{"title":"用于微运动研究的线性位移传感器的校准","authors":"M. DiSilvestro, T. Dietz","doi":"10.1109/SFICON.2004.1287161","DOIUrl":null,"url":null,"abstract":"This paper describes a calibration method for linear displacement sensors that includes quantifiable errors on the order of 1 /spl mu/m, and that has been adopted in an effort to ensure accurate measurements of micromotion within orthopaedic implants during simulated use. This method results in a calibrated linear displacement sensor system with known quantifiable error. For the components used in this study, this error is on the order of 1 /spl mu/m. It is suggested that this or a similar method be used for calibration of linear displacement sensors to be used to measure micromotion of orthopaedic implants to ensure accuracy and reliability of data and interpreted results.","PeriodicalId":381233,"journal":{"name":"ISA/IEEE Sensors for Industry Conference, 2004. Proceedings the","volume":"133 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2004-08-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"3","resultStr":"{\"title\":\"Calibration of linear displacement sensors for micromotion studies\",\"authors\":\"M. DiSilvestro, T. Dietz\",\"doi\":\"10.1109/SFICON.2004.1287161\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This paper describes a calibration method for linear displacement sensors that includes quantifiable errors on the order of 1 /spl mu/m, and that has been adopted in an effort to ensure accurate measurements of micromotion within orthopaedic implants during simulated use. This method results in a calibrated linear displacement sensor system with known quantifiable error. For the components used in this study, this error is on the order of 1 /spl mu/m. It is suggested that this or a similar method be used for calibration of linear displacement sensors to be used to measure micromotion of orthopaedic implants to ensure accuracy and reliability of data and interpreted results.\",\"PeriodicalId\":381233,\"journal\":{\"name\":\"ISA/IEEE Sensors for Industry Conference, 2004. Proceedings the\",\"volume\":\"133 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2004-08-24\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"3\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"ISA/IEEE Sensors for Industry Conference, 2004. Proceedings the\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/SFICON.2004.1287161\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"ISA/IEEE Sensors for Industry Conference, 2004. Proceedings the","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/SFICON.2004.1287161","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Calibration of linear displacement sensors for micromotion studies
This paper describes a calibration method for linear displacement sensors that includes quantifiable errors on the order of 1 /spl mu/m, and that has been adopted in an effort to ensure accurate measurements of micromotion within orthopaedic implants during simulated use. This method results in a calibrated linear displacement sensor system with known quantifiable error. For the components used in this study, this error is on the order of 1 /spl mu/m. It is suggested that this or a similar method be used for calibration of linear displacement sensors to be used to measure micromotion of orthopaedic implants to ensure accuracy and reliability of data and interpreted results.