{"title":"[血钙离子含量:对不同方法因素的依赖]。","authors":"K Kölling","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>In the present study the influence of different methodical factors (analyzer with ion-selective electrodes used, anticoagulation of blood samples with heparin, storing of samples at different temperatures) on the concentration of ionized calcium in blood was investigated. It could be shown that nowadays' systems for analysis of ionized calcium (ORION SS-20, AVL 980, NOVA 2 and 8, ICA 1) yield reproducible values for human as well as canine blood. System-related differences point to a varying degree of interferences on the ion-selective electrodes used. The concentration of ionized calcium is strongly dependent on different preanalytic factors. Heparin leads to a concentration-related reduction of calcium ions in blood. This negative impact of heparin can be overcome by substitution of the solvent for CaCl2 (2.5 mmol.l-1) without meaningful falsification of the measured values. At 4 degrees C blood samples can be stored without detectable changes in Ca++ concentration for 8 hours.</p>","PeriodicalId":23904,"journal":{"name":"Zentralblatt fur Veterinarmedizin. Reihe A","volume":"38 8","pages":"629-38"},"PeriodicalIF":0.0000,"publicationDate":"1991-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[The ionized calcium fraction of the blood: dependence on different methodical factors].\",\"authors\":\"K Kölling\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>In the present study the influence of different methodical factors (analyzer with ion-selective electrodes used, anticoagulation of blood samples with heparin, storing of samples at different temperatures) on the concentration of ionized calcium in blood was investigated. It could be shown that nowadays' systems for analysis of ionized calcium (ORION SS-20, AVL 980, NOVA 2 and 8, ICA 1) yield reproducible values for human as well as canine blood. System-related differences point to a varying degree of interferences on the ion-selective electrodes used. The concentration of ionized calcium is strongly dependent on different preanalytic factors. Heparin leads to a concentration-related reduction of calcium ions in blood. This negative impact of heparin can be overcome by substitution of the solvent for CaCl2 (2.5 mmol.l-1) without meaningful falsification of the measured values. At 4 degrees C blood samples can be stored without detectable changes in Ca++ concentration for 8 hours.</p>\",\"PeriodicalId\":23904,\"journal\":{\"name\":\"Zentralblatt fur Veterinarmedizin. Reihe A\",\"volume\":\"38 8\",\"pages\":\"629-38\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1991-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Zentralblatt fur Veterinarmedizin. Reihe A\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Zentralblatt fur Veterinarmedizin. Reihe A","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
本研究考察了不同方法因素(离子选择电极分析仪、肝素抗凝血、不同温度保存)对血钙离子浓度的影响。可以看出,目前用于分析电离钙的系统(ORION SS-20, AVL 980, NOVA 2和8,ICA 1)对人血和犬血都有可重复性。系统相关的差异指向不同程度的离子选择电极上的干扰。离子钙的浓度强烈依赖于不同的分析前因素。肝素导致血液中钙离子浓度相关的减少。肝素的这种负面影响可以通过用CaCl2 (2.5 mmol.l-1)代替溶剂来克服,而不会对测量值产生有意义的伪造。在4℃条件下,血液样本可保存8小时,ca2 +浓度无明显变化。
[The ionized calcium fraction of the blood: dependence on different methodical factors].
In the present study the influence of different methodical factors (analyzer with ion-selective electrodes used, anticoagulation of blood samples with heparin, storing of samples at different temperatures) on the concentration of ionized calcium in blood was investigated. It could be shown that nowadays' systems for analysis of ionized calcium (ORION SS-20, AVL 980, NOVA 2 and 8, ICA 1) yield reproducible values for human as well as canine blood. System-related differences point to a varying degree of interferences on the ion-selective electrodes used. The concentration of ionized calcium is strongly dependent on different preanalytic factors. Heparin leads to a concentration-related reduction of calcium ions in blood. This negative impact of heparin can be overcome by substitution of the solvent for CaCl2 (2.5 mmol.l-1) without meaningful falsification of the measured values. At 4 degrees C blood samples can be stored without detectable changes in Ca++ concentration for 8 hours.