新研制的根管长度测量装置的电子测量。根尖孔直径、k锉大小和根管灌洗剂的影响。

Dentistry in Japan Pub Date : 1990-01-01
T Saito, Y Yamashita
{"title":"新研制的根管长度测量装置的电子测量。根尖孔直径、k锉大小和根管灌洗剂的影响。","authors":"T Saito,&nbsp;Y Yamashita","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>A new electronic root canal measuring device called \"ENDEX\" which is able to determine the root canal length under moistened conditions using a relative value of different alternative electric currents is discussed in this study. This study is designed to investigate the efficiencies of the new device, in vitro, with respect to change in the root canal environment; namely the influence of the size of K-file, the diameter of the apical foramen and the various root canal irrigants. According to the present study, it was confirmed that the location of the apical foramen was clearly detected under moistened root canal with saline solution using any size of K-file. The locations of the tip of K-files were 0.5 mm inside the apical foramen when the relative value displayed 0.3 Volt, so that the apical constriction could not disturb during root canal measurement. One of the big problem with prior electrical measurement devices was not being able to determine the canal length under moistened conditions. However with this improved device, intracanal irrigants, such as saline, 5% NaOC1, 14% EDTA and 3% H2O2 did not interfere in detecting apical foramen irregardless of the size of K-file and the size of the apical foramen.</p>","PeriodicalId":75798,"journal":{"name":"Dentistry in Japan","volume":"27 1","pages":"65-72"},"PeriodicalIF":0.0000,"publicationDate":"1990-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Electronic determination of root canal length by newly developed measuring device. Influences of the diameter of apical foramen, the size of K-file and the root canal irrigants.\",\"authors\":\"T Saito,&nbsp;Y Yamashita\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>A new electronic root canal measuring device called \\\"ENDEX\\\" which is able to determine the root canal length under moistened conditions using a relative value of different alternative electric currents is discussed in this study. This study is designed to investigate the efficiencies of the new device, in vitro, with respect to change in the root canal environment; namely the influence of the size of K-file, the diameter of the apical foramen and the various root canal irrigants. According to the present study, it was confirmed that the location of the apical foramen was clearly detected under moistened root canal with saline solution using any size of K-file. The locations of the tip of K-files were 0.5 mm inside the apical foramen when the relative value displayed 0.3 Volt, so that the apical constriction could not disturb during root canal measurement. One of the big problem with prior electrical measurement devices was not being able to determine the canal length under moistened conditions. However with this improved device, intracanal irrigants, such as saline, 5% NaOC1, 14% EDTA and 3% H2O2 did not interfere in detecting apical foramen irregardless of the size of K-file and the size of the apical foramen.</p>\",\"PeriodicalId\":75798,\"journal\":{\"name\":\"Dentistry in Japan\",\"volume\":\"27 1\",\"pages\":\"65-72\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1990-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Dentistry in Japan\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Dentistry in Japan","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

本研究讨论了一种新型的电子根管测量装置“ENDEX”,该装置能够利用不同交变电流的相对值来确定潮湿条件下的根管长度。本研究旨在研究新装置在体外对根管环境变化的影响;即k锉大小、根尖孔直径和不同根管冲洗剂的影响。本研究证实,使用任意大小的k锉,在用生理盐水湿润根管的情况下,可以清楚地检测根尖孔的位置。当相对值为0.3伏特时,k锉的尖端位置在根尖孔内0.5 mm,这样在测量根管时不会干扰根尖收缩。以前的电子测量设备的一个大问题是不能确定潮湿条件下的运河长度。然而,使用这种改进的装置,无论k锉的大小和根尖孔的大小,盐水、5% NaOC1、14% EDTA和3% H2O2等管内冲洗液都不会干扰根尖孔的检测。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Electronic determination of root canal length by newly developed measuring device. Influences of the diameter of apical foramen, the size of K-file and the root canal irrigants.

A new electronic root canal measuring device called "ENDEX" which is able to determine the root canal length under moistened conditions using a relative value of different alternative electric currents is discussed in this study. This study is designed to investigate the efficiencies of the new device, in vitro, with respect to change in the root canal environment; namely the influence of the size of K-file, the diameter of the apical foramen and the various root canal irrigants. According to the present study, it was confirmed that the location of the apical foramen was clearly detected under moistened root canal with saline solution using any size of K-file. The locations of the tip of K-files were 0.5 mm inside the apical foramen when the relative value displayed 0.3 Volt, so that the apical constriction could not disturb during root canal measurement. One of the big problem with prior electrical measurement devices was not being able to determine the canal length under moistened conditions. However with this improved device, intracanal irrigants, such as saline, 5% NaOC1, 14% EDTA and 3% H2O2 did not interfere in detecting apical foramen irregardless of the size of K-file and the size of the apical foramen.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信