P. Conde, Antonio J. González, L. Hernández, P. Bellido, E. Crespo, A. Iborra, L. Moliner, J. Rigla, M. Rodríguez-Álvarez, F. Sánchez, M. Seimetz, A. Soriano, L. Vidal, J. Benlloch
{"title":"利用光分布拟合方法确定三维光子冲击","authors":"P. Conde, Antonio J. González, L. Hernández, P. Bellido, E. Crespo, A. Iborra, L. Moliner, J. Rigla, M. Rodríguez-Álvarez, F. Sánchez, M. Seimetz, A. Soriano, L. Vidal, J. Benlloch","doi":"10.1109/NSSMIC.2014.7430889","DOIUrl":null,"url":null,"abstract":"In Positron Emission Tomography (PET) detectors based in monolithic scintillators, the spatial resolution is limited by the accuracy in the determination of the interaction coordinates from the 511 keV photons. When linear algorithms, such as Center of Gravity (CoG) are used a poor estimation of the interaction positions, specially towards the edges is the major limitation in spatial resolution. A novel PET detector block, where complete information of Light Distribution (LD) for each event is available, allows to fit each event to a theoretical model, improving the estimation of the interaction coordinates, and minimizing border effects. In this work, by means of the LD fitting approach, we were able to obtain an average spatial resolution of 1.2 mm in the entire scintillator volume and an average depth of interaction (DOI) resolution of 1.5 mm. Moreover, splitting the data in three DOI regions, we obtained an average spatial resolution of 1.0 mm at the DOI region closer to the photodetectors. Finally, it is remarkable that the implementation of the LD fitting approach is capable of processing up to 50 kcps in a octacore system.","PeriodicalId":144711,"journal":{"name":"2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)","volume":"109 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2014-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"8","resultStr":"{\"title\":\"3-D photon impact determination using fitting approaches to the Light Distribution\",\"authors\":\"P. Conde, Antonio J. González, L. Hernández, P. Bellido, E. Crespo, A. Iborra, L. Moliner, J. Rigla, M. Rodríguez-Álvarez, F. Sánchez, M. Seimetz, A. Soriano, L. Vidal, J. Benlloch\",\"doi\":\"10.1109/NSSMIC.2014.7430889\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"In Positron Emission Tomography (PET) detectors based in monolithic scintillators, the spatial resolution is limited by the accuracy in the determination of the interaction coordinates from the 511 keV photons. When linear algorithms, such as Center of Gravity (CoG) are used a poor estimation of the interaction positions, specially towards the edges is the major limitation in spatial resolution. A novel PET detector block, where complete information of Light Distribution (LD) for each event is available, allows to fit each event to a theoretical model, improving the estimation of the interaction coordinates, and minimizing border effects. In this work, by means of the LD fitting approach, we were able to obtain an average spatial resolution of 1.2 mm in the entire scintillator volume and an average depth of interaction (DOI) resolution of 1.5 mm. Moreover, splitting the data in three DOI regions, we obtained an average spatial resolution of 1.0 mm at the DOI region closer to the photodetectors. Finally, it is remarkable that the implementation of the LD fitting approach is capable of processing up to 50 kcps in a octacore system.\",\"PeriodicalId\":144711,\"journal\":{\"name\":\"2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)\",\"volume\":\"109 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2014-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"8\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/NSSMIC.2014.7430889\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/NSSMIC.2014.7430889","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 8

摘要

在基于单片闪烁体的正电子发射层析成像(PET)探测器中,从511 keV光子确定相互作用坐标的精度限制了空间分辨率。当使用线性算法,如重心(CoG)时,对相互作用位置的估计很差,特别是对边缘的估计是空间分辨率的主要限制。一种新的PET检测器块,其中每个事件的光分布(LD)的完整信息是可用的,允许将每个事件拟合到理论模型中,改进相互作用坐标的估计,并最小化边界效应。在这项工作中,通过LD拟合方法,我们能够获得整个闪烁体体积的平均空间分辨率为1.2 mm,平均相互作用深度(DOI)分辨率为1.5 mm。此外,将数据分成三个DOI区域,我们在离光电探测器较近的DOI区域获得了平均1.0 mm的空间分辨率。最后,值得注意的是,LD拟合方法的实现能够在八核系统中处理高达50 kcps的数据。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
3-D photon impact determination using fitting approaches to the Light Distribution
In Positron Emission Tomography (PET) detectors based in monolithic scintillators, the spatial resolution is limited by the accuracy in the determination of the interaction coordinates from the 511 keV photons. When linear algorithms, such as Center of Gravity (CoG) are used a poor estimation of the interaction positions, specially towards the edges is the major limitation in spatial resolution. A novel PET detector block, where complete information of Light Distribution (LD) for each event is available, allows to fit each event to a theoretical model, improving the estimation of the interaction coordinates, and minimizing border effects. In this work, by means of the LD fitting approach, we were able to obtain an average spatial resolution of 1.2 mm in the entire scintillator volume and an average depth of interaction (DOI) resolution of 1.5 mm. Moreover, splitting the data in three DOI regions, we obtained an average spatial resolution of 1.0 mm at the DOI region closer to the photodetectors. Finally, it is remarkable that the implementation of the LD fitting approach is capable of processing up to 50 kcps in a octacore system.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信