the effect of both the same polishing method and those with and without thermal aging on the surface roughness of conventional, CAD/CAM milling and 3D printing denture base materials.
 Materials and Methods: A total of 30 round shaped specimens were obtained by 3 different methods (n=10): Conventional, CAD/CAM milling and 3D-Printing. After applying the same polishing technique to all groups, surface roughness values were measured. Profilometer device was used for surface roughness measurement. Then, after the thermal aging of all samples, surface roughness values were measured and the roughness values between no-thermocycling and thermocycling were compared. Tukey, Mann Whitney U and Kruskal Wallis tests were used statistically. P values of ≤ 0.05 were considered significant. 
 Results: As a result of the same polishing process, there was a difference in surface roughness in all groups. While the highest surface roughness values were seen in 3D-printing, the lowest roughness value was seen in the CAD/CAM milling and was statistically significant (p<0.05). Thermocycling did not show a statistically significant difference in surface roughness (p>0.05).
 Conclusions: The same polishing process caused different surface roughness values in the denture base materials obtained with different methods, and the lowest surface roughness value was seen in the CAD/CAM milling.\",\"PeriodicalId\":10781,\"journal\":{\"name\":\"Cumhuriyet Dental Journal\",\"volume\":\"68 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-09-29\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Cumhuriyet Dental Journal\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.7126/cumudj.1317851\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Dentistry\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cumhuriyet Dental Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.7126/cumudj.1317851","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Dentistry","Score":null,"Total":0}
Comparison of the Effect of the Same Polishing Method on the Surface Roughness of Conventional, CAD/CAM Milling and 3D Printing Denture Base Materials
Objective: The aim of this study was to evaluate the effect of both the same polishing method and those with and without thermal aging on the surface roughness of conventional, CAD/CAM milling and 3D printing denture base materials.
Materials and Methods: A total of 30 round shaped specimens were obtained by 3 different methods (n=10): Conventional, CAD/CAM milling and 3D-Printing. After applying the same polishing technique to all groups, surface roughness values were measured. Profilometer device was used for surface roughness measurement. Then, after the thermal aging of all samples, surface roughness values were measured and the roughness values between no-thermocycling and thermocycling were compared. Tukey, Mann Whitney U and Kruskal Wallis tests were used statistically. P values of ≤ 0.05 were considered significant.
Results: As a result of the same polishing process, there was a difference in surface roughness in all groups. While the highest surface roughness values were seen in 3D-printing, the lowest roughness value was seen in the CAD/CAM milling and was statistically significant (p<0.05). Thermocycling did not show a statistically significant difference in surface roughness (p>0.05).
Conclusions: The same polishing process caused different surface roughness values in the denture base materials obtained with different methods, and the lowest surface roughness value was seen in the CAD/CAM milling.