{"title":"生物化学实验室两种不同品牌采血管的性能比较","authors":"A. V. Ozdemir, S. Mızrak, Metin Demir","doi":"10.18502/ijml.v9i3.10906","DOIUrl":null,"url":null,"abstract":"Background and Aims: Blood collection tubes (BCT) with gel separators are diagnostic devices that can affect all laboratory processes. In this study, the aim was to compare the performance of a new brand of BCT with the sample tubes that are well-known and widely used in the market. \nMaterials and Methods: Blood samples were taken from 50 healthy volunteers to two brands of BCTs (KWS and BD Vacutainer SST II Advance) by providing standard conditions. The nine most requested test parameters were determined and analyzed simultaneously. The results were analyzed using the SPSS 21 software. \nResults: There was no significant difference in the results of most tests of the new brand tubes compared to the Becton-Dickinson brand tube, and the bias calculated for the tests except calcium and potassium was detected to be lower than the desirable bias. \nConclusion: It was determined that the test parameters analyzed in both tubes were in statistical agreement. The percentage of bias value of calcium and potassium tests in KWS branded tubes is higher than the desired value. In addition, constant and proportional errors were determined for the calcium test. These test parameters should be analyzed in larger patient groups. Therefore, KWS tubes can be accepted for routine laboratory operations.","PeriodicalId":183358,"journal":{"name":"International Journal of Medical Laboratory","volume":"1 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2022-10-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Performance Comparison of Two Different Branded Blood Collection Tubes Used in Biochemistry Laboratory\",\"authors\":\"A. V. Ozdemir, S. Mızrak, Metin Demir\",\"doi\":\"10.18502/ijml.v9i3.10906\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Background and Aims: Blood collection tubes (BCT) with gel separators are diagnostic devices that can affect all laboratory processes. In this study, the aim was to compare the performance of a new brand of BCT with the sample tubes that are well-known and widely used in the market. \\nMaterials and Methods: Blood samples were taken from 50 healthy volunteers to two brands of BCTs (KWS and BD Vacutainer SST II Advance) by providing standard conditions. The nine most requested test parameters were determined and analyzed simultaneously. The results were analyzed using the SPSS 21 software. \\nResults: There was no significant difference in the results of most tests of the new brand tubes compared to the Becton-Dickinson brand tube, and the bias calculated for the tests except calcium and potassium was detected to be lower than the desirable bias. \\nConclusion: It was determined that the test parameters analyzed in both tubes were in statistical agreement. The percentage of bias value of calcium and potassium tests in KWS branded tubes is higher than the desired value. In addition, constant and proportional errors were determined for the calcium test. These test parameters should be analyzed in larger patient groups. Therefore, KWS tubes can be accepted for routine laboratory operations.\",\"PeriodicalId\":183358,\"journal\":{\"name\":\"International Journal of Medical Laboratory\",\"volume\":\"1 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-10-18\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International Journal of Medical Laboratory\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.18502/ijml.v9i3.10906\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Medical Laboratory","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.18502/ijml.v9i3.10906","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
背景和目的:带有凝胶分离器的采血管(BCT)是一种可以影响所有实验室流程的诊断设备。在这项研究中,目的是比较一个新品牌的BCT与市场上知名和广泛使用的样管的性能。材料与方法:选取50名健康志愿者,分别使用两种品牌的bct (KWS和BD Vacutainer SST II Advance)进行标准条件下的血液采集。同时确定并分析了9个最需要的试验参数。采用SPSS 21软件对结果进行分析。结果:与Becton-Dickinson品牌试管相比,新品牌试管的大多数测试结果无显著差异,除钙和钾外,其他测试的计算偏差均低于理想偏差。结论:两种试管的试验参数分析具有统计学上的一致性。KWS品牌试管中钙钾试验的偏置值百分比高于期望值。此外,确定了钙试验的恒定误差和比例误差。这些测试参数应在更大的患者群体中进行分析。因此,KWS试管可用于常规实验室操作。
Performance Comparison of Two Different Branded Blood Collection Tubes Used in Biochemistry Laboratory
Background and Aims: Blood collection tubes (BCT) with gel separators are diagnostic devices that can affect all laboratory processes. In this study, the aim was to compare the performance of a new brand of BCT with the sample tubes that are well-known and widely used in the market.
Materials and Methods: Blood samples were taken from 50 healthy volunteers to two brands of BCTs (KWS and BD Vacutainer SST II Advance) by providing standard conditions. The nine most requested test parameters were determined and analyzed simultaneously. The results were analyzed using the SPSS 21 software.
Results: There was no significant difference in the results of most tests of the new brand tubes compared to the Becton-Dickinson brand tube, and the bias calculated for the tests except calcium and potassium was detected to be lower than the desirable bias.
Conclusion: It was determined that the test parameters analyzed in both tubes were in statistical agreement. The percentage of bias value of calcium and potassium tests in KWS branded tubes is higher than the desired value. In addition, constant and proportional errors were determined for the calcium test. These test parameters should be analyzed in larger patient groups. Therefore, KWS tubes can be accepted for routine laboratory operations.