{"title":"医院电磁干扰环境初步调查","authors":"P. Boisvert, B. Segal, T. Pavlasek, S. Rétfalvi, A. Sebe, P. Caron","doi":"10.1109/ISEMC.1991.148219","DOIUrl":null,"url":null,"abstract":"Electric field levels were predicted using a free-space propagation model, and then measured in three major hospitals in downtown Montreal. The measurements were performed using industry-standard techniques and novel broadband, omnidirectional, triaxial, electrically small (BOTES) techniques. Although the absolute values of predictions were often unreliable, they acceptably predicted the type of electromagnetic environment observed at different hospitals. Results from both sets of measurements were closely related, suggesting continued future development of lower-cost BOTES measurement techniques. None of the fields encountered were above the limit prescribed by the Food and Drug Administration (FDA) susceptibility standard for medical devices. Correlations of survey results and of previous reports of device malfunctions in the rooms surveyed suggest that the FDA standard should be mandatory rather than voluntary.<<ETX>>","PeriodicalId":243730,"journal":{"name":"IEEE 1991 International Symposium on Electromagnetic Compatibility","volume":"10 47 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1991-08-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"21","resultStr":"{\"title\":\"Preliminary survey of the electromagnetic interference environment in hospitals\",\"authors\":\"P. Boisvert, B. Segal, T. Pavlasek, S. Rétfalvi, A. Sebe, P. Caron\",\"doi\":\"10.1109/ISEMC.1991.148219\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Electric field levels were predicted using a free-space propagation model, and then measured in three major hospitals in downtown Montreal. The measurements were performed using industry-standard techniques and novel broadband, omnidirectional, triaxial, electrically small (BOTES) techniques. Although the absolute values of predictions were often unreliable, they acceptably predicted the type of electromagnetic environment observed at different hospitals. Results from both sets of measurements were closely related, suggesting continued future development of lower-cost BOTES measurement techniques. None of the fields encountered were above the limit prescribed by the Food and Drug Administration (FDA) susceptibility standard for medical devices. Correlations of survey results and of previous reports of device malfunctions in the rooms surveyed suggest that the FDA standard should be mandatory rather than voluntary.<<ETX>>\",\"PeriodicalId\":243730,\"journal\":{\"name\":\"IEEE 1991 International Symposium on Electromagnetic Compatibility\",\"volume\":\"10 47 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1991-08-12\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"21\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"IEEE 1991 International Symposium on Electromagnetic Compatibility\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/ISEMC.1991.148219\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE 1991 International Symposium on Electromagnetic Compatibility","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ISEMC.1991.148219","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 21

摘要

利用自由空间传播模型预测电场水平,然后在蒙特利尔市中心的三家主要医院进行测量。测量采用工业标准技术和新型宽带、全向、三轴、电小(BOTES)技术进行。虽然预测的绝对值通常是不可靠的,但它们可以接受地预测在不同医院观察到的电磁环境类型。两组测量结果密切相关,表明未来将继续发展低成本的BOTES测量技术。所遇到的所有领域均未超过美国食品和药物管理局(FDA)医疗器械敏感性标准规定的限制。调查结果的相关性和先前调查房间设备故障的报告表明,FDA的标准应该是强制性的,而不是自愿的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Preliminary survey of the electromagnetic interference environment in hospitals
Electric field levels were predicted using a free-space propagation model, and then measured in three major hospitals in downtown Montreal. The measurements were performed using industry-standard techniques and novel broadband, omnidirectional, triaxial, electrically small (BOTES) techniques. Although the absolute values of predictions were often unreliable, they acceptably predicted the type of electromagnetic environment observed at different hospitals. Results from both sets of measurements were closely related, suggesting continued future development of lower-cost BOTES measurement techniques. None of the fields encountered were above the limit prescribed by the Food and Drug Administration (FDA) susceptibility standard for medical devices. Correlations of survey results and of previous reports of device malfunctions in the rooms surveyed suggest that the FDA standard should be mandatory rather than voluntary.<>
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信