超声浸泡治疗技术在幻体中的应用评价

A. P. M. Moraes, B. S. Fagundes, L. S. de Lima, T. P. Omena, J. F. Costa, R. M. Costa, W. Pereira
{"title":"超声浸泡治疗技术在幻体中的应用评价","authors":"A. P. M. Moraes, B. S. Fagundes, L. S. de Lima, T. P. Omena, J. F. Costa, R. M. Costa, W. Pereira","doi":"10.1109/PAHCE.2014.6849628","DOIUrl":null,"url":null,"abstract":"The present work evaluate the heating pattern of therapeutic ultrasound using immersion technique in a phantom of fat tissue made of silicone in different distances. Five tests were performed in each of six positions: 0, 1, 2, 3, 5 and 10 cm apart from the transducer surface, at 1 MHz, continuous mode and effective intensity 1.05 W·cm-2 for 180s. In the most superficial thermocouple, there was decrease in the temperature from the distance 0 cm to 1 cm and the following distances (2, 3, 5 and 10 cm) also experienced reduction above 90% of the average value at 0 cm. In thermocouples 2, 3 and 4 there was no significant difference with increasing distance from the transducer to the phantom.","PeriodicalId":196438,"journal":{"name":"2014 Pan American Health Care Exchanges (PAHCE)","volume":"3 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2014-04-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluation of the therapeutic ultrasound immersion technique applied in phantom\",\"authors\":\"A. P. M. Moraes, B. S. Fagundes, L. S. de Lima, T. P. Omena, J. F. Costa, R. M. Costa, W. Pereira\",\"doi\":\"10.1109/PAHCE.2014.6849628\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The present work evaluate the heating pattern of therapeutic ultrasound using immersion technique in a phantom of fat tissue made of silicone in different distances. Five tests were performed in each of six positions: 0, 1, 2, 3, 5 and 10 cm apart from the transducer surface, at 1 MHz, continuous mode and effective intensity 1.05 W·cm-2 for 180s. In the most superficial thermocouple, there was decrease in the temperature from the distance 0 cm to 1 cm and the following distances (2, 3, 5 and 10 cm) also experienced reduction above 90% of the average value at 0 cm. In thermocouples 2, 3 and 4 there was no significant difference with increasing distance from the transducer to the phantom.\",\"PeriodicalId\":196438,\"journal\":{\"name\":\"2014 Pan American Health Care Exchanges (PAHCE)\",\"volume\":\"3 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2014-04-07\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2014 Pan American Health Care Exchanges (PAHCE)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/PAHCE.2014.6849628\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2014 Pan American Health Care Exchanges (PAHCE)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/PAHCE.2014.6849628","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

本研究在不同距离的硅胶脂肪组织体中,利用浸没技术评价了治疗性超声的加热模式。在距离换能器表面0、1、2、3、5和10 cm的6个位置分别进行5次测试,频率为1 MHz,连续模式,有效强度为1.05 W·cm-2,持续180s。在最浅层的热电偶中,从0 cm到1 cm的距离温度下降,并且接下来的距离(2、3、5和10 cm)也经历了高于0 cm平均值90%的下降。在热电偶2、3和4中,随着传感器到模体距离的增加,没有显著差异。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Evaluation of the therapeutic ultrasound immersion technique applied in phantom
The present work evaluate the heating pattern of therapeutic ultrasound using immersion technique in a phantom of fat tissue made of silicone in different distances. Five tests were performed in each of six positions: 0, 1, 2, 3, 5 and 10 cm apart from the transducer surface, at 1 MHz, continuous mode and effective intensity 1.05 W·cm-2 for 180s. In the most superficial thermocouple, there was decrease in the temperature from the distance 0 cm to 1 cm and the following distances (2, 3, 5 and 10 cm) also experienced reduction above 90% of the average value at 0 cm. In thermocouples 2, 3 and 4 there was no significant difference with increasing distance from the transducer to the phantom.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信