D. Gamino, D. Lorias, A. Minor, R. Martinez
{"title":"用于微创手术无生命模型的手术器械插入记录系统","authors":"D. Gamino, D. Lorias, A. Minor, R. Martinez","doi":"10.1109/PAHCE.2013.6568267","DOIUrl":null,"url":null,"abstract":"This paper presents the results obtained from a recording system for the insertion of surgical instrumental which is guided by imaging equipment in minimally invasive surgeries focused on the lumbar region. The system involves hardware and software tools; the latter are capable of yielding angles, intracorporeal length, acceleration, and velocity of the instrumental by means of image processing. The system was subjected to tests in order to validate its precision and accuracy. The accuracy of intracorporeal length has an absolute error of less than ±1.5 mm, and a variation of ±1 mm in its precision (99%), whereas the angles can be measured with a precision of ±1° (99%).","PeriodicalId":151015,"journal":{"name":"2013 Pan American Health Care Exchanges (PAHCE)","volume":"584 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2013-07-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Recording system for the insertion of surgical instrumental applied to an inanimate model of minimally invasive surgery\",\"authors\":\"D. Gamino, D. Lorias, A. Minor, R. Martinez\",\"doi\":\"10.1109/PAHCE.2013.6568267\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This paper presents the results obtained from a recording system for the insertion of surgical instrumental which is guided by imaging equipment in minimally invasive surgeries focused on the lumbar region. The system involves hardware and software tools; the latter are capable of yielding angles, intracorporeal length, acceleration, and velocity of the instrumental by means of image processing. The system was subjected to tests in order to validate its precision and accuracy. The accuracy of intracorporeal length has an absolute error of less than ±1.5 mm, and a variation of ±1 mm in its precision (99%), whereas the angles can be measured with a precision of ±1° (99%).\",\"PeriodicalId\":151015,\"journal\":{\"name\":\"2013 Pan American Health Care Exchanges (PAHCE)\",\"volume\":\"584 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2013-07-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2013 Pan American Health Care Exchanges (PAHCE)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/PAHCE.2013.6568267\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2013 Pan American Health Care Exchanges (PAHCE)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/PAHCE.2013.6568267","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

本文介绍了一种在成像设备引导下的腰椎微创手术中手术器械插入记录系统的结果。该系统包括硬件和软件工具;后者能够通过图像处理的方式产生仪器的角度、体内长度、加速度和速度。对该系统进行了测试,以验证其精度和准确性。体内长度测量精度的绝对误差小于±1.5 mm,精度变化为±1 mm(99%),而角度测量精度为±1°(99%)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Recording system for the insertion of surgical instrumental applied to an inanimate model of minimally invasive surgery
This paper presents the results obtained from a recording system for the insertion of surgical instrumental which is guided by imaging equipment in minimally invasive surgeries focused on the lumbar region. The system involves hardware and software tools; the latter are capable of yielding angles, intracorporeal length, acceleration, and velocity of the instrumental by means of image processing. The system was subjected to tests in order to validate its precision and accuracy. The accuracy of intracorporeal length has an absolute error of less than ±1.5 mm, and a variation of ±1 mm in its precision (99%), whereas the angles can be measured with a precision of ±1° (99%).
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信