{"title":"采用180nm技术的心电监护系统","authors":"Kumari Srashti Singh, P. Bansod, R. Gamad","doi":"10.1109/IEMENTech48150.2019.8981126","DOIUrl":null,"url":null,"abstract":"This paper present analog front end comprising amplifier suitable for ECG acquisition. It is designed using 180nm CMOS SCL technology using Cadence Virtuoso environment with 1.8V power supply. ECG monitoring setup consists of protection circuit, an analog front end pre-amplifier, buffer, low pass filter and main amplifier. Protection circuit is used for protection from voltage surges which can occur between electrodes in electrocardiogram, during the application of defibrillator or due to electrostatic discharge (ESD) event. The ECG signal is amplified in stages, instrumentation amplifier and main amplifier. Low pass filter is designed for noise cancellation and buffer is used for impedance matching between two stages of ECG front end.","PeriodicalId":243805,"journal":{"name":"2019 3rd International Conference on Electronics, Materials Engineering & Nano-Technology (IEMENTech)","volume":"54 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2019-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"4","resultStr":"{\"title\":\"ECG Monitoring System using180nm Technology\",\"authors\":\"Kumari Srashti Singh, P. Bansod, R. Gamad\",\"doi\":\"10.1109/IEMENTech48150.2019.8981126\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This paper present analog front end comprising amplifier suitable for ECG acquisition. It is designed using 180nm CMOS SCL technology using Cadence Virtuoso environment with 1.8V power supply. ECG monitoring setup consists of protection circuit, an analog front end pre-amplifier, buffer, low pass filter and main amplifier. Protection circuit is used for protection from voltage surges which can occur between electrodes in electrocardiogram, during the application of defibrillator or due to electrostatic discharge (ESD) event. The ECG signal is amplified in stages, instrumentation amplifier and main amplifier. Low pass filter is designed for noise cancellation and buffer is used for impedance matching between two stages of ECG front end.\",\"PeriodicalId\":243805,\"journal\":{\"name\":\"2019 3rd International Conference on Electronics, Materials Engineering & Nano-Technology (IEMENTech)\",\"volume\":\"54 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2019-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"4\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2019 3rd International Conference on Electronics, Materials Engineering & Nano-Technology (IEMENTech)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/IEMENTech48150.2019.8981126\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2019 3rd International Conference on Electronics, Materials Engineering & Nano-Technology (IEMENTech)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/IEMENTech48150.2019.8981126","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
This paper present analog front end comprising amplifier suitable for ECG acquisition. It is designed using 180nm CMOS SCL technology using Cadence Virtuoso environment with 1.8V power supply. ECG monitoring setup consists of protection circuit, an analog front end pre-amplifier, buffer, low pass filter and main amplifier. Protection circuit is used for protection from voltage surges which can occur between electrodes in electrocardiogram, during the application of defibrillator or due to electrostatic discharge (ESD) event. The ECG signal is amplified in stages, instrumentation amplifier and main amplifier. Low pass filter is designed for noise cancellation and buffer is used for impedance matching between two stages of ECG front end.